- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00821684036310), received between Mar 24, 2017 and Jan 13, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K061566); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 13, 2026Detachment of Device or Device Component
LIMITED INFORMATION REPORTED THAT DURING THE IMPLANT OF 2 VISIPRO STENTS ABOUT 8 YEARS AGO, A "PIECE OF THE BLADE BROKE OFF" AND HAS NOW MIGRATED THROUGH THE PATIENTS BODY. IT IS NOW BEHIND THEIR LEFT KNEE, BUT THE ORIGINAL POSITION WAS UNKNOWN AND THEY WERE ORIGINALLY TOLD THAT THE FOREIGN OBJECT WOULD NOT MIGRATE PAST THE STENT BUT IT NOW HAS. THE PATIENT DOES NOT REPORT ANY ISSUES ASSOCIATED WITH THIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24053294 ↗MalfunctionMar 24, 2017Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING AN ATTEMPT TO TREAT A LESION IN THE LEFT HYPOGASTRIC ARTERY USING A VISI PRO STENT, THE STENT COULD NOT CROSS THE LESION SUCCESSFULLY. IT WAS REPORTED THAT THE LESION WAS PRE-DILATED AND THAT NO EMBOLIC PROTECTION WAS USED. IT WAS REPORTED TO BE MODERATELY CALCIFIED, 10-15 MM IN LENGTH AND WITH A DIAMETER OF 6MM. THE PHYSICIAN REPORTED ENCOUNTERING RESISTANCE WHILE ADVANCING THE DEVICE AND THAT NO EXCESSIVE FORCE WAS USED. IT WAS REPORTED THAT THE PHYSICIAN REMOVED THE ENTIRE DELIVERY SYSTEM. DEFORMATION OF STENT WAS NOTED ONCE THE VISI-PRO HAD BEEN REMOVED FROM THE PATIENT AND WAS ON THE TABLE. IT WAS REPORTED THAT THE DEVICE TIP COMPONENT WAS DAMAGED WHILE ATTEMPTING TO DELIVER THE STENT THROUGH THE VESSEL. AN EVERFLEX 7X20MM STENT WAS USED TO COMPLETE THE PROCEDURE. IT WAS REPORTED THAT THERE WAS NO INJURY TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6431024 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K061566. One submission often covers many models, so check the product and lot details in each record.
- Z-1548-2008 · Jul 1, 2008 · Terminatedev3" sterile EO, single Use, Visi-Pro Balloon-Expandable Biliary Stent System. Model Numbers: PXB35-06-27-135, PXB35-06-37-080, PXB35-07-37-080, PXB35-08-17-080 , PXB35-08-27
- Z-1547-2008 · Jul 1, 2008 · Terminatedev3" 55442-2920, sterile EO, single Use, Visi-Pro Balloon-Expandable Peripheral Stent System. Model Numbers: PXB35-05-12-080, PXB35-06-17-135, PXB35-06-27-080, PXB35-06-57-135, PX
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.