Visi-Pro PXB35-07-37-080Recalls & reported problems
Covidien LP · Balloon-Expandable Biliary Stent System · Class 2 · Implant
All models of Visi-Pro (66)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00821684036426), received between Feb 2, 2018 and Jan 13, 2026 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K061566); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Detachment of Device or Device Component1
- Inaccurate Delivery1
- Premature Activation1
Patient problems
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 13, 2026Detachment of Device or Device Component
LIMITED INFORMATION REPORTED THAT DURING THE IMPLANT OF 2 VISIPRO STENTS ABOUT 8 YEARS AGO, A "PIECE OF THE BLADE BROKE OFF" AND HAS NOW MIGRATED THROUGH THE PATIENTS BODY. IT IS NOW BEHIND THEIR LEFT KNEE, BUT THE ORIGINAL POSITION WAS UNKNOWN AND THEY WERE ORIGINALLY TOLD THAT THE FOREIGN OBJECT WOULD NOT MIGRATE PAST THE STENT BUT IT NOW HAS. THE PATIENT DOES NOT REPORT ANY ISSUES ASSOCIATED WITH THIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24053250 ↗InjuryOct 16, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PHYSICIAN USED A VISI-PRO BALLOON EXPANDABLE STENT WITH A NON-MEDTRONIC 6F SHEATH AND A 0.035 STIFF GUIDEWIRE, TO TREAT A SEVERELY CALCIFIED LESION WITH 90% STENOSIS IN THE MID RIGHT COMMON ILIAC ARTERY. THE DEVICE WAS PREPPED AS PER IFU WITH NO ISSUES IDENTIFIED. THE LESION WAS NOT PRE-DILATED. NO EMBOLIC PROTECTION WAS USED. IT WAS REPORTED THAT RESISTANCE WAS ENCOUNTERED ADVANCING THE DEVICE TO THE LESION BUT NO EXCESSIVE FORCE WAS USED. IT WAS REPORTED THAT THE STENT CAME OFF THE BALLOON DELIVERY SYSTEM. SEVERAL ATTEMPTS TO ADVANCE THE BALLOON INTO THE STENT WERE UNSUCCESSFUL. A NEW 7X40 BALLOON WAS ADVANCED AND USED TO DILATE THE STENT AT THE TARGET LESION SITE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7969380 ↗InjuryMar 23, 2018Premature Activation; Inaccurate Delivery; Premature Activation; Inaccurate Delivery
THE PHYSICIAN WAS USING ONE VISIPRO DEVICE TO TREAT A SLIGHTLY TORTUOUS FIBROUS LESION, LOCATED IN THE MID RIGHT FEMORAL / FEMORAL BYPASS AREA, EXHIBITING 75% STENOSIS. EMBOLIC PROTECTION WAS NOT USED. THE DEVICE WAS REMOVED FROM ITS PACKAGING AND PREPPED PER THE IFU WITH NO ISSUES NOTED. THE LESION WAS NOT PREDILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT, RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE AND EXCESSIVE FORCE WAS NOT USED. IT WAS REPORTED THAT WHILE THE PHYSICIAN WAS ADVANCING THE DEVICE TO THE LESION SITE, THE STENT DEPLOYED AT AN UNINTENDED SITE. THE PHYSICIAN USED A LONGER SHEATH AND DEPLOYED A SECOND STENT IN THE ORIGINAL APPROPRIATE POSITION. NO PATIENT INJURY WAS REPORTED.
Patient problems: Patient Problem/Medical Problem
MAUDE report 7365581 ↗MalfunctionFeb 2, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE PHYSICIAN WAS ATTEMPTING TO USE A VISI-PRO BALLOON EXPANDABLE STENT TO TREAT A MODERATELY TORTUOUS AND MODERATELY CALCIFIED LEFT PROXIMAL COMMON ILIAC ARTERY LESION EXHIBITING 95% STENOSIS. A NON-MDT GUIDEWIRE WAS USED. EMBOLIC PROTECTION WAS NOT USED. THE DEVICE WAS REMOVED FROM ITS PACKAGING AND INSPECTED WITH NO ISSUES NOTED. THE IFU WAS FOLLOWED DURING THE PROCEDURE. IT WAS REPORTED THAT AS THE PHYSICIAN ATTEMPTED TO ADVANCE THE DEVICE TOWARDS THE LESION, IT COULD BE SEEN UNDER FLUOROSCOPY THAT THE STENT WAS BEGINNING TO MIGRATE ON THE BALLOON. THE PHYSICIAN WAS ABLE TO SUCCESSFULLY REMOVE THE STENT OVER THE WIRE AND THROUGH THE SHEATH WITH NO FURTHER MIGRATION AND NO INJURY TO THE PATIENT. THE PHYSICIAN REPLACED THE DEVICE WITH ANOTHER VISI-PRO STENT AND COMPLETED THE PROCEDURE WITHOUT FURTHER COMPLICATIONS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7239181 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K061566. One submission often covers many models, so check the product and lot details in each record.
- Z-1547-2008 · Jul 1, 2008 · Terminatedev3" 55442-2920, sterile EO, single Use, Visi-Pro Balloon-Expandable Peripheral Stent System. Model Numbers: PXB35-05-12-080, PXB35-06-17-135, PXB35-06-27-080, PXB35-06-57-135, PX
- Z-1548-2008 · Jul 1, 2008 · Terminatedev3" sterile EO, single Use, Visi-Pro Balloon-Expandable Biliary Stent System. Model Numbers: PXB35-06-27-135, PXB35-06-37-080, PXB35-07-37-080, PXB35-08-17-080 , PXB35-08-27
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page