Visi-Pro PXB35-07-57-080Recalls & reported problems
Covidien LP · Balloon-Expandable Biliary Stent System · Class 2 · Implant
All models of Visi-Pro (66)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821684036440), received between Sep 6, 2017 and Sep 6, 2017 — classified as 1 injury.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K061566); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated1
- Difficult or Delayed Positioning1
Patient problems
- Hematoma1
- Hemorrhage/Blood Loss/Bleeding1
- Patient Problem/Medical Problem1
- Perforation1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 6, 2017Difficult or Delayed Positioning; Device Dislodged or Dislocated; Difficult or Delayed Positioning; Device Dislodged or Dislocated
THE PHYSICIAN WAS ATTEMPTING TO USE A VISI-PRO BALLOON EXPANDABLE STENT TO TREAT A LESION WITH A SEVERE DEGREE OF TORTUOSITY AND CALCIFICATION AND DEMONSTRATING (B)(4) STENOSIS AT THE COMMON ILIAC ARTERY AND SUPERFICIAL FEMORAL ARTERY (SFA). THE DEVICE WAS PREPPED PER THE IFU. NO EMBOLIC PROTECTION WAS USED DURING THE PROCEDURE. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE AND EXCESSIVE FORCE USED DURING DELIVERY AND WITHDRAWAL. IT WAS REPORTED THAT DEPLOYMENT ISSUES WERE EXPERIENCED AND THE PHYSICIAN WAS UNABLE TO DEPLOY THE STENT RESULTING IN A DELIVERY FAILURE. THE STENT DISLODGED DURING REMOVAL, WHICH REQUIRED INTERVENTION. THIS DISLODGED STENT WAS NOT REMOVED AND IT REMAINED IN PLACE. IT WAS REPORTED THAT A BALLOON WAS PASSED THROUGH THE DISLODGED STENT AND IT WAS DEPLOYED. THE PHYSICIAN USED ANOTHER STENT TO COMPLETE THE PROCEDURE. PATIENT INJURY WAS REPORTED AND COMPLICATIONS FOLLOWING THE INCIDENT INCLUDED VESSEL DAMAGE/INJURY, PERFORATION, HEMATOMA AND HEMORRHAGE/BLEED AT THE ACCESS SITE. PATIENT EXPERIENCED RETROPERITONEAL BLEEDING. A COMMON FEMORAL CUT-DOWN PROCEDURE WAS PERFORMED.
Patient problems: Hematoma; Hemorrhage/Blood Loss/Bleeding; Patient Problem/Medical Problem; Perforation
MAUDE report 6844191 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K061566. One submission often covers many models, so check the product and lot details in each record.
- Z-1547-2008 · Jul 1, 2008 · Terminatedev3" 55442-2920, sterile EO, single Use, Visi-Pro Balloon-Expandable Peripheral Stent System. Model Numbers: PXB35-05-12-080, PXB35-06-17-135, PXB35-06-27-080, PXB35-06-57-135, PX
- Z-1548-2008 · Jul 1, 2008 · Terminatedev3" sterile EO, single Use, Visi-Pro Balloon-Expandable Biliary Stent System. Model Numbers: PXB35-06-27-135, PXB35-06-37-080, PXB35-07-37-080, PXB35-08-17-080 , PXB35-08-27
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page