Visi-Pro PXB35-08-17-080Recalls & reported problems
Covidien LP · Balloon-Expandable Biliary Stent System · Class 2 · Implant
All models of Visi-Pro (66)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00821684036464), received between Oct 25, 2017 and Dec 5, 2018 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K061566); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Difficult To Position1
- Malposition of Device1
Patient problems
- No Consequences Or Impact To Patient1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 5, 2018Malposition of Device; Malposition of Device
PHYSICIAN USED A VISI PRO BALLOON EXPANDABLE STENT DURING TREATMENT OF A 20MM PLAQUE LESION IN THE PATIENT¿S PROXIMAL COMMON ILIAC ARTERY OF DIAMETER 8MM. MODERATE TORTUOSITY AND CALCIFICATION WERE REPORTED. THE IFU WAS FOLLOWED AND THE DEVICE WAS PREPPED WITHOUT ISSUE. PRE-DILATION WAS CARRIED OUT USING A SIZE 5 PRE-DILATION DEVICE. NO EMBOLIC PROTECTION WAS USED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT, NO EXCESSIVE FORCE WAS USED BUT RESISTANCE WAS EXPERIENCED. IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED IN VIVO DURING POSITIONING AS THE STENT CAME OFF THE BALLOON. THE STENT WAS DEPLOYED IN THE INTENDED LESION IN THE PATIENTS ILIAC USING ANOTHER PTA.
Patient problems: Patient Problem/Medical Problem
MAUDE report 8132386 ↗MalfunctionOct 25, 2017Difficult To Position; Detachment of Device or Device Component; Difficult To Position; Detachment of Device or Device Component
THE PHYSICIAN WAS ATTEMPTING TO USE A VISI-PRO BALLOON EXPANDABLE STENT TO TREAT A 2 MM PLAQUE LESION WITH MODERATE CALCIFICATION IN THE PROXIMAL CELIAC ARTERY. THE LESION HAD 90% STENOSIS. THE VESSEL WAS 6 MM IN DIAMETER AND MODERATELY TORTUOUS. THE DEVICE WAS PREPPED AS PER THE IFU WITH NO ISSUES NOTED. THE LESION WAS NOT PRE-DILATED. THE PROCEDURE USED A 6X70 SHEATH. IT WAS REPORTED THAT THE PHYSICIAN EXPERIENCED DIFFICULTY IN ADVANCING THE DEVICE AND THE STENT DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. NO EXCESSIVE FORCE WAS USED. THE VISI-PRO STENT WOULD NOT ADVANCE THROUGH THE LESION AND THE PHYSICIAN IMMEDIATELY PULLED THE STENT BACK INTO THE SHEATH AND THE STENT CAME OFF. IT WAS REPORTED THAT THE UNDEPLOYED STENT WAS TAKEN OUT ALONG WITH THE SHEATH AND THE PROCEDURE WAS COMPLETED USING A NEW 6X90 SHEATH AND A NEW STENT 6X27 VISI-PRO BALLOON EXPANDABLE STENT. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6978921 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K061566. One submission often covers many models, so check the product and lot details in each record.
- Z-1548-2008 · Jul 1, 2008 · Terminatedev3" sterile EO, single Use, Visi-Pro Balloon-Expandable Biliary Stent System. Model Numbers: PXB35-06-27-135, PXB35-06-37-080, PXB35-07-37-080, PXB35-08-17-080 , PXB35-08-27
- Z-1547-2008 · Jul 1, 2008 · Terminatedev3" 55442-2920, sterile EO, single Use, Visi-Pro Balloon-Expandable Peripheral Stent System. Model Numbers: PXB35-05-12-080, PXB35-06-17-135, PXB35-06-27-080, PXB35-06-57-135, PX
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page