EverCross™ AB35W07040080 (UDI 00821684061831)Recalls & reported problems
Ev3, Inc · PTA AB35W07040080 EVERCROSS 035 V05 · Class 2
All models of EverCross (114)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00821684061831), received between Aug 8, 2018 and Nov 6, 2018 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
1 other recall cover devices cleared under the same FDA submission (K103322); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Burst Container or Vessel1
- Detachment of Device or Device Component1
- Difficult to Remove1
- Inflation Problem1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 6, 2018Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT THE PHYSICIAN ATTEMPTED TO USE A 7X40 EVERCROSS BALLOON TO TREAT A MODERATELY CALCIFIED SOFT TISSUE LESION IN THE MID LEFT SUPERFICIAL FEMORAL ARTERY. THE IFU WAS FOLLOWED AND THE DEVICE WAS PREPPED WITHOUT ISSUE. IT WAS REPORTED THAT THE BALLOON BURST DURING INFLATION AT 8 ATM. IT WAS REPORTED THAT THE PHYSICIAN THEN ATTEMPTED TO USE A 7X60 EVERCROSS BALLOON. IT WAS REPORTED THAT THE EVERCROSS BALLOON BURST DURING INFLATION AT 8 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER EVERCROSS 7X60. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8042654 ↗InjuryAug 8, 2018Inflation Problem; Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove
IT WAS REPORTED THAT THE PHYSICIAN WAS ATTEMPTING TO USE AN EVERCROSS BALLOON DILATION CATHETER TO TREAT AN AV FISTULA CLOT IN THE RIGHT ARM. THE DEVICE WAS REPORTED TO HAVE BEEN PREPPED AS PER IFU. THERE WAS NO DAMAGE NOTED TO THE PACKAGING OR TO THE DEVICE UPON REMOVAL FROM THE TRAY. NO EMBOLIC PROTECTION WAS USED. NO RESISTANCE WAS ENCOUNTERED ADVANCING THE DEVICE AND THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. THE VESSEL DIAMETER WAS REPORTED TO BE 7MM. POST BALLOON INFLATION DIFFICULTIES WERE EXPERIENCED DURING REMOVAL. THE BALLOON CATHETER AND TIP WERE REPORTED TO HAVE DETACHED AT THE CATHETER SHAFT INSIDE THE BALLOON. THESE WERE SNARED AND REMOVED, USING 2 X GOOSENECK SNARES - A 10MM LOOP AND A 25MM LOOP. HOWEVER, A PORTION OF THE BALLOON REMAINED IN THE PATIENT & IT WAS REPORTED THAT THE FISTULA FLOW WAS NOT FULLY RESTORED. THE DECISION WAS THEN MADE TO HOLD PRESSURE AND OCCLUDE THE FISTULA COMPLETELY.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient
MAUDE report 7761257 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K103322. One submission often covers many models, so check the product and lot details in each record.
- Z-1489-2014 · Apr 9, 2014 · Terminatedev3 EverCross 0.035" OTW PTA Dilatation Catheter, 6mm x 120mm x 135cm, REF: AB35W06120135. and ev3 EverCross 0.035" OTW PTA Dilatation Catheter, 6mm x 40mm x, 80cm REF: AB35W
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page