Skip to content

EverCross™ AB35W09060080 (UDI 00821684062173)Recalls & reported problems

Ev3, Inc · PTA AB35W09060080 EVERCROSS 035 V05 · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821684062173), received between Sep 21, 2017 and Sep 21, 2017 — classified as 1 injury.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

1 other recall cover devices cleared under the same FDA submission (K103322); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Burst Container or Vessel1

Patient problems

  • Patient Problem/Medical Problem1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 21, 2017Burst Container or Vessel; Burst Container or Vessel

    THE PHYSICIAN PERFORMED A BILATERAL STENTING PROCEDURE OF BOTH RIGHT AND LEFT COMMON ILIAC ARTERIES WITH UNKNOWN STENTS USING ¿KISSING¿ TECHNIQUE. IT WAS REPORTED THAT POST STENT DEPLOYMENT, THE NEED FOR FURTHER ANGIOPLASTY WAS IDENTIFIED IN THE RIGHT MID COMMON ILIAC ARTERY. THE LESION WAS CALCIFIED AND THE ARTERY HAD LITTLE TORTUOSITY. AN EVERCROSS 9MMX40MM PTA BALLOON WAS USED TO PERFORM THE ANGIOPLASTY PROCEDURE. THE DEVICE WAS REMOVED FROM ITS PACKAGING AND INSPECTED WITH NO ISSUES NOTED. THE DEVICE WAS PREPPED AS PER THE IFU. NO EMBOLIC PROTECTION WAS USED. IT WAS REPORTED THAT THIS BALLOON BURST ON INFLATION WITHOUT ANY INCIDENT. A SECOND EVERCROSS 9MMX60MM PTA BALLOON WAS SELECTED FOR USE. IT WAS REPORTED THAT A CIRCUMFERENTIAL/ RADIAL BURST OCCURRED DURING INFLATION FOR THIS BALLOON. IT WAS NOTED THAT AN INCH OF THE TIP WAS MISSING. THIS WAS CONFIRMED BY FURTHER IMAGING WHICH DEMONSTRATED THE BROKEN PIECE STILL IN SITU. THE PHYSICIAN HAD TROUBLE IN REMOVING THE FRAGMENTS OF THE BALLOON FROM THE ARTERY. THIS BROKEN TIP COULDN'T BE RETRIEVED VIA SNARE. A SURGICAL INTERVENTION UNDER GENERAL ANESTHESIA WAS REQUIRED TO REMOVE FRAGMENTS. THE SURGERY WAS REPORTED TO BE SUCCESSFU

    Patient problems: Patient Problem/Medical Problem

    MAUDE report 6881802 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K103322. One submission often covers many models, so check the product and lot details in each record.

  • Z-1489-2014 · Apr 9, 2014 · Terminated
    ev3 EverCross 0.035" OTW PTA Dilatation Catheter, 6mm x 120mm x 135cm, REF: AB35W06120135. and ev3 EverCross 0.035" OTW PTA Dilatation Catheter, 6mm x 40mm x, 80cm REF: AB35W

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page