Skip to content

EverCross™ AB35W10040135 (UDI 00821684062265)Recalls & reported problems

Ev3, Inc · PTA AB35W10040135 EVERCROSS 035 V05 · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821684062265), received between Jun 7, 2018 and Jun 7, 2018 — classified as 1 injury.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Burst Container or Vessel1

Patient problems

  • Patient Problem/Medical Problem1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 7, 2018Burst Container or Vessel; Burst Container or Vessel

    THE PHYSICIAN WAS USING AN EVERCROSS BALLOON WITH A NON-MEDTRONIC 8F SHEATH AND A 0.035 NON-MEDTRONIC GUIDEWIRE FOR TREATMENT OF A 200 MM, 70% (STENOSED) SOFT TISSUE CLOT IN THE PATIENT¿S MID LEFT COMMON FEMORAL VEIN. THE VESSEL DIAMETER WAS REPORTED TO BE BETWEEN 16-18 MM. A NON-MEDTRONIC INFLATION DEVICE WAS USED WITH ISOVUE AND SALINE 1:3. THE DEVICE WAS PREPPED WITHOUT ISSUE. THE DEVICE DID PASS THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS ENCOUNTERED NOR EXCESSIVE FORCE USED WHEN ADVANCING THE DEVICE TO THE LESION. IT WAS REPORTED THAT THE BALLOON BURST AT A PRESSURE OF 9 ATMS FOR 10 SECONDS ON FIRST INFLATION. THE BALLOON FRAGMENTS WERE RETRIEVED WITH CYSTOGRASPERS AND NON MEDTRONIC CATHETERS WERE PLACED BILATERALLY. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Patient Problem/Medical Problem

    MAUDE report 7577667 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →