EverCross™ AB35W10060135 (UDI 00821684062289)Recalls & reported problems
Ev3, Inc · PTA AB35W10060135 EVERCROSS 035 V05 · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821684062289), received between Mar 1, 2019 and Mar 1, 2019 — classified as 1 malfunction.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
1 other recall cover devices cleared under the same FDA submission (K103322); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Burst Container or Vessel1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 1, 2019Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT A PHYSICIAN WAS ATTEMPTING TO USE AN EVERCROSS 035 BALLOON DILATATION CATHETER TO TREAT A LESION IN THE PROXIMAL COMMON ILIAC ARTERY WITH SEVERE CALCIFICATION AND MODERATE TORTUOSITY WITH 85% STENOSIS. ARTERY DIAMETER REPORTED AS 10MM, LESION LENGTH 50MM. A NON-MEDTRONIC 7FR SHEATH AND 0.035 GUIDEWIRE WERE ALSO USED. NO DAMAGE NOTED TO PACKAGING NOTED, NO ISSUES NOTED WHEN REMOVING THE DEVICES FROM THE HOOP/TRAY, DEVICES WAS PREPPED AS PER IFU. A NON-MEDTRONIC INFLATION DEVICE WAS USED FOR BALLOON INFLATION. IT WAS REPORTED THAT A CIRCUMFERENTIAL/RADIAL BALLOON BURST OCCURRED AT 9ATMS DURING SECOND INFLATION. RESISTANCE WAS NOT ENCOUNTERED, NO EXCESSIVE FORCED USED. THERE WAS NO INTERVENTION REQUIRED TO REMOVE THE BALLOON FRAGMENTS AND NO VESSEL DAMAGE. A NEW BALLOON WAS USED TO COMPLETE THE SURGERY. NO PATIENT INJURY REPORTED FOR THIS EVENT
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8381672 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K103322. One submission often covers many models, so check the product and lot details in each record.
- Z-1489-2014 · Apr 9, 2014 · Terminatedev3 EverCross 0.035" OTW PTA Dilatation Catheter, 6mm x 120mm x 135cm, REF: AB35W06120135. and ev3 EverCross 0.035" OTW PTA Dilatation Catheter, 6mm x 40mm x, 80cm REF: AB35W
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.