The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821684062340), received between Sep 6, 2019 and Sep 6, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

1 other recall cover devices cleared under the same FDA submission (K103322); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Burst Container or Vessel1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 6, 2019Burst Container or Vessel; Burst Container or Vessel

    PHYSICIAN ATTEMPTED TO USE AN EVERCROSS PTA BALLOON WITH A NON-MEDTRONIC 6FR SHEATH (COOK) AND A NON-MEDTRONIC 0.035' GUIDEWIRE (ABBOTT) TO TREAT A MODERATELY CALCIFIED LESION PRESENTING 70% STENOSIS IN THE MID SUPERFICIAL FEMORAL ARTERY. ARTERY DIAMETER AND LESION LENGTH WERE 5MM AND 10 MM, RESPECTIVELY. THE DEVICE WAS PREPPED PER THE IFU. A NON-MEDTRONIC INFLATION DEVICE (BOSTON) WAS USED TO INFLATE THE BALLOON. THE BALLOON DID NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS EXPERIENCED DURING ADVANCEMENT. IT WAS REPORTED THAT A LONGITUDINAL BALLOON BURST OCCURRED DURING FIRST BALLOON INFLATION AT 7 ATM. ANOTHER EVERCROSS BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE IS NO PATIENT INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8968862 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K103322. One submission often covers many models, so check the product and lot details in each record.

  • Z-1489-2014 · Apr 9, 2014 · Terminated
    ev3 EverCross 0.035" OTW PTA Dilatation Catheter, 6mm x 120mm x 135cm, REF: AB35W06120135. and ev3 EverCross 0.035" OTW PTA Dilatation Catheter, 6mm x 40mm x, 80cm REF: AB35W