Skip to content

NanoCross™ Elite AB14W030080150 (UDI 00821684066447)Recalls & reported problems

Ev3, Inc · PTA AB14W030080150 NANOCROSS 014 V04 · Class 2

All models of NanoCross Elite (63)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821684066447), received between Nov 8, 2017 and Nov 8, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Burst Container or Vessel1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 8, 2017Burst Container or Vessel; Burst Container or Vessel

    PHYSICIAN INTENDED TO USE A NANOCROSS PTA DILATATION CATHETER FOR THE TREATMENT OF A SEVERELY CALCIFIED CHRONIC TOTAL OCCLUSION (CTO) LESION IN THE SFA. A 6 FRENCH NON-MEDTRONIC 55CM SHEATH AND A 0.035¿ NON-MEDTRONIC GUIDEWIRE WERE USED. NO EMBOLIC PROTECTION WAS USED. AN UNKNOWN INFLATION DEVICE WAS USED WITH CONTRAST AND HEPARINIZED SALINE. DEVICE WAS PREPPED WITHOUT ISSUE. IT WAS REPORTED THAT A CIRCUMFERENTIAL/RADIAL BALLOON BURST OCCURRED AT AN INFLATION PRESSURE OF 14ATMS DURING THE PROCEDURE. DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE BUT NO EXCESSIVE FORCE WAS USED. PROCEDURE COMPLETED BY PTA AND STENTING WITH 2 EVERFLEX 200MM STENTS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7012250 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →