Skip to content

Protégé™ GPS™ SERB65-09-20-120 (UDI 00821684071397)Recalls & reported problems

Ev3, Inc · STENT SERB65-09-20-120 PROTEGE GPS V05 · Class 2 · Implant

All models of Protégé GPS (26)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821684071397), received between Feb 22, 2019 and Feb 22, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Premature Activation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Premature Activation1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 22, 2019Premature Activation; Premature Activation

    IT WAS REPORTED THAT A PHYSICIAN WAS ATTEMPTING TO USE A PROTEGE GPS SELF-EXPANDING STENT WITH A 6FR SHEATH TO TREAT A LESION IN THE MID BILIARY ARTERY WITH NO TORTUOSITY OR CALCIFICATION. AN 0.035" WHOLEY WORE WAS USED. NO ISSUES WITH PACKAGING, NO ISSUES IN REMOVING THE DEVICE FROM THE HOOP/TRAY. DEVICE WAS PREPPED AS PER IFU. IT WAS REPORTED THAT DEPLOYMENT ISSUES OCCURRED, PARTIAL DEPLOYMENT OCCURRED. THERE WAS NO DAMAGE TO THE DEPLOYMENT MECHANISMS OR DEVICE HANDLE PRIOR TO DEPLOYMENT ISSUE, THE THUMBSCREW/LOCK-PIN WAS CHECKED FOR SECUREMENT PRIOR TO PROCEDURE. THE LESION WAS NOT PRE-DILATED AND THE STENT DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE, NO EXCESSIVE FORCE WAS USED. THIS PATIENT HAD A BANDING PROCEDURE (OPEN-HEART) IMMEDIATELY PRECEDING THIS STENTING PROCESS. THE PHYSICIAN ADVANCED THE FIRST STENT WITH NO ISSUE. THE PATIENT BEGAN TO BLEED. AFTER THE CV SURGEON REPAIRED THE BLEEDING THE PHYSICIAN RESTARTED. THE PROTÉGÉ WOULDN'T ADVANCE. IT WAS THEN REPORTED THAT THE DISTAL EDGE OF THE STENT WAS OUT OF THE DELIVERY SYSTEM. A NEW STENT WAS OPENED, PREPPED AND DEPLOYED WITHOUT ISSUE. NO PATIENT INJURY REPORTED

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8361086 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page