Bagley MR-6LARecalls & reported problems
Gyrus ACMI, LLC · Bagley Self-Dilating Autoclavable Semi-Rigid Ureteroscope · Class 2
All models of Bagley (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00821925006232), received between May 31, 2023 and Sep 2, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Material Twisted/Bent2
- Crack1
- Dent in Material1
- Failure to Clean Adequately1
- Material Deformation1
- No Display/Image1
- Overheating of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 2, 2026Crack
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50573104 ↗MalfunctionJul 16, 2026Break
THE CUSTOMER RETURNED THE URETEROSCOPE AND ACCESSORIES, FLEXIBLE, RIGID TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE EVALUATION, THE OBJECTIVE WINDOW WAS FOUND TO BE BROKEN INSIDE THE OBJECTIVE DISTAL CHANNELS. THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE MOST LIKELY CAUSE OF THE COMPLAINT WAS TRACED TO COMPONENT FAILURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25846794 ↗MalfunctionMay 22, 2026Overheating of Device
THE URETEROSCOPE AND ACCESSORIES WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF WHEN DISCONNECTING THE TWO PIECES, IT WAS EXTREMELY DIFFICULT TO SEPARATE. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, BURN FIBERS WERE FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25262350 ↗MalfunctionApr 3, 2026Material Deformation; Failure to Clean Adequately
IT WAS OBSERVED DURING DEVICE EVALUATION THAT THE URETEROSCOPE AND ACCESSORIES (FLEXIBLE, RIGID) HAD WHITE FLUID IN THE DISTAL CHANNEL AND A BENT DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24781612 ↗MalfunctionApr 1, 2026Dent in Material
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE URETEROSCOPE AND ACCESSORIES, FLEXIBLE/RIGID PRESENTED SHARP DENT ON THE DISTAL EDGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24755471 ↗MalfunctionNov 7, 2025Break; Poor Quality Image; Break; Poor Quality Image
IT WAS REPORTED THAT THE SUBJECT DEVICE HAD A BLURRY VIEW THROUGH THE SCOPE AND DAMAGE TO THE LENS. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23494335 ↗MalfunctionJul 10, 2025Material Twisted/Bent; Material Twisted/Bent
DURING THE DEVICE EVALUATION, THE URETEROSCOPE EXHIBITED A BENT OUTER TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22474751 ↗MalfunctionSep 16, 2024Break; No Display/Image; Break; No Display/Image
IT WAS REPORTED THE URETEROSCOPE DISTAL TIP WAS DAMAGED AND/OR BROKEN WITH A SHARP EDGE, COULD NOT SEE THE PICTURE. THERE WERE NO REPORTS OF PATIENT HARM OR IMPACT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20233737 ↗MalfunctionJan 29, 2024Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED, THE URETEROSCOPE DISTAL TIP IS BENT. THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18596486 ↗MalfunctionMay 31, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS, THE URETEROSCOPE WAS FOUND TO HAVE A BLURRY IMAGE DURING DEVICE REPROCESSING FOLLOWING A THERAPEUTIC PROCEDURE. INSPECTION AND TESTING OF THE RETURNED DEVICE FOUND DISTAL TIP DAMAGE. THERE WAS NO PATIENT INVOLVEMENT. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17034883 ↗
FDA clearance and classification
Cleared through Special 510(k) K052044 on Aug 16, 2005 (decision: Substantially Equivalent). Submitted by Acmi Corporation as “Acmi Autoclavable Ureteroscope, Models Mr-6a And Mr-6al”. The FDA review took 19 days. FDA record ↗
The FDA regulates product code FGB (ureteroscope and accessories, flexible/rigid) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Common questions about the Bagley MR-6LA
Has the Bagley MR-6LA been recalled?
No FDA recall record in our data names this exact device identifier (DI 00821925006232), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Bagley MR-6LA?
The FDA's MAUDE database lists 10 adverse event reports for this device, received between May 31, 2023 and Sep 2, 2026. In those reports, the most common specific device problems are “Break” (4), “Material Twisted/Bent” (2) and “Crack” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Bagley MR-6LA?
None of the reports are classified as deaths, 0 as injuries and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Bagley MR-6LA FDA approved?
It was cleared through Special 510(k) K052044 on Aug 16, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Bagley MR-6LA, and is it still sold?
The FDA's device database lists Gyrus ACMI, LLC as the company responsible for it (brand Bagley, model MR-6LA). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Ureteroscope and accessories, flexible/rigid devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall1,122 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls393 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall303 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls187 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls170 reports
- ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-190-10-3DOlympus Medical Systems Corp. · Ureteroscope And Accessories, Flexible/Rigid186 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P7ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall121 reports
- n.a. 091279-01Karl Storz GmbH & Co. KG · Ureteroscope and accessories, flexible/rigid111 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page