Bagley REXMR-6ARecalls & reported problems
Gyrus ACMI, LLC · Bagley Self-Dilating Autoclavable Semi-Rigid Ureteroscope · Class 2
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00821925007826), received between Jun 22, 2021 and Apr 26, 2024 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Sharp Edges2
- Overheating of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Burn(s)1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 26, 2024Break; Break
IT WAS REPORTED THE URETEROSCOPE'S TIP WAS DAMAGED, AND IT WAS SHARP. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19192259 ↗MalfunctionJul 26, 2022Sharp Edges; Sharp Edges
CUSTOMER REPORTED WITH AN ISSUE OF "ROUGH PLACE ON END OF SHAFT HAS BURR". THE ISSUE FOUND DURING REPROCESSING. THERE WAS NO PATIENT HARM, NO USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15107042 ↗InjuryJun 9, 2022Overheating of Device; Overheating of Device
THE CUSTOMER REPORTS DURING AN UNSPECIFIED PROCEDURE USING A SEMI RIGID URETEROSCOPE, THE SCOPE GOT HOT WHEN CONNECTED TO THE LIGHT CORD AND THE PATIENT EXPERIENCED A BURN ON HIS PENIS. THE SEVERITY OF THE BURN IS UNKNOWN. IT IS UNKNOWN IF WHAT/IF THE PATIENT REQUIRED ANY TREATMENT OR INTERVENTION FOR THE BURN. ADDITIONAL DETAILS HAVE BEEN REQUESTED FROM THE CUSTOMER, AT THIS TIME, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Burn(s)
MAUDE report 14644865 ↗MalfunctionJul 14, 2021Break; Break
IT WAS REPORTED THAT THE SCOPE IS DAMAGED. THE ISSUE FOUND DURING AN UNKNOWN EVENT. THERE WAS NO PATIENT INVOLVEMENT REPORTED DUE TO THE EVENT. NO USER INJURY REPORTED. DEVICE RETURN EVALUATION FOUND THE OUTER TUBE AT THE DISTAL TIP OF THE DEVICE WAS DAMAGED WITH A SHARP EDGES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12167743 ↗MalfunctionJun 22, 2021Sharp Edges; Sharp Edges
THE SERVICE CENTER WAS INFORMED THAT A CUSTOMER URETEROSCOPE, 33CM WAS RETURNED DUE TO A REPORT OF "POOR VISIBILITY". UPON INSPECTION AND TESTING, THE DISTAL END OF THE OUTER TUBE WAS OBSERVED TO BE DENTED AND SHARP. NO PATIENT HARM OR INVOLVEMENT WAS REPORTED TO OLYMPUS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12041313 ↗
Other Ureteroscope and accessories, flexible/rigid devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall1,125 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls397 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall305 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls190 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls170 reports
- ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-190-10-3DOlympus Medical Systems Corp. · Ureteroscope And Accessories, Flexible/Rigid186 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P7ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall121 reports
- n.a. 091279-01Karl Storz GmbH & Co. KG · Ureteroscope and accessories, flexible/rigid118 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.