Cook G09500Recalls & reported problems
Cook Incorporated · Biliary Drainage Catheter Ultrathane · Class U · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00827002095007), received between Apr 18, 2017 and Jun 30, 2022 — classified as 5 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Material Separation” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Material Separation3
- Break2
- Detachment of Device or Device Component2
- Leak/Splash2
- Component Falling1
- Insufficient Device Problem Information1
Patient problems
- No Code Available5
- No Known Impact Or Consequence To Patient2
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- No Information1
- No Patient Involvement1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 30, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A BILE DUCT DRAINAGE PROCEDURE AN ULTRATHANE MAC-LOC LOCKING LOOP BILIARY DRAINAGE CATHETER SEPARATED AT THE HUB WHEN THE USER WAS FIXING THE CATHETER TO THE PATIENT'S SKIN. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING EVENT DETAILS HAS BEEN REQUESTED, BUT IS CURRENTLY UNAVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14877048 ↗MalfunctionMay 5, 2020Material Separation; Material Separation
AS REPORTED, A (B)(6) YEAR-OLD MALE PATIENT REQUIRED PLACEMENT OF AN ULTRATHANE MAC-LOC LOCKING LOOP BILIARY DRAINAGE CATHETER DURING A BILIARY DRAINAGE EXCHANGE PROCEDURE. WHILE PREPARING THE DRAIN FOR PLACEMENT, THE OPERATOR REPORTED THE HUB CAME OFF WHEN INSERTING THE PLASTIC STIFFENER. THE CATHETER DID NOT MAKE PATIENT CONTACT. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10026484 ↗InjurySep 6, 2018Material Separation; Material Separation
IT WAS REPORTED AN ULTRATHANE MAC-LOC LOCKING LOOP BILIARY DRAINAGE CATHETER WAS PLACED INTO THE TRANSPLANTED KIDNEY AND BLADDER OF A PATIENT DURING A NEPHROURETERAL DRAINAGE PROCEDURE. HUB SEPARATION FROM THE CATHETER WAS REPORTED TO HAVE HAPPENED TWO TIMES, ONE WEEK APART. IN THE SECOND OCCURRENCE, IT WAS FOUND POST PROCEDURE THAT THE CATHETER HAD SEPARATED FROM THE HUB. THE DEFECTIVE CATHETER WAS REMOVED AND REPLACED WITH A LIKE DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE FIRST OCCURRENCE IS DETAILED IN MEDWATCH 1820334-2018-02359.
Patient problems: No Code Available
MAUDE report 7852462 ↗InjurySep 6, 2018Material Separation; Material Separation
IT WAS REPORTED THAT AN ULTRATHANE MAC-LOC LOCKING LOOP BILIARY DRAINAGE CATHETER WAS PLACED INTO THE TRANSPLANTED KIDNEY AND BLADDER OF A PATIENT DURING A NEPHROURETERAL DRAINAGE PROCEDURE AT AN UNSPECIFIED DATE. HUB SEPARATION FROM THE CATHETER WAS REPORTED TO HAVE HAPPENED TWO TIMES, ONE WEEK APART. IN THE FIRST OCCURRENCE, IT WAS FOUND POST PROCEDURE THAT THE CATHETER HAD SEPARATED FROM THE HUB. THE DEFECTIVE CATHETER WAS REMOVED AND REPLACED WITH A LIKE DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE SECOND OCCURRENCE IS DETAILED IN MEDWATCH 1820334-2018-02532.
Patient problems: No Code Available
MAUDE report 7852370 ↗InjuryAug 20, 2018Leak/Splash; Leak/Splash
IT WAS REPORTED THE PATIENT RETURNED ON (B)(6) 2018 DUE TO LEAKING OF THE ULTRATHANE MAC-LOC LOCKING LOOP BILIARY DRAINAGE CATHETER . THE DRAINAGE CATHETER WAS INITIALLY PLACED ON (B)(6) 2018 FOR A BILIARY DRAIN PROCEDURE. THE DEVICE WAS SUCCESSFULLY REPLACED WITH A SIMILAR DEVICE FROM A DIFFERENT LOT. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EVENTS DUE TO THIS OCCURRENCE.
Patient problems: No Code Available
MAUDE report 7796661 ↗InjuryNov 29, 2017Break; Break
THE CUSTOMER REPORTED THAT THE PATIENT HAD TO RETURN FOR ADDITIONAL MEDICAL INTERVENTION WHEN THE HUB OF THE ULTRATHANE MAC-LOC LOCKING LOOP BILIARY DRAINAGE CATHETER BROKE FOLLOWING SURGERY. THE DEVICE WAS REPORTEDLY IMPLANTED ON (B)(6) 2017 AND REPLACED ON (B)(6) 2017. ADDITIONAL INFORMATION WAS REQUESTED FROM BUT NOT PROVIDED BY THE CUSTOMER. THE PRODUCT WAS RETURNED FOR EVALUATION; HOWEVER, AS OF THE DATE OF THIS REPORT, THE INVESTIGATION IS STILL PENDING.
Patient problems: No Code Available
MAUDE report 7069242 ↗MalfunctionNov 14, 2017Insufficient Device Problem Information
THE DEVICE LEAKED NO ADVERSE EFFECTS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7028198 ↗MalfunctionNov 14, 2017Leak/Splash
THE DEVICE LEAKED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Information
MAUDE report 7028177 ↗MalfunctionNov 9, 2017Component Falling
THE COMPLAINANT REPORTED THAT A DRESSING WAS BEING PLACED ON THE CATHETER AND THE HUB FELL OFF AFTER THE BILIARY DRAIN PLACEMENT WAS OVER. THE HUB DISCONNECTED WHERE THE CATHETER AND HUB MEET. THE PATIENT WAS STILL ON THE TABLE FROM THE INITIAL CATHETER PLACEMENT. AS A RESULT OF THE MALFUNCTION, THE DOCTOR PLACED A NEW CATHETER AND THE PATIENT WAS ABLE TO LEAVE THE ROOM. THERE WERE NO REPORTS OF ANY SECTION OF THE DEVICE REMAINING INSIDE THE PATIENT'S BODY AND THE INITIAL REPORTER NOTED THAT THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS AS A RESULT OF THIS PRODUCT PROBLEM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7014418 ↗InjuryJul 20, 2017Break; Break
INTERNATIONAL CUSTOMER REPORTED THAT A PATIENT WAS UNDERGOING PLACEMENT OF AN ULTRATHANE MAC-LOC LOCKING LOOP BILIARY DRAINAGE CATHETER FOR AN UNSPECIFIED INDICATION. THE CUSTOMER REPORTS THAT "THE DRAIN BROKE ITSELF DURING THE PROCEDURE." THE BREAK OCCURRED AT THE "CROSSING OF THE TUBE AND THE MACLOCK SYSTEM." THE DEVICE WAS EXPLANTED AND EXCHANGED WITH A NEW DEVICE DURING THE INITIAL IMPLANT PROCEDURE. THE HANDLING CONDITIONS AND CIRCUMSTANCES SURROUNDING THE EVENT WERE NOT PROVIDED. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE
Patient problems: No Code Available
MAUDE report 6728075 ↗MalfunctionApr 18, 2017Detachment of Device or Device Component; Detachment of Device or Device Component
CUSTOMER REPORTS EXAMINING A NEW DEVICE THAT HAD NOT TOUCHED ANY PATIENT AFTER HAVING A PATIENT WITH MULTIPLE HUB DETACHMENTS OF THE ULTRATHANE MAC-LOC LOCKING LOOP BILIARY DRAINAGE CATHETER. THE FACILITY DECIDED TO "TEST A FRESHLY OPENED DEVICE (WITH NO PATIENT CONTACT) AND THE HUB POPPED OFF WITH APPROXIMATELY 3 POUNDS OF FORCE." THE TESTING CONDITIONS AND ACTUAL FORCES APPLIED TO THE DEVICE ARE NOT KNOW; THE TEST WAS NOT CONDUTED USING CALIBRATED MEASURING EQUIPMENT. THERE WAS NO PATIENT INVOLVEMENT FOR THIS EVALUATION. THE DEVICE IS NOT AVAILABLE FOR EVALUATION. THIS IS ONE OF TWO REPORTS BEING SUBMITTED AS THE CUSTOMER REPORTED TWO SEPARATE EVENTS. SEE MDR NUMBERS 1820334-2017-00678 AND 1820334-2017-00679 FOR ALL ASSOCIATED REPORTS. THE SAME CUSTOMER IS INVOLVED FOR BOTH EVENTS.
Patient problems: No Patient Involvement
MAUDE report 6501207 ↗
Other CATHETER, NEPHROSTOMY devices
- Cook G09501Cook Incorporated · CATHETER, NEPHROSTOMY280 reports
- Dawson Mueller G09705Cook Incorporated · CATHETER, NEPHROSTOMY2 deaths215 reports
- Cook G09497Cook Incorporated · CATHETER, NEPHROSTOMY129 reports
- Dawson Mueller G11020Cook Incorporated · CATHETER, NEPHROSTOMY100 reports
- Cook G09502Cook Incorporated · CATHETER, NEPHROSTOMY86 reports
- NephroMax Kit M0062101180Boston Scientific Corporation · CATHETER, NEPHROSTOMY1 recall25 reports
- NephroMax Kit M0062101600Boston Scientific Corporation · CATHETER, NEPHROSTOMY1 recall11 reports
- Dawson Mueller G13768Cook Incorporated · CATHETER, NEPHROSTOMY59 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page