Skip to content

NForce G19072Recalls & reported problems

Cook Incorporated · NForce, Nitinol Helical Stone Extractor · Class 2

All models of NForce (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00827002190726), received between Aug 30, 2017 and Dec 24, 2019 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Difficult to Open or Close” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Difficult to Open or Close2
  • Break1
  • Device Damaged Prior to Use1
  • Difficult or Delayed Positioning1

Patient problems

  • No Consequences Or Impact To Patient5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 24, 2019Difficult to Open or Close; Difficult to Open or Close

    NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9516344 ↗
  • MalfunctionMay 8, 2019Device Damaged Prior to Use; Device Damaged Prior to Use

    THERE IS NO NEW EVENT INFORMATION TO REPORT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8590631 ↗
  • MalfunctionMay 11, 2018Difficult or Delayed Positioning; Difficult or Delayed Positioning

    IT WAS REPORTED THAT THE TIP OF THE NFORCE NITINOL HELICAL STONE EXTRACTOR DID NOT RETRACT BACK WITHIN THE SHEATH. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.  ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7506550 ↗
  • MalfunctionFeb 28, 2018Break; Break

    IT WAS REPORTED DURING A URETEROSCOPY PROCEDURE BEING PERFORMED ON A (B)(6), FEMALE PATIENT, THE NFORCE NITINOL HELICAL STONE EXTRACTOR BROKE WHILE IN USE. AS REPORTED, ONE OF THE WIRES BROKE AT THE DISTAL END OF THE SHEATH. THE DEVICE WAS REMOVED. ANOTHER NFORCE NITONAL HELICAL STONE EXTRACTOR WAS OPENED AND WOULD NOT CLOSE RIGHT OUT OF THE PACKAGE. THIS DEVICE WAS NOT USED ON THE PATIENT. AS REPORTED, THERE WAS NO UNINTENDED PART OF THE DEVICE LEFT IN THE PATIENT¿S BODY. NO ADDITIONAL PROCEDURES WERE REQUIRED AND THERE WERE NO ADVERSE EFFECTS TO THE PATIENT BECAUSE OF THESE OCCURRENCES. TWO REPORTS HAVE BEEN FILED TO CAPTURE THE TWO REPORTED FAILURE MODES OCCURRING DURING THE SAME PROCEDURE. THE BROKEN BASKET WIRE IS REPORTED IN MFR. REPORT # 1820334-2018-00590. THE BASKET NOT CLOSING IN MFR. REPORT # 1820334-2018- 00591.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7303151 ↗
  • MalfunctionAug 30, 2017Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT THE PHYSICIAN WAS PERFORMING AN URETERORENOLITHOTRIPSY PROCEDURE BY USING AN NFORCE NITINOL HELICAL STONE EXTRACTOR DEVICE. THE PHYSICIAN NOTICED THAT THE BASKET OF THE DEVICE WOULD NOT CLOSE PROPERLY. IT WAS FURTHER REPORTED THAT THE PHYSICIAN HAD TO CUT CATHETER TO REMOVE THE DEVICE FROM THE PATIENT. THE PRODUCT WAS REPLACED AND THE INTENDED PROCEDURE WAS COMPLETED SUCCESSFULLY. NO UNINTENDED SECTION OF THE DEVICE REMAINED INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION WAS REQUESTED FROM THE CUSTOMER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6829594 ↗

FDA clearance and classification

The FDA regulates product code FFL (dislodger, stone, basket, ureteral, metal) under 21 CFR 876.4680 as a Class II device. 21 CFR 876.4680 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NForce G19072

Has the NForce G19072 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00827002190726), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NForce G19072?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Aug 30, 2017 and Dec 24, 2019. In those reports, the most common specific device problems are “Difficult to Open or Close” (2), “Break” (1) and “Device Damaged Prior to Use” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NForce G19072?

None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NForce G19072 FDA approved?

Devices of this type (product code FFL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cook Incorporated.

Who makes the NForce G19072, and is it still sold?

The FDA's device database lists Cook Incorporated as the company responsible for it (brand NForce, model G19072). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other DISLODGER, STONE, BASKET, URETERAL, METAL devices

All DISLODGER, STONE, BASKET, URETERAL, METAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page