Cook G21466Recalls & reported problems
Cook Ireland Ltd · Nasal Pancreatic Drainage Set · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002214668), received between Nov 2, 2017 and Jul 3, 2019 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
- Restricted Flow rate1
Patient problems
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJul 3, 2019Restricted Flow rate; Restricted Flow rate
THE DEVICE WAS USED FOR DRAINAGE DUE TO CHRONIC PANCREATITIS. DRAINAGE AND COLLECTION OF PANCREATIC JUICE COULD NOT BE MADE DUE TO A KINK IN THE CATHETER NEAR THE PAPILLA. (IT CANNOT BE DETERMINED IF THE USER NOTICED THE KINK DURING DEVICE INSERTION OR AFTER DEVICE PLACEMENT.) AFTER LEAVING THE KINKED CATHETER PLACED IN THE PANCREAS DUCT FOR A WHILE, THE USER REMOVED IT AND PLACED ANOTHER CATHETER. (THE DATE OF REMOVAL CANNOT BE DETERMINED.) THERE HAVE BEEN NO ADVERSE EFFECTS TO THE PATIENT REPORTED. FDA MDR REPORTING REQUIRED: THIS EVENT MEETS THE CRITERIA OF AN FDA ¿SERIOUS INJURY¿ REPORT AS PER FDA GUIDELINES ¿MEDICAL DEVICE REPORTING FOR MANUFACTURERS (2016)¿ SECTION 2.13 AND 2.15 AS SURGICAL INTERVENTION WAS REQUIRED TO REMOVE KINKED STENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8757793 ↗MalfunctionOct 4, 2018Break; Break
THIS FILE IS BEING REPORTED UNDER THE FDA MDR MALFUNCTION PRECEDENCE "CATHETER FRACTURE/ BREAKAGE" THE NPDS CATHETER WAS MOVED TO CHANGE THE POSITION OF THE NPDS CATHETER ON THE X-RAY. THEN, THE CATHETER WAS BREAK OFF. SO THEY REMOVED THE CATHETER. THEY WENT THROUGH THE PROCEDURE AGAIN FROM THE BEGINNING. THEN, ANOTHER NPDS CATHETER WAS INSERTED AND ENDED THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7933357 ↗MalfunctionNov 2, 2017Break; Break
THIS FOLLOW UP MDR IS BEING SUBMITTED TO UPDATE THE INVESTIGATION CONCLUSIONS. THE DRAINAGE CATHETER CAUGHT SOMEWHERE/SOMETHING AND BROKE DURING THE PLACEMENT PROCEDURE. ADDITIONAL INFORMATION PROVIDED BY THE SALES REP ON OCT-10TH-2017: THE USER PLACED THE DRAINAGE CATHETER IN THE PATIENT WHO HAD PANCREAS CYST WHICH WAS LOCATED VERY PROXIMALLY. DUE TO THE LOCATION OF THE PANCREAS CYST, THE DRAINAGE HOLES OF THE CATHETER WERE LOCATED IN THE DUODENUM. SO HE THOUGHT IT WOULD NOT DRAIN PROPERLY AND DECIDED TO REMOVE IT. DURING REMOVING, THE CATHETER GOT CAUGHT BY SOMETHING AND STRETCHED, BUT HE KEPT PULLING IT AND THE CATHETER TIP GOT BROKE, BUT A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PROCEDURE WAS ENDED WITHOUT PLACING OTHER CATHETERS. ADDITIONAL INFORMATION PROVIDED BY THE SALES REP ON OCT-27TH-2017: I TALKED THE DOCTOR WHO WAS AT THE PROCEDURE AS AN ASSISTANT DOCTOR AND OBTAINED THE INFORMATION BELOW. - THE PROXIMAL SIDE OF THE CATHETER WAS CUT AFTER PLACING IT AT THE TARGET SITE. - WHILE THE CATHETER WAS PLACED, ESWL WAS PERFORMED. DURING PERFORMING IT, THE DOCTOR DETERMINED TO REMOVED THE CATHETER BECAUSE HE THOUGHT IT WOULD NOT DRAIN PROPERLY DUE TO
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6997882 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page