Leung G21520Recalls & reported problems
Cook Ireland Ltd · Leung, Nasal Biliary Drainage Set · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002215207), received between May 2, 2019 and May 2, 2019 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break2
Patient problems
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 2, 2019Break; Break
AS REPORTED TO CUSTOMER RELATIONS " THE INITIAL ERCP CANNULATION WAS PERFORMED. THE GUIDE WIRE WAS PASSED INTO THE COMMON BILE DUCT WITHOUT PROBLEMS. THE FIRST NASAL BILIARY DEVICE (B)(4) WAS FED OVER THE GUIDE WIRE (VISIGLIDE .025 - 450). THE TIP OF THE DEVICE BROKE (SEPARATED) WHILE FEEDING THE WIRE THROUGH. THIS OCCURRED OUTSIDE OF THE PATIENT AND SCOPE. THE DEVICE WAS REMOVED FROM THE WIRE AND SAVED TO SEND BACK TO THE MANUFACTURER. THE SECOND DEVICE (B)(4) WAS FED OVER THE GUIDE WIRE THROUGH THE SCOPE WITHOUT RESISTANCE AND WAS ADVANCED INTO THE COMMON BILE DUCT. THE GUIDE WIRE WAS REMOVED AND ASPIRATION BEGAN. WHEN THE GUIDE WIRE WAS ADVANCED BACK THROUGH THE CATHETER AND THE TUBE WAS EXTRACTED, A TWO INCH PIECE OF THE TIP REMAINED INSIDE THE PATIENT. ADDITIONAL ATTEMPTS TO REMOVE THE PIECE IN THE LIVER WAS DIFFICULT, RESULTING IN ADDITIONAL PROCEDURE TIME. A SPYGLASS DEVICE WAS INSERTED OVER THE WIRE TO RETRIEVE THE DISLODGED PIECE WITH SUCCESS. THE PROCEDURE WAS COMPLETED WITH A STANDARD BILIARY STENT." FDA MDR REPORTING REQUIRED: EVENT IS FDA MDR REPORTABLE BASED ON THE REQUIREMENT FOR SURGICAL INTERVENTION IN ORDER TO REMOVE THE 2 INCH PIECE OF THE TIP FROM THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8572620 ↗MalfunctionMay 2, 2019Break; Break
AS REPORTED TO CUSTOMER RELATIONS " THE INITIAL ERCP CANNULATION WAS PERFORMED. THE GUIDE WIRE WAS PASSED INTO THE COMMON BILE DUCT WITHOUT PROBLEMS. THE FIRST NASAL BILIARY DEVICE (B)(4) WAS FED OVER THE GUIDE WIRE (VISIGLIDE .025 - 450). THE TIP OF THE DEVICE BROKE (SEPARATED) WHILE FEEDING THE WIRE THROUGH. THIS OCCURRED OUTSIDE OF THE PATIENT AND SCOPE. THE DEVICE WAS REMOVED FROM THE WIRE AND SAVED TO SEND BACK TO THE MANUFACTURER. THE SECOND DEVICE (B)(4) WAS FED OVER THE GUIDE WIRE THROUGH THE SCOPE WITHOUT RESISTANCE AND WAS ADVANCED INTO THE COMMON BILE DUCT. THE GUIDE WIRE WAS REMOVED AND ASPIRATION BEGAN. WHEN THE GUIDE WIRE WAS ADVANCED BACK THROUGH THE CATHETER AND THE TUBE WAS EXTRACTED, A TWO INCH PIECE OF THE TIP REMAINED INSIDE THE PATIENT. ADDITIONAL ATTEMPTS TO REMOVE THE PIECE IN THE LIVER WAS DIFFICULT, RESULTING IN ADDITIONAL PROCEDURE TIME. A SPYGLASS DEVICE WAS INSERTED OVER THE WIRE TO RETRIEVE THE DISLODGED PIECE WITH SUCCESS. THE PROCEDURE WAS COMPLETED WITH A STANDARD BILIARY STENT." FDA MDR REPORTING REQUIRED: REPORTING BASED ON THE DECISION FOR PR (B)(4) (SURGICAL INTERVENTION) AS THIS WILL ESTABLISH A PRECEDENCE FOR ENBD 'CATHETER FRACTURE/BREAKAGE'.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8572610 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page