Zimmon G21526Recalls & reported problems
Cook Ireland Ltd · Zimmon, Biliary Stent Set · Class 2 · Implant
All models of Zimmon (43)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00827002215269), received between Apr 25, 2018 and Aug 21, 2019 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Deformation4
- Kinked2
- Break1
Patient problems
- No Known Impact Or Consequence To Patient4
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 21, 2019Material Deformation; Material Deformation
THE USER STRAIGHTENED THE STENT WITH A STRAIGHTENER SLOWLY, THEN ATTEMPTED TO ADVANCE IT OVER A WIRE GUIDE. HOWEVER, THE STENT COULD NOT ADVANCE OVER THE WIRE GUIDE DUE TO A KINK IN THE PIGTAIL. ANOTHER ZEBD-7-12 WAS USED INSTEAD AND THE PROCEDURE WAS COMPLETED. THERE HAVE BEEN NO ADVERSE EFFECTS TO THE PATIENT REPORTED. FDA MDR REPORTING REQUIRED: EVENT IS FDA MDR REPORTABLE BASED ON THE DEVICE MALFUNCTION REPORTING PRECEDENCE FOR THIS DEVICE FAMILY FOR THE ISSUE OF ¿STENT KINKED/BENT¿. NO ADVERSE EFFECTS TO THE PATIENT HAVE BEEN REPORTED AS OCCURRING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8915868 ↗MalfunctionMay 16, 2019Material Deformation; Material Deformation
AFTER PLACING OLYMPUS'S WIRE GUIDE "VISIGLIDE 0.025INCH" IN THE B6 AREA, THE PHYSICIAN STRAIGHTENED THE STENT WITH THE STRAIGHTENER IN ORDER TO ADVANCE THE STENT OVER THE WIRE GUIDE. HOWEVER, BECAUSE THE WIRE GUIDE COULD NOT PASS THROUGH THE STENT DUE TO KINK, ANOTHER STENT WAS USED INSTEAD. THERE HAVE BEEN NO ADVERSE EFFECTS TO THE PATIENT REPORTED. FDA MDR REPORTING REQUIRED: EVENT IS FDA MDR REPORTABLE BASED ON THE DEVICE MALFUNCTION REPORTING PRECEDENCE FOR THIS DEVICE FAMILY FOR THE ISSUE OF ¿STENT KINKED/BENT'. NO ADVERSE EFFECTS TO THE PATIENT HAVE BEEN REPORTED AS OCCURRING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8615231 ↗MalfunctionMay 13, 2019Material Deformation; Material Deformation
THE USER STRAIGHTENED THE PIG TAIL (ZEBD-7-12/ LOT C1554025) USING THE STRAIGHTENER AND ATTEMPTED TO ADVANCE IT OVER A WIRE GUIDE (M-THROUGH 0.025 ANGLE TIP / ASAHI INTECC) BUT THE WIRE GUIDE WOULD NOT PASS THROUGH THE STENT LUMEN. HE FOUND THERE WAS A KINK IN THE PIG TAIL AND REPLACED IT WITH ANOTHER ZEBD-7-12 TO COMPLETE THE PROCEDURE. FDA MDR REPORTING REQUIRED: EVENT IS FDA MDR REPORTABLE BASED ON THE DEVICE MALFUNCTION REPORTING PRECEDENCE FOR THIS DEVICE FAMILY FOR THE ISSUE OF ¿STENT KINKED/BENT'. NO ADVERSE EFFECTS TO THE PATIENT HAVE BEEN REPORTED AS OCCURRING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8603031 ↗MalfunctionJan 9, 2019Material Deformation; Material Deformation
THE STENT KINKED WHEN THE USER WAS STRAIGHTENING THE PIG TAIL USING THE STRAIGHTENER. ANOTHER DEVICE WAS USED INSTEAD FOR THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8230215 ↗MalfunctionJun 14, 2018Break; Kinked; Break; Kinked
THE USER STRAIGHTENED THE PIG TAIL OF THE STENT USING THE STRAIGHTENER AND ATTEMPTED TO ADVANCE IT OVER A WIRE GUIDE (OLYMPUS' VISIGLIDE ANGLED 0.025 INCH) THAT WAS PLACED IN THE HEPATIC PORTAL REGION. BUT THE WIRE GUIDE WOULD NOT BE INSERTED INTO THE STENT LUMEN DUE TO A KINK OF THE PIG TAIL. HE REPLACED IT WITH ANOTHER ZEBD-7-12 TO COMPLETE THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7600740 ↗MalfunctionApr 25, 2018Kinked; Kinked
UPON OPENING THE PACKAGE, THE USER FOUND THAT THE STENT WAS KINKED ALREADY. HE USED ANOTHER ZEBD-7-12 FOR THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7458342 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page