Skip to content

Cotton-Huibregtse G21881Recalls & reported problems

Cook Ireland Ltd · Cotton-Huibregtse Biliary Stent · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002218819), received between Jan 4, 2018 and Jan 4, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Kinked” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Kinked1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 4, 2018Kinked; Kinked

    DURING INSERTION OF STENT AND INTRODUCER INTO SCOPE, STENT WOULD BUCKLE UPON INSERTION INTO WIRE LOCKING CAP. TECH REMOVED THE STENT AND GRABBED ANOTHER STENT AND PHYSICIAN PLACED STENT WITHOUT DIFFICULTY. THEY WENT TO PLACE ANOTHER STENT FOR THIS PROCEDURE AND IT DID NOT BUCKLE. IT WAS PLACED SUCCESSFULLY. DM OBSERVED PHYSICIAN HAVING DIFFICULTY INSERTING FIRST STENT INTO SCOPE. THEN DM GOT CLOSER TO PHYSICIAN AND NOTICED STENT BUCKLING AT THE DISTAL FLAP.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7163216 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page