Zimmon G21982Recalls & reported problems

Cook Ireland Ltd · Zimmon Biliary Stent · Class 2 · Implant

All models of Zimmon (43)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002219823), received between Aug 10, 2025 and Aug 10, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Off-Label Use” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Off-Label Use1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 10, 2025Off-Label Use; Off-Label Use

    OPERATOR HAD DIFFICULTY DEPLOYING STENT OVER WIRE GUIDE AND AS STENT WAS VERY STIFF AND WAS BUCKLING UNDER PRESSURE. I HAVE BEEN ADVISED BY THE DOCTOR THAT DURING A PROCEDURE ON (B)(6) 2025, THAT THE 10FR X 9MM ZIMMON STENT WAS TOO STIFF. WHEN TRYING TO STRAIGHTEN OVER STIFF GUIDEWIRE, THE STENT BUCKLES, AT RISK OF FRACTURE, RATHER THAN BEING MALLEABLE. A 7FR X 12MM ZIMMON BILIARY STENT WAS USED INSTEAD. COOK CELECT PLATINUM VENA CAVA SET FOR FEMORAL AND JUGULAR VEIN APPROACH. 1. DOES THE COMPLAINT RELATE TO DEVICE REPLACEMENT, DEVICE REMOVAL OR WAS IT AN OBSERVATION MADE PRIOR TO PATIENT CONTACT? DEVICE PLACEMENT. 2. WHAT WAS THE TARGET LOCATION FOR THE STENT? GALL BLADDER. 3. PLEASE DESCRIBE THE STORAGE CONDITIONS OF THE DEVICE PRIOR TO USE ESPECIALLY THOSE PERTAINING TO TEMPERATURE AND LIGHT EXPOSURE. AS PER IFU¿S. 4. WHAT IS THE REORDER NUMBER, DIAMETER AND LENGTH OF THE WIRE GUIDE THAT WAS USED WITH THIS DEVICE IN THIS PROCEDURE? TERUMO 0.0035 STIFF WIRE GUIDE. 5. WAS THE WIRE GUIDE LUBRICATED PRIOR TO USE? YES. 6. WAS THE WIRE GUIDE INSPECTED FOR DAMAGE PRIOR TO USE? YES. 7. WAS THE DEVICE AT THE CENTRE OF THE COMPLAINT INSPECTED FOR DAMAGE PRIOR TO USE? NO. 8. WERE PREVIOUS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22768299 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page