Skip to content

Geenen G22107Recalls & reported problems

Cook Ireland Ltd · Geenen Pancreatic Stent · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002221079), received between Oct 5, 2017 and Nov 27, 2018 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break2
  • Use of Device Problem1

Patient problems

  • Foreign Body In Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 27, 2018Break; Use of Device Problem; Break; Use of Device Problem

    THIS FILE IS BEING REPORTED BASED ON THE FDA MDR MALFUNCTION PRECEDENCE "STENT FRACTURE". THE GPSO-5-3 COULDN'T GET IN THE PANCRATIC DUCT BECAUSE THE GPSO-5-3 DISTAL FLAP WAS BROKEN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8105123 ↗
  • InjuryOct 5, 2017Break; Break

    THIS FOLLOW UP MDR IS BEING SUBMITTED TO INCLUDE THE INVESTIGATION CONCLUSIONS. STENT BROKEN. "AS PER COMPLAINT FORM: IN (B)(6) 2017, THE STENT WAS IMPLANTED BY THE SAME HOSPITAL AND PHYSICIAN. ON THE (B)(6) THE STENT SHOULD BE REMOVED. STENT WAS IN AN OPTIMAL POSITION, BUT COMPLETELY DISCOLOURED DARK. WHEN REMOVING, THE PART OF THE STENT WAS IN THE DUODENUM, FELL APART INTO THE SMALLEST PIECES. REST PIECE WITHOUT HARM WAS IN THE PANCREATIC DUCT.

    Patient problems: Foreign Body In Patient

    MAUDE report 6917313 ↗