Skip to content

Oasis G22150Recalls & reported problems

Cook Ireland Ltd · Oasis, One Action Stent Introduction System · Class 2

All models of Oasis (15)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002221505), received between Apr 19, 2018 and Jul 21, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Operational Issue” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Operational Issue1
  • Improper or Incorrect Procedure or Method1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 21, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    THE DOCTOR INTENDED TO USE TTSO-8.5-13 ALONG WITH OA-8.5 FOR DRAINAGE IN THE CBD STRICTURE AREA. AS USUAL, TTSO-8.5-13 WAS PRELOADED INTO OA-8.5 AND APPROACHED INTO THE CBD. THE INNER GUIDING CATHETER WAS POSITIONED ALONG THE WIRE, BUT DURING THE PROCESS OF INSERTING THE STENT, THE INNER GUIDING CATHETER DID NOT SEPARATE WELL AND THE SHEATH STRETCHED. THERE WAS SIGNIFICANT RESISTANCE, AND AFTER MUCH DIFFICULTY, TTSO-8.5-13 WAS POSITIONED, AND THE INNER GUIDING CATHETER WAS SEPARATED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22561895 ↗
  • MalfunctionDec 12, 2019Use of Device Problem; Use of Device Problem

    THE DEVICE WAS USED FOR BILIARY STENT PLACEMENT. THE TARGET SITE WAS THE LOWER BILE DUCT, AND THE STENT COULD REACH THERE WITH A PUSHER. HOWEVER, IT COULD NOT BE RELEASED. THEREFORE, ANOTHER MANUFACTURE¿S DEVICE WAS USED INSTEAD.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9456767 ↗
  • MalfunctionApr 19, 2018Device Operational Issue; Device Operational Issue

    LEFT HEPATICUS WAS REACHED BY WIRE GUIDE, THAN USED DILATOR 8,5 FR AFTER DILATATION WAS USED OA-8.5 WITH PLASTIC STENT PBS. IT WASN'T POSSIBLE TO GO THROUGH AND THERE WAS BROKEN GUIDING AND ALSO PUSHING CATHETER. WIRE GUIDE WAS STUCK INSIDE THE GUIDING CATHETER. ALL DEVICES WERE REMOVED FROM THE PATIENT. SECOND ATTEMPT TO DRAIN THE LEFT HEPATICUS WASN'T PERFORMED. PATIENT HAS DRAINED RIGHT HEPATICUS ONLY. REPORTING DUE TO THE PRECEDENCE OF "DIFFICULT ADVANCEMENT" FOR THIS PRODUCT FAMILY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7442749 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page