Zimmon G22164Recalls & reported problems
Cook Ireland Ltd · Zimmon Biliary Stent · Class 2 · Implant
All models of Zimmon (43)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00827002221642), received between May 8, 2019 and Jan 29, 2020 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Deformation3
- Difficult to Advance1
- Material Twisted/Bent1
Patient problems
- No Known Impact Or Consequence To Patient4
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 29, 2020Material Twisted/Bent; Material Twisted/Bent
WHEN THEY USED WIRE GUIDE IN TO THE ZSO-7-6 AFTER USING THE STRAIGHTENER FOR STENT WAS STRAIGHTENED WITH THE PIGTAIL, WIRE GUIDE COULD NOT INSERT IN TO THE ZSO-7-6. BECAUSE PIGTAIL WAS KINKED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9642463 ↗MalfunctionOct 8, 2019Material Deformation; Material Deformation
EVENT IS FDA MDR REPORTABLE BASED ON THE DEVICE MALFUNCTION REPORTING PRECEDENCE FOR THIS DEVICE FAMILY FOR THE ISSUE OF ¿STENT KINKED/ BENT'. WHEN THEY USED WIRE GUIDE IN TO THE ZSO-7-6 AFTER USING THE STRAIGHTENER FOR STENT WAS STRAIGHTENED WITH THE PIGTAIL, WIRE GUIDE COULD NOT INSERT IN TO THE ZSO-7-6. BEACUASE PIGTAIL WAS KINKED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9165123 ↗MalfunctionSep 18, 2019Material Deformation; Material Deformation
EVENT IS FDA MDR REPORTABLE BASED ON THE DEVICE MALFUNCTION REPORTING PRECEDENCE FOR THIS DEVICE FAMILY FOR THE ISSUE OF STENT KINKED/BENT. WHEN THEY USED WIRE GUIDE IN TO THE ZSO-7-6 AFTER USING THE STRAIGHTENER FOR STENT WAS STRAIGHTENED WITH THE PIGTAIL, WIRE GUIDE COULD NOT INSERT IN TO THE ZSO-7-6. BECAUSE PIGTAIL WAS KINKED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9081817 ↗MalfunctionAug 12, 2019Difficult to Advance; Difficult to Advance
EVENT IS FDA MDR REPORTABLE BASED ON THE DEVICE MALFUNCTION REPORTING PRECEDENCE FOR THIS DEVICE FAMILY FOR THE ISSUE OF ¿DIFFICULT ADVANCEMENT'. THEY TRIED TO PUT THE WIRE GUIDE IN THE ZSO-7-6 FOR PLACE THE STENT IN THE BILE DUCT, BUT THE WIRE GUIDE COULD NOT PASS THROUGH IN THE ZSO-7-6.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8882581 ↗MalfunctionMay 8, 2019Material Deformation; Material Deformation
THEY TRIED TO PUT THE WIRE GUIDE IN THE ZSO-7-6 FOR PLACE THE STENT IN THE BILE DUCT, BUT THE WIRE GUIDE COULD NOT PASS THROUGH IN THE ZSO-7-6. FDA MDR REPORTING REQUIRED: EVENT IS FDA MDR REPORTABLE BASED ON THE DEVICE MALFUNCTION REPORTING PRECEDENCE FOR THIS DEVICE FAMILY FOR THE ISSUE OF 'STENT KINKED/BENT' AND ¿DIFFICULT ADVANCEMENT'. NO ADVERSE EFFECTS TO THE PATIENT HAVE BEEN REPORTED AS OCCURRING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8591430 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page