Soehendra G22205Recalls & reported problems
Cook Incorporated · Soehendra Stent Retriever · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002222052), received between Jun 7, 2016 and Aug 28, 2018 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Material Separation1
- Off-Label Use1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 28, 2018Off-Label Use; Detachment of Device or Device Component; Off-Label Use; Detachment of Device or Device Component
DURING A ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A SOEHENDRA STENT RETRIEVER. THE CUSTOMER LIKE TO DRILL THE PANCREAS DUCT AND THE TIP OF THE RETRIEVER GOT LOST [SIC]. THE FOLLOWING ADDITIONAL INFORMATION WAS PROVIDED ON 08-AUGUST-2018: THE USER DILATED WITH THE STENT RETRIEVER. THEY TURNED THE RETRIEVER CLOCKWISE IN THE PANCREATIC DUCT. SURGERY IS AN OPTION TO REMOVE THE DETACHED TIP, BUT HAS NOT BEEN DONE YET. AN ATTEMPT WAS MADE TO RETRIEVE THE TIP, BUT WAS UNSUCCESSFUL. OUR ATTEMPTS TO COLLECT ADDITIONAL INFORMATION ON IF THE PATIENT REQUIRED SURGERY WERE UNSUCCESSFUL. IT IS CURRENTLY UNKNOWN IF SURGERY WAS REQUIRED, IF THE TIP OF THE DEVICE WAS LEFT TO PASS NATURALLY, OR THE IMPACT TO THE PATIENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 7822898 ↗MalfunctionJun 7, 2016Material Separation; Material Separation
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK SOEHENDRA STENT RETRIEVER. THE CUSTOMER REPORTED THAT THE CONNECTION THREAD OF THE STENT RETRIEVER DETACHED FROM CATHETER. DUE TO THIS, AN ADDITIONAL PROCEDURE HAD TO BE PERFORMED IN ORDER CORRECT THE FAULT AND REMOVE STENT. THE FOLLOWING WAS RECEIVED ON (B)(6) 2016 WITH THE COMPLAINT FORM: "THE PLAN WAS TO REMOVE THE PLASTIC STENT WITH THE STENT RETRIEVER AND PLACE A NEW STENT AFTERWARDS. WHEN THE STENT RETRIEVER WAS IN PLACE, THE FRONT PART DETACHED FROM THE CATHETER AND STAYED IN THE PLASTIC STENT. I WILL RETURN EVERYTHING, INCLUDING THE STENT. THEREFORE, THEY HAD TO TAKE A SNARE TO REMOVE THE STENT [AND THIS IS] WHAT CAUSED A LOSS OF THE WIRE POSITION [ANOTHER MANUFACTURER'S WIRE GUIDE]. IT WASN'T POSSIBLE TO PASS THE STRICTURE AGAIN, [WHICH IS] WHY A SPYGLASS PROCEDURE IS PLANNED."
Patient problems: Foreign Body In Patient; No Consequences Or Impact To Patient
MAUDE report 5702985 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page