The Web G22776Recalls & reported problems
Cook Incorporated · The Web Extraction Basket · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00827002227767), received between Feb 19, 2016 and Feb 20, 2025 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Difficult to Remove1
- Material Deformation1
- Material Split, Cut or Torn1
- Mechanical Jam1
Patient problems
- No Consequences Or Impact To Patient3
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Patient Involvement1
- Obstruction/Occlusion1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryFeb 20, 2025Difficult to Remove; Difficult to Remove
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY, ERCP, THE PHYSICIAN USED A COOK WEB EXTRACTION BASKET. IT WAS REPORTED THAT THE BASKET BECAME IMPACTED IN THE BATTERY AMPOULE [DUCT/PAPILLA]. ALL POSSIBLE MANEUVERS WERE PERFORMED TO REMOVE IT BUT IT WAS NOT POSSIBLE, THEREFORE THE PATIENT WAS TAKEN TO SURGERY. THE BASKET WAS RECOVERED BY SURGERY (EXPLORATORY LAPAROTOMY). ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 21425474 ↗InjuryAug 23, 2019Fracture; Mechanical Jam; Fracture; Mechanical Jam
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED TWO (2) COOK THE WEB EXTRACTION BASKETS. THE ERCP WITH WIDE PAPILLOTOMY WAS PERFORMED, SHOWING THE OUTFLOW OF PURULENT MATERIAL. CHOLANGIOPANCREATOGRAPHY DEMONSTRATED COMPACTED CALCULUS IN THE DISTAL THIRD OF THE BILE DUCT. THEY PASSED THE EXTRACTOR BASKET, WITH IMPACTION OF THIS (SIC). THE USER ATTEMPTED LITHOTRIPSY WITH LITHOTRIPTER, WITH EXTRACTOR BASKET BREAKING. A NEW BASKET WAS USED AND IT BROKE IN THE BILIARY TRACT. AFTER NUMEROUS ATTEMPTS TO WITHDRAW, WE DECIDED TO SUSPEND THE PROCEDURE AND INDICATE EMERGENCY LAPAROTOMY. THE BROKEN BASKET REMAINED INSIDE THE PATIENT¿S BODY. THE PATIENT WAS TAKEN TO SURGERY FOR AN EMERGENCY LAPAROTOMY TO HAVE THE BASKET REMOVED. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: Foreign Body In Patient; Obstruction/Occlusion
MAUDE report 8928096 ↗MalfunctionJul 26, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
DURING A STONE REMOVAL PROCEDURE, THE PHYSICIAN USED A COOK WEB II MEMORY EXTRACTION BASKET. AFTER SEVERAL PASSES, THE BASKET BECAME DIFFICULT TO CLOSE. THE NURSE USED FORCE TO CLOSE THE BASKET, [AND] WHERE [SIC] ONE OF THE WIRES APPEARED TO COME THOUGH THE CATHETER [SHEATH]. NO CONTRAST COULD BE USED AS HOLE PRESENTED IN LUMEN OF THE DEVICE CATHETER [DRIVE WIRE COMES THROUGH SHEATH]. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8833724 ↗MalfunctionJun 4, 2019Fracture; Fracture
AFTER A LITHOTRIPSY PROCEDURE, THE PHYSICIAN TESTED ONE (1) PRIOR TO USE COOK WEB EXTRACTION BASKET. THE WIRE OF THE BASKET TORE CLOSE TO HANDLE WHILE DOING AN EMERGENCY LITHOTRIPSY [SEE RELATED EMDR 1037905-2019-00304]. THE STONE CAME OUT OF THE BASKET. AFTER THE PROCEDURE, THE USERS TRIED TO REPLICATE THE MALFUNCTION ON A DEVICE OF THE SAME LOT NUMBER OUTSIDE THE PATIENT WITH THE SAME RESULT. WHEN COOK SALES REPRESENTATIVE WAS THERE TO COLLECT THE DEVICES FOR RETURN, A FURTHER CHECK (OUTSIDE PATIENT) WAS TRIED WITH ANOTHER BASKET WITH SAME LOT NUMBER. THE WIRES TORE AS WELL [ALSO RELATED EMDR 1037905-2019-00304]. THEY CHECKED WITH SAME PRODUCT OF ANOTHER LOT NUMBER, BUT AS WELL THE WIRES TORE [SUBJECT OF THIS REPORT]. THIS OCCURRED PRIOR TO PATIENT CONTACT; THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Patient Involvement
MAUDE report 8667653 ↗MalfunctionJun 4, 2019Fracture; Fracture
DURING A LITHOTRIPSY PROCEDURE, THE PHYSICIAN USED ONE (1) AND TESTED TWO (2) PRIOR TO USE COOK WEB EXTRACTION BASKETS. THE WIRE OF THE BASKET TORE CLOSE TO HANDLE WHILE DOING AN EMERGENCY LITHOTRIPSY [SUBJECT OF THIS REPORT]. THE STONE CAME OUT OF THE BASKET. AFTER THE PROCEDURE, THE USERS TRIED TO REPLICATE THE MALFUNCTION ON A DEVICE OF THE SAME LOT NUMBER OUTSIDE THE PATIENT WITH THE SAME RESULT. WHEN COOK SALES REPRESENTATIVE WAS THERE TO COLLECT THE DEVICES FOR RETURN, A FURTHER CHECK (OUTSIDE PATIENT) WAS TRIED WITH ANOTHER BASKET WITH SAME LOT NUMBER. THE WIRES TORE AS WELL [ALSO SUBJECT OF THIS REPORT]. THEY CHECKED WITH SAME PRODUCT OF ANOTHER LOT NUMBER, BUT AS WELL THE WIRES TORE [SEE RELATED EMDR 1037905-2019-00305]. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8667618 ↗MalfunctionFeb 19, 2016Material Deformation; Material Deformation
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE ON A PATIENT OF UNKNOWN AGE AND GENDER, THE PHYSICIAN USED A COOK THE WEB EXTRACTION BASKET. THE PLASTIC SHEATH SEPARATED FROM THE WIRES WHILE CRANKING DOWN ON THE SOHENDRA LITHOTRIPTER. THE PHYSICIAN REMOVED [THE DEVICE] WITH NO HARM TO THE PATIENT AND WAS ABLE TO USE ANOTHER OF THE SAME DEVICE TO COMPLETE THE PROCEDURE. THE COOK REPRESENTATIVE PROVIDED THE FOLLOWING CLARIFICATION ON (B)(6) 2016: WHILE CRANKING DOWN ON THE HANDLE OR LITHOTRIPSY CRANK, THE PLASTIC SHEATH SEPARATED I.E. CUT AND DUG INTO THE LITHOTRIPSY WIRE. THE CUSTOMER CUT THE PLASTIC FROM THE HANDLE WHICH WAS NOT WHERE THE PROBLEM WAS. THE PROBLEM WAS THAT THE DISTAL END WHERE THE BASKET WAS TRYING TO CRUSH THE STONE. IT [THE BASKET] HAD CUT INTO THE PLASTIC AT THE DISTAL END.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5447138 ↗
FDA clearance and classification
The FDA regulates product code FFL (dislodger, stone, basket, ureteral, metal) under 21 CFR 876.4680 as a Class II device. 21 CFR 876.4680 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the The Web G22776
Has the The Web G22776 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00827002227767), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the The Web G22776?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Feb 19, 2016 and Feb 20, 2025. In those reports, the most common specific device problems are “Fracture” (3), “Difficult to Remove” (1) and “Material Deformation” (1); the most common patient problems are “Device Embedded In Tissue or Plaque” (1), “Foreign Body In Patient” (1) and “Obstruction/Occlusion” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the The Web G22776?
None of the reports are classified as deaths, 2 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the The Web G22776 FDA approved?
Devices of this type (product code FFL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cook Incorporated.
Who makes the The Web G22776, and is it still sold?
The FDA's device database lists Cook Incorporated as the company responsible for it (brand The Web, model G22776). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other DISLODGER, STONE, BASKET, URETERAL, METAL devices
- Zero Tip M0063901050Boston Scientific Corporation · DISLODGER, STONE, BASKET, URETERAL, METAL286 reports
- NCircle G18778Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL222 reports
- NCircle G46206Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL195 reports
- NCircle G17520Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL127 reports
- NGage G48297Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL96 reports
- NGage G48294Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL92 reports
- NCircle G19110Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL87 reports
- Bard® SkyLite™ Tipless Nitinol Stone Basket 041900C. R. Bard, Inc. · DISLODGER, STONE, BASKET, URETERAL, METAL78 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page