Oasis G25380Recalls & reported problems
Cook Ireland Ltd · Oasis One Action Stent Introduction System, with Pre-Loaded Cotton Leung Biliary Stent · Class 2 · Implant
All models of Oasis (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002253803), received between Nov 13, 2019 and Oct 29, 2025 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Activation, Positioning or Separation Problem1
- Difficult to Advance1
- Failure to Advance1
- Improper or Incorrect Procedure or Method1
- Physical Resistance/Sticking1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 29, 2025Failure to Advance; Physical Resistance/Sticking; Failure to Advance; Physical Resistance/Sticking
DESCRIPTION OF EVENT IMPOSSIBLE TO JETTISON THE PROSTHESIS AS PER CC FORM". UNABLE TO REMOVE THE STENT FROM THE ENDOSCOPE , THEY REMOVED THE PATIENT'S ENDOSCOPE AND REMOVED THE STENT STUCK AND THEN PUT ANOTHER BOSTON SCIENTIFIC BRAND STENT. PATIENT/EVENT INFO - NOTES 1. WHAT WAS THE TARGET LOCATION FOR THE STENT? BILE DUCT 3. PLEASE DESCRIBE THE STORAGE CONDITIONS OF THE DEVICE PRIOR TO USE ESPECIALLY THOSE PERTAINING TO TEMPERATURE AND LIGHT EXPOSURE. NORMAL 4. WHAT IS THE REORDER NUMBER, DIAMETER AND LENGTH OF THE WIRE GUIDE THAT WAS USED WITH THIS DEVICE IN THIS PROCEDURE?* BRAND PIOLAX DIAMETER 0,35¿¿ 5. WAS THE WIRE GUIDE LUBRICATED PRIOR TO USE? YES 6. WAS THE WIRE GUIDE INSPECTED FOR DAMAGE PRIOR TO USE? YES 7. WAS THE DEVICE AT THE CENTRE OF THE COMPLAINT INSPECTED FOR DAMAGE PRIOR TO USE? YES 8. WERE PREVIOUS PROCEDURES I.E. SPHINCTEROTOMY ETC. CARRIED OUT PRIOR TO PLACING THE COMPLAINT DEVICE? SPHINCTEROTOMY 9. IF YES, PLEASE INDICATE THE TYPE OF PROCEDURE PERFORMED. SPHINCTEROTOMY. 10. DID THE PATIENT INVOLVED EXHIBIT ALTERED ANATOMY OR TORTUOUS ANATOMY? NO 11. IF NOT WITH THE DEVICE IN QUESTION, HOW WAS THE PROCEDURE FINISHED?* ANOTHER STENT ( BOSTON SCIENTIFIC BRAND) 1
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23413814 ↗MalfunctionFeb 10, 2020Improper or Incorrect Procedure or Method; Difficult to Advance; Improper or Incorrect Procedure or Method; Difficult to Advance
FINAL MDR BEING SUBMITTED DUE TO COMPLETION OF THE INVESTIGATION ON (B)(6) 2021.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 9685903 ↗InjuryNov 21, 2019Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
COMPLAINT RECEIVED FROM DM VIA E-MAIL ON 05NOV2019--DID 05NOV2019. AS REPORTED TO CUSTOMER RELATIONS: "I RECEIVED A PRODUCT COMPLAINT ON 10/25 REGARDING OUR STENT INTRODUCER IN RELATION TO A BILIARY PROCEDURE. CUSTOMER: THE (B)(6) HOSPITAL, (B)(6). CONTACT: (B)(6), LPN (LEAD ENDOSCOPY TECHNICIAN). GPN: G25380, LOT C1629388. DESCRIPTION: DURING AN ERCP, THE ENDOSCOPIST AND TECHNICIAN USED A STENT INTRODUCER TO ADVANCE A PLASTIC BILIARY STENT INTO THE WORKING CHANNEL OF A DUODENOSCOPE. UNKNOWINGLY, THE STENT INTRODUCER WAS ADVANCED BEYOND THE BIOPSY PORT AND INTO THE SCOPE. BECAUSE THE TEAM WERE UNAWARE OF THIS, NO ATTEMPT TO RETRIEVE THE INTRODUCER WAS MADE. WHEN THE SAME SCOPE WAS USED FOR ANOTHER PROCEDURE, IT WAS DISCOVERED THAT THE STENT INTRODUCER WAS STILL INSIDE THE SCOPE, AND WAS ADVANCED DOWN AND OUT OF THE SCOPE BY A NEW STENT. THE CUSTOMER RECOMMENDS THAT COOK CHANGE THE COLOR OF THE STENT INTRODUCER (IT IS CURRENTLY BLACK) OR NOT INCLUDE IT AT ALL IN THE STENT SET, AS THE CUSTOMER STATES THEY TYPICALLY DO NOT USE IT. AS FAR AS I KNOW, THE PATIENT HAS NOT BEEN AFFECTED BY THIS EVENT." ADDITIONAL INFORMATION PROVIDED ON 05NOV2019. "WHAT ENDOSCOPE WAS INVOLVED? A DUODENOSCO
Patient problems: Foreign Body In Patient
MAUDE report 9358164 ↗InjuryNov 13, 2019Unintended Movement; Unintended Movement
AS REPORTED BY THE INITIAL REPORTER: "I RECEIVED A PRODUCT COMPLAINT ON (B)(6) REGARDING OUR STENT INTRODUCER IN RELATION TO A BILIARY PROCEDURE. ACCORDING TO THE INITIAL REPORTER, "DURING AN ERCP, THE ENDOSCOPIST AND TECHNICIAN USED A STENT INTRODUCER TO ADVANCE A PLASTIC BILIARY STENT INTO THE WORKING CHANNEL OF A DUODENOSCOPE. UNKNOWINGLY, THE STENT INTRODUCER WAS ADVANCED BEYOND THE BIOPSY PORT AND INTO THE SCOPE. BECAUSE THE TEAM WERE UNAWARE OF THIS, NO ATTEMPT TO RETRIEVE THE INTRODUCER WAS MADE. WHEN THE SAME SCOPE WAS USED FOR ANOTHER PROCEDURE, IT WAS DISCOVERED THAT THE STENT INTRODUCER WAS STILL INSIDE THE SCOPE, AND WAS ADVANCED DOWN AND OUT OF THE SCOPE BY A NEW STENT." IT WAS FURTHER REPORTER THAT "THE PATIENT HAS NOT BEEN AFFECTED BY THIS EVENT." ADDITIONAL INFORMATION PROVIDED ON 05NOV2019, A DUODENOSCOPE WAS USED DURING THE PROCEDURE.
Patient problems: No Code Available
MAUDE report 9317284 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.