Skip to content

Oasis G25380Recalls & reported problems

Cook Ireland Ltd · Oasis One Action Stent Introduction System, with Pre-Loaded Cotton Leung Biliary Stent · Class 2 · Implant

All models of Oasis (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002253803), received between Nov 13, 2019 and Oct 29, 2025 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation, Positioning or Separation Problem1
  • Difficult to Advance1
  • Failure to Advance1
  • Improper or Incorrect Procedure or Method1
  • Physical Resistance/Sticking1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1
  • No Code Available1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 29, 2025Failure to Advance; Physical Resistance/Sticking; Failure to Advance; Physical Resistance/Sticking

    DESCRIPTION OF EVENT IMPOSSIBLE TO JETTISON THE PROSTHESIS AS PER CC FORM". UNABLE TO REMOVE THE STENT FROM THE ENDOSCOPE , THEY REMOVED THE PATIENT'S ENDOSCOPE AND REMOVED THE STENT STUCK AND THEN PUT ANOTHER BOSTON SCIENTIFIC BRAND STENT. PATIENT/EVENT INFO - NOTES 1. WHAT WAS THE TARGET LOCATION FOR THE STENT? BILE DUCT 3. PLEASE DESCRIBE THE STORAGE CONDITIONS OF THE DEVICE PRIOR TO USE ESPECIALLY THOSE PERTAINING TO TEMPERATURE AND LIGHT EXPOSURE. NORMAL 4. WHAT IS THE REORDER NUMBER, DIAMETER AND LENGTH OF THE WIRE GUIDE THAT WAS USED WITH THIS DEVICE IN THIS PROCEDURE?* BRAND PIOLAX DIAMETER 0,35¿¿ 5. WAS THE WIRE GUIDE LUBRICATED PRIOR TO USE? YES 6. WAS THE WIRE GUIDE INSPECTED FOR DAMAGE PRIOR TO USE? YES 7. WAS THE DEVICE AT THE CENTRE OF THE COMPLAINT INSPECTED FOR DAMAGE PRIOR TO USE? YES 8. WERE PREVIOUS PROCEDURES I.E. SPHINCTEROTOMY ETC. CARRIED OUT PRIOR TO PLACING THE COMPLAINT DEVICE? SPHINCTEROTOMY 9. IF YES, PLEASE INDICATE THE TYPE OF PROCEDURE PERFORMED. SPHINCTEROTOMY. 10. DID THE PATIENT INVOLVED EXHIBIT ALTERED ANATOMY OR TORTUOUS ANATOMY? NO 11. IF NOT WITH THE DEVICE IN QUESTION, HOW WAS THE PROCEDURE FINISHED?* ANOTHER STENT ( BOSTON SCIENTIFIC BRAND) 1

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23413814 ↗
  • MalfunctionFeb 10, 2020Improper or Incorrect Procedure or Method; Difficult to Advance; Improper or Incorrect Procedure or Method; Difficult to Advance

    FINAL MDR BEING SUBMITTED DUE TO COMPLETION OF THE INVESTIGATION ON (B)(6) 2021.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 9685903 ↗
  • InjuryNov 21, 2019Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem

    COMPLAINT RECEIVED FROM DM VIA E-MAIL ON 05NOV2019--DID 05NOV2019. AS REPORTED TO CUSTOMER RELATIONS: "I RECEIVED A PRODUCT COMPLAINT ON 10/25 REGARDING OUR STENT INTRODUCER IN RELATION TO A BILIARY PROCEDURE. CUSTOMER: THE (B)(6) HOSPITAL, (B)(6). CONTACT: (B)(6), LPN (LEAD ENDOSCOPY TECHNICIAN). GPN: G25380, LOT C1629388. DESCRIPTION: DURING AN ERCP, THE ENDOSCOPIST AND TECHNICIAN USED A STENT INTRODUCER TO ADVANCE A PLASTIC BILIARY STENT INTO THE WORKING CHANNEL OF A DUODENOSCOPE. UNKNOWINGLY, THE STENT INTRODUCER WAS ADVANCED BEYOND THE BIOPSY PORT AND INTO THE SCOPE. BECAUSE THE TEAM WERE UNAWARE OF THIS, NO ATTEMPT TO RETRIEVE THE INTRODUCER WAS MADE. WHEN THE SAME SCOPE WAS USED FOR ANOTHER PROCEDURE, IT WAS DISCOVERED THAT THE STENT INTRODUCER WAS STILL INSIDE THE SCOPE, AND WAS ADVANCED DOWN AND OUT OF THE SCOPE BY A NEW STENT. THE CUSTOMER RECOMMENDS THAT COOK CHANGE THE COLOR OF THE STENT INTRODUCER (IT IS CURRENTLY BLACK) OR NOT INCLUDE IT AT ALL IN THE STENT SET, AS THE CUSTOMER STATES THEY TYPICALLY DO NOT USE IT. AS FAR AS I KNOW, THE PATIENT HAS NOT BEEN AFFECTED BY THIS EVENT." ADDITIONAL INFORMATION PROVIDED ON 05NOV2019. "WHAT ENDOSCOPE WAS INVOLVED? A DUODENOSCO

    Patient problems: Foreign Body In Patient

    MAUDE report 9358164 ↗
  • InjuryNov 13, 2019Unintended Movement; Unintended Movement

    AS REPORTED BY THE INITIAL REPORTER: "I RECEIVED A PRODUCT COMPLAINT ON (B)(6) REGARDING OUR STENT INTRODUCER IN RELATION TO A BILIARY PROCEDURE. ACCORDING TO THE INITIAL REPORTER, "DURING AN ERCP, THE ENDOSCOPIST AND TECHNICIAN USED A STENT INTRODUCER TO ADVANCE A PLASTIC BILIARY STENT INTO THE WORKING CHANNEL OF A DUODENOSCOPE. UNKNOWINGLY, THE STENT INTRODUCER WAS ADVANCED BEYOND THE BIOPSY PORT AND INTO THE SCOPE. BECAUSE THE TEAM WERE UNAWARE OF THIS, NO ATTEMPT TO RETRIEVE THE INTRODUCER WAS MADE. WHEN THE SAME SCOPE WAS USED FOR ANOTHER PROCEDURE, IT WAS DISCOVERED THAT THE STENT INTRODUCER WAS STILL INSIDE THE SCOPE, AND WAS ADVANCED DOWN AND OUT OF THE SCOPE BY A NEW STENT." IT WAS FURTHER REPORTER THAT "THE PATIENT HAS NOT BEEN AFFECTED BY THIS EVENT." ADDITIONAL INFORMATION PROVIDED ON 05NOV2019, A DUODENOSCOPE WAS USED DURING THE PROCEDURE.

    Patient problems: No Code Available

    MAUDE report 9317284 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →

Sources and updates