Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00827002303652), received between Apr 17, 2018 and Apr 15, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture2
  • Device Contaminated During Manufacture or Shipping1
  • Device Operates Differently Than Expected1
  • Material Separation1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient2
  • No Code Available1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 15, 2024Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19112842 ↗
  • MalfunctionMar 31, 2022Material Rupture; Material Twisted/Bent; Material Rupture; Material Twisted/Bent

    AS REPORTED, AN ULTRAXX NEPHROSTOMY BALLOON AND SET RUPTURED AT AN UNKNOWN PRESSURE. THE DEVICE WAS USED FOR ACCESS DURING A PERCUTANEOUS NEPHROLITHOTOMY. THE BALLOON WAS "EXTREMELY DIFFICULT" TO INFLATE AND THE KNOB WAS HARD TO TURN. AFTER THE RUPTURE, THE SHEATH WAS PLACED OVER THE BALLOON AND BENT IN TWO PLACES (IT IS UNKNOWN WHICH DEVICE BENT). A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. NO UNINTENDED SECTION OF THE DEVICE REMAINED INSIDE THE PATIENT. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THE OCCURRENCE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13971384 ↗
  • MalfunctionNov 15, 2019Material Separation; Material Separation

    IT IS REPORTED DURING THE REMOVAL OF A LEFT STAGHORN CALCULUS IN AN ADULT MALE PATIENT USING A ULTRAXX NEPHROSTOMY BALLOON AND SET, WHEN ATTEMPTING TO DILATE A RETRACTOR, THE BALLOON BROKE INSIDE THE PATIENT AND THE FLUID DRAINED OUT. THE ISSUE OCCURRED DURING THE BALLOON INFLATION. THE BALLOON WAS WITHDRAWN AND A SECOND BALLOON WAS USED WITHOUT COMPLICATION TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THIS OCCURRENCE. NO SECTION OF THE DEVICE REMAINED INSIDE THE PATIENT'S BODY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9329633 ↗
  • MalfunctionJun 12, 2019Material Rupture; Material Rupture

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8692144 ↗
  • MalfunctionApr 17, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected

    IT WAS REPORTED AN ULTRAXX NEPHROSTOMY BALLOON SET WAS DILATED IN A PCNL TO ESTABLISH A NEPHROSTOMY TRACT INTO THE KIDNEY FOR STONE MANIPULATION, HOWEVER, THE SHEATH WOULD NOT ADVANCE. A SECOND ATTEMPT WAS MADE AT AN ALTERNATIVE PUNCTURE SITE WHERE THE SHEATH WOULD NOT ADVANCE AGAIN. THE PHYSICIAN STATED THERE WAS SIGNIFICANT SPACE BETWEEN THE OUTER DIAMETER OF THE BALLOON AND THE INNER DIAMETER OF THE SHEATH DESPITE COMPLETE INFLATION OF THE BALLOON. THE CASE WAS ABORTED AND THE PATIENT¿S STONES REMAIN WITHIN THE KIDNEY. THE PHYSICIAN HAS REFERRED THE PATIENT TO ANOTHER FACILITY FOR A PROBABLE PCNL AS FURTHER TREATMENT. TO DATE, NO KNOWN ADVERSE CONSEQUENCES TO THE PATIENT RESULTED FROM THIS INCIDENT.

    Patient problems: No Code Available

    MAUDE report 7438557 ↗