NCompass G32701Recalls & reported problems

Cook Incorporated · NCompass, Nitinol Stone Extractor 4 Wire Basket · Class 2

All models of NCompass (5)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Cook Medical NCompass nitinol stone extractor is a four-wire basket used to grab and remove kidney and ureter stones. The 37 reports mostly describe the basket sticking open or shut, wires breaking, and the metal separating or twisting, often discovered when staff opened the package before a procedure and then finished the case with another basket. There are no deaths and one injury. No recalls name this device. Reports have slowed, from 17 in the earlier 12 months to 6 in the most recent.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
36
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 00827002327016), received between Feb 27, 2017 and Feb 26, 2026 — classified as 1 injury, 35 malfunctions.

The most frequently reported specific device problem is “Difficult to Open or Close” (16 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 7 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 8 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Open or Close16
  • Break14
  • Material Separation8
  • Contamination /Decontamination Problem1
  • Deformation Due to Compressive Stress1
  • Device Operates Differently Than Expected1
  • Fracture1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions22
  • No Consequences Or Impact To Patient8
  • No Known Impact Or Consequence To Patient4
  • No Patient Involvement2

Recent adverse event reports

The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 26, 2026Difficult to Open or Close

    IT WAS REPORTED THE NCOMPASS NITINOL TIPLESS STONE EXTRACTOR¿S BASKET WAS UNABLE TO CLOSE COMPLETELY PRIOR TO USE AN UNSPECIFIED KIDNEY PROCEDURE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-TYPE DEVICE. A CUSTOMER ATTACHED A VIDEO APPEARS TO SHOW THE BASKET ONLY PARTIALLY CLOSING AND FULLY OPENING. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24454836 ↗
  • MalfunctionFeb 13, 2026Material Separation; Difficult to Open or Close

    IT WAS REPORTED THE BASKET OF AN NCOMPASS NITINOL TIPLESS STONE EXTRACTOR WAS UNABLE TO BE OPENED PRIOR TO USE DURING A TRANSURETHRAL LASER LITHOTRIPSY PROCEDURE IN THE URETER/RENAL PELVIS. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-TYPE DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24333217 ↗
  • MalfunctionFeb 13, 2026Difficult to Open or Close

    IT WAS REPORTED THE BASKET OF AN NCOMPASS NITINOL TIPLESS STONE EXTRACTOR WAS UNABLE TO BE OPENED PRIOR TO A URETEROSCOPY LITHOTRIPSY PROCEDURE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-TYPE DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. THE DEVICE DID NOT MAKE PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24332569 ↗
  • MalfunctionDec 22, 2025Difficult to Open or Close; Difficult to Open or Close

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23866986 ↗
  • MalfunctionDec 16, 2025Break; Material Deformation; Break; Material Deformation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23818146 ↗
  • MalfunctionOct 21, 2025Difficult to Open or Close; Difficult to Open or Close

    AS REPORTED, DURING AN UNSPECIFIED PROCEDURE, THE BASKET OF 'NCOMPASS NITINOL TIPLESS STONE EXTRACTOR' COULD NOT BE OPENED. THE ISSUE WAS DISCOVERED AFTER REMOVING THE OUTER PACKAGING; HENCE THE PRODUCT COULD NOT BE USED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING ANOTHER NEW DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23352210 ↗
  • MalfunctionAug 29, 2025Difficult to Open or Close; Material Twisted/Bent; Difficult to Open or Close; Material Twisted/Bent

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22927155 ↗
  • MalfunctionAug 29, 2025Break; Break

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22927040 ↗
  • MalfunctionAug 18, 2025Material Separation; Difficult to Open or Close; Material Separation; Difficult to Open or Close

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22830635 ↗
  • MalfunctionAug 17, 2025Difficult to Open or Close; Difficult to Open or Close

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22824457 ↗
  • MalfunctionAug 15, 2025Deformation Due to Compressive Stress; Difficult to Open or Close; Deformation Due to Compressive Stress; Difficult to Open or Close

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22821732 ↗
  • MalfunctionAug 14, 2025Break; Break

    IT WAS REPORTED THAT DURING A URETERORENOSCOPY OF THE KIDNEY, THE BASKET OF A NCOMPASS NITINOL TIPLESS STONE EXTRACTOR BROKE. A SECOND NCOMPASS NITINOL TIPLESS STONE EXTRACTOR WAS OPENED AND THE BASKET BROKE IN HALF (COVERED UNDER PATIENT IDENTIFIER (B)(6). THE PROCEDURE WAS COMPLETED WITHOUT A BASKET DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. THERE WAS NO UNINTENDED SECTION OF THE DEVICE THAT REMAINED INSIDE THE PATIENT¿S BODY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22808946 ↗
  • MalfunctionAug 14, 2025Break; Break

    ADDITIONAL INFORMATION WAS RECEIVED 14APR2026. AN UNSPECIFIED LASER WAS USED. THERE WERE NO STONE FRAGMENTS LEFT IN THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22808913 ↗
  • MalfunctionAug 14, 2025Break; Difficult to Open or Close; Break; Difficult to Open or Close

    CORRECTION: THE CUSTOMER PROVIDED A PHOTO OF THE NCOMPASS NITINOL TIPLESS STONE EXTRACTOR THAT SHOWED THE BASKET WIRES BROKEN. THE DEVICE WAS RETURNED, AND A PRELIMINARY EVALUATION WAS PERFORMED. A VISUAL EXAMINATION OF THE DEVICE NOTED THE BASKET APPEARED TO BE USED AND ALL 4 BASKET WIRES WERE SEVERED, CHARRED, AND FRAZZLED LOOKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22808687 ↗
  • MalfunctionJul 17, 2025Break; Break

    IT WAS REPORTED TWO NCOMPASS NITINOL TIPLESS STONE EXTRACTORS WERE BROKEN PRIOR TO USE FOR AN UNSPECIFIED PROCEDURE. THE ISSUE WAS FOUND WHEN THE PACKAGE WAS OPENED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT HAS NOT YET BEEN RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22530025 ↗
  • MalfunctionMay 30, 2025Break; Break

    IT WAS REPORTED DURING A URETEROSCOPY (URS) URETER PROCEDURE, THE YELLOW SHEATH AND HANDLE OF AN NCOMPASS NITINOL TIPLESS STONE EXTRACTOR WERE SEPARATED SO THE BASKET COULD NOT BE OPENED AND CLOSED. ANOTHER NEW PRODUCT WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22117630 ↗
  • MalfunctionMay 30, 2025Break; Material Separation; Break; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22117293 ↗
  • MalfunctionMay 30, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THE NCOMPASS NITINOL TIPLESS STONE EXTRACTOR WAS UNABLE TO OPEN AND CLOSE DURING AN URETEROSCOPY PROCEDURE. REPORTEDLY, THE YELLOW SHEATH AND HANDLE PARTS WERE FOUND TO BE SEPARATED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-LIKE DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL INTERVENTION DUE TO THIS OCCURRENCE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22117153 ↗
  • MalfunctionMay 29, 2025Break; Break

    IT WAS REPORTED DURING AN UNKNOWN PROCEDURE, THE YELLOW SHEATH AND THE HANDLE OF AN NGAGE NITINOL STONE EXTRACTOR WERE SEPARATED, SO THE BASKET COULD NOT BE OPENED AND CLOSED. ANOTHER NEW PRODUCT WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22107072 ↗
  • MalfunctionMay 28, 2025Fracture; Fracture

    IT WAS REPORTED, DURING A STONE EXTRACTION PROCEDURE, THAT THE BASKET WIRE OF A NCOMPASS NITINOL TIPLESS STONE EXTRACTOR SNAPPED. ADDITIONAL INFORMATION HAS BEEN REQUESTED REGARDING THE PATIENT AND THE EVENT. AT THE TIME OF THIS REPORT, NO FURTHER INFORMATION HAS BEEN PROVIDED. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22095902 ↗
  • MalfunctionMay 28, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED, DURING PREPARATION OF AN NCOMPASS NITINOL TIPLESS STONE EXTRACTOR, THE BASKET WAS UNABLE TO OPEN AND CLOSE DURING AN UNSPECIFIED PROCEDURE. IT WAS FOUND THAT THE YELLOW SHEATH AND HANDLE PARTS WERE SEPARATED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-LIKE DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL INTERVENTION DUE TO THIS OCCURRENCE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22095111 ↗
  • MalfunctionMay 27, 2025Material Separation; Material Separation

    IT WAS REPORTED DURING PREPARATION OF AN NCOMPASS NITINOL TIPLESS STONE EXTRACTOR THE BASKET WAS UNABLE TO OPEN AND CLOSE DURING PRIOR TO AN UNSPECIFIED PROCEDURE. THE YELLOW SHEATH AND HANDLE PARTS WERE SEPARATED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-LIKE DEVICE. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL INTERVENTION DUE TO THIS OCCURRENCE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22090236 ↗
  • MalfunctionJul 26, 2019Difficult to Open or Close; Difficult to Open or Close

    THERE IS NO NEW PATIENT OR EVENT INFORMATION TO REPORT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8832914 ↗
  • MalfunctionJun 24, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT IS REPORTED THAT A DISTRIBUTOR FOUND FOREIGN MATTER SEALED INSIDE AN COMPASS NITINOL TIPLESS STONE EXTRACTOR PACKAGE. THE PACKAGE WAS NOT OPENED. THE DEVICE DID NOT MAKE CONTACT WITH ANY PATIENT.

    Patient problems: No Patient Involvement

    MAUDE report 8727988 ↗
  • MalfunctionMay 14, 2019Break; Break

    IT IS REPORTED WHILE PREPARING FOR AN UNSPECIFIED PROCEDURE USING A NCOMPASS NITINOL TIPLESS STONE EXTRACTOR, THE USER OPENED THE PACKAGE AND IT WAS IMMEDIATELY NOTED THE DEVICE WAS BROKEN AND THE BASKET WAS MISSING FROM THE END. THIS DEVICE HAD NO PATIENT CONTACT. A SECOND BASKET WAS OPENED, AND USED TO CARRY ON WITH THE INTENDED PROCEDURE. NO ADVERSE EFFECTS TO THE PATIENT HAVE BEEN REPORTED AS A RESULT OF THIS ALLEGED PRODUCT MALFUNCTION. THE CUSTOMER DECLINED TO ANSWER ANY ADDITIONAL QUESTIONS DUE TO THE FACT THAT THE DEVICE MADE NO PATIENT CONTACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8608566 ↗

FDA clearance and classification

The FDA regulates product code FFL (dislodger, stone, basket, ureteral, metal) under 21 CFR 876.4680 as a Class II device. 21 CFR 876.4680 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NCompass G32701

Has the NCompass G32701 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00827002327016), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NCompass G32701?

The FDA's MAUDE database lists 36 adverse event reports for this device, received between Feb 27, 2017 and Feb 26, 2026. In those reports, the most common specific device problems are “Difficult to Open or Close” (16), “Break” (14) and “Material Separation” (8). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NCompass G32701?

None of the reports are classified as deaths, 1 as an injury and 35 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NCompass G32701 FDA approved?

Devices of this type (product code FFL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cook Incorporated.

Who makes the NCompass G32701, and is it still sold?

The FDA's device database lists Cook Incorporated as the company responsible for it (brand NCompass, model G32701). Its record shows it as in commercial distribution.