NCompass G32702Recalls & reported problems
Cook Incorporated · NCompass, Nitinol Stone Extractor 3 Wire Basket · Class 2
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00827002327023), received between Sep 27, 2017 and Aug 4, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Difficult to Open or Close” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Open or Close5
- Break3
- Component Missing1
- Detachment Of Device Component1
- Device Damaged Prior to Use1
- Fracture1
- Kinked1
- Loss of or Failure to Bond1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 4, 2026Difficult to Open or Close
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50132575 ↗MalfunctionApr 1, 2026Break; Fracture
AS REPORTED, DURING URETEROSCOPIC STONE REMOVAL PROCEDURE, THE TIP OF THE 'NCOMPASS NITINOL STONE EXTRACTOR' FELL OFF. THE NCOMPASS WAS OPENED AND TESTED PRIOR TO USE WITH NO ISSUES. IT WAS USED TO REMOVE THREE STONE FRAGMENTS. WHEN THE USER ATTEMPTED TO REMOVE THE FOURTH FRAGMENT, THE TIP FELL OFF INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY RETRIEVING THE FRAGMENT AND USING ANOTHER NCOMPASS. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24754946 ↗MalfunctionFeb 6, 2026Loss of or Failure to Bond; Difficult to Open or Close
IT WAS REPORTED, TWO NCOMPASS NITINOL STONE EXTRACTORS WOULD NOT OPEN AND CLOSE DURING A LAPAROSCOPIC CHOLECYSTECTOMY. THIS REPORT WILL BE FOR DEVICE # 2. REFERENCE PATIENT IDENTIFIER (B)(6) FOR DEVICE # 1. A THIRD SAME TYPE DEVICE WAS USED TO COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24277902 ↗MalfunctionFeb 6, 2026Break; Difficult to Open or Close
IT WAS REPORTED, TWO NCOMPASS NITINOL STONE EXTRACTORS WOULD NOT OPEN AND CLOSE DURING A LAPAROSCOPIC CHOLECYSTECTOMY. THIS REPORT WILL BE FOR DEVICE # 1. REFERENCE PATIENT IDENTIFIER (B)(6) FOR DEVICE # 2. A THIRD SAME TYPE DEVICE WAS USED TO COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24277834 ↗MalfunctionSep 9, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED, THE NCOMPASS NITINOL STONE EXTRACTOR WAS STUCK IN THE OPEN POSITION AND WOULD NOT CLOSE, PRIOR TO AN UNSPECIFIED PROCEDURE. THE DEVICE WAS INSPECTED AND THERE WERE NO OBVIOUS DEFECTS. THE ISSUE WAS FOUND WHILE TESTING THE FUNCTION OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER SAME LIKE DEVICE. THE DEVICE DID NOT MAKE PATIENT CONTACT. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23003712 ↗MalfunctionDec 18, 2018Material Separation; Material Separation
NO NEW INFORMATION HAS BEEN PROVIDED DUE TO PRIVACY RESTRICTIONS OF THE FACILITY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8173188 ↗MalfunctionJan 8, 2018Kinked; Difficult to Open or Close; Kinked; Difficult to Open or Close
IT WAS REPORTED THE NCOMPASS NITINOL STONE EXTRACTOR WIRE KINKED APPROXIMATELY 3CM FROM THE BASKET HANDLE, PREVENTING OPERATION OF THE BASKET. NO UNINTENDED PORTION OF THE DEVICE REMAINED INSIDE THE PATIENT¿S BODY. NO ADDITIONAL PROCEDURES WERE REQUIRED DUE TO THIS OCCURRENCE. NO ADVERSE EFFECTS OR CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7171670 ↗MalfunctionDec 19, 2017Break; Detachment Of Device Component
THE COMPLAINANT REPORTED THAT THE BASKET WAS OPENED AND CLOSED BEFORE USING DOWN THE URETEROSCOPE; THE BASKET SUBSEQUENTLY "BROKE OFF." THE BASKET COULD NOT BE USED AND ANOTHER BASKET WAS USED FOR CASE. THERE WERE NO REPORTED ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THIS PRODUCT PROBLEM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7128495 ↗MalfunctionSep 27, 2017Device Damaged Prior to Use; Component Missing; Device Damaged Prior to Use; Component Missing
DR.(B)(6) AND HIS ASSISTANT DISCOVERED BEFORE THE OPERATION THAT SHEATH NEAR THE HANDLE WAS TAKEN OFF FROM THE NCOMPASS NITINOL STONE EXTRACTOR AND IT WAS SQUASHED. THE DEVICE WAS NOT USED. THE DEVICE DID NOT MAKE CONTACT WITH THE PATIENT. THERE WERE NO ADVERSE CONSEQUENCES OR EFFECTS TO THE PATIENT AS A RESULT OF THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6896695 ↗
Other DISLODGER, STONE, BASKET, URETERAL, METAL devices
- Zero Tip M0063901050Boston Scientific Corporation · DISLODGER, STONE, BASKET, URETERAL, METAL287 reports
- NCircle G18778Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL226 reports
- NCircle G46206Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL197 reports
- NCircle G17520Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL129 reports
- NGage G48297Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL96 reports
- NGage G48294Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL92 reports
- NCircle G19110Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL88 reports
- Bard® SkyLite™ Tipless Nitinol Stone Basket 041900C. R. Bard, Inc. · DISLODGER, STONE, BASKET, URETERAL, METAL78 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.