Skip to content

Atlas Wire G32824Recalls & reported problems

Cook Incorporated · Atlas Wire Stone Extractor 4 Wire Flat-Wire Basket · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002328242), received between Jul 3, 2019 and Oct 12, 2023 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Break1
  • Difficult to Open or Close1
  • Material Frayed1
  • Material Separation1
  • Material Split, Cut or Torn1

Patient problems

  • No Consequences Or Impact To Patient3
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 12, 2023Material Split, Cut or Torn; Material Split, Cut or Torn

    AS REPORTED, DURING A UNILATERAL RIGID URETERORENAL LITHOTRIPSY PROCEDURE, THE TIP OF THE EXTRACTOR, LOCATED BEFORE THE BASKET, WAS CRACKED. A NEW SAME TYPE DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO SECTION OF THE DEVICE REMAINED INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. THERE WERE NO ADVERSE EFFECTS ON THE PATIENT DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17928203 ↗
  • MalfunctionOct 22, 2020Material Frayed; Material Frayed

    AS REPORTED, PRIOR TO USE IN A RIGID URETERORENOLITHOTRIPSY PROCEDURE, THE EXTRACTOR WIRE OF A ATLAS-WIRE STONE EXTRACTOR FRAYED WHEN TESTING THE DEVICE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10718736 ↗
  • MalfunctionSep 29, 2020Break; Break

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10599688 ↗
  • MalfunctionJul 3, 2019Material Separation; Difficult to Open or Close; Material Separation; Difficult to Open or Close

    IT WAS INITIALLY REPORTED THAT DURING A FLEXIBLE URETEROSCOPY WITH LASER LITHOTRIPSY THROUGH A FLEXOR URETERAL ACCESS SHEATH, MULTIPLE COOK EXTRACTOR DEVICES MALFUNCTIONED AND WOULD NOT OPEN, INCLUDING ONE DEVICE SEPARATING (REPORTED UNDER PATIENT IDENTIFIERS (B)(6)). ULTIMATELY, THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER COOK EXTRACTOR DEVICE. FIVE (5) DEVICES WERE RETURNED TO THE MANUFACTURER. UPON COMPLETION OF THE DEVICE FAILURE ANALYSIS FORMS ON 02JUL2019 THE FOLLOWING INFORMATION HAS BEEN PROVIDED: TWO (2) NCIRCLE TIPLESS STONE EXTRACTOR DEVICES WERE RETURNED. THE FIRST DEVICE, (REPORTED UNDER PATIENT IDENTIFIER (B)(6)) WAS NOTED TO MALFUNCTION IN TERMS OF NOT OPENING AND CLOSING PROPERLY. THE SECOND DEVICE (REPORTED UNDER PATIENT IDENTIFIER 270371) WAS NOTED TO BE "SEVERED" OR SEPARATED. THREE (3) ATLAS-WIRE STONE EXTRACTOR DEVICES WERE RETURNED. DURING THE INVESTIGATION, IT WAS DISCOVERED THAT ALL THREE OF THESE DEVICES WERE NOTED TO MALFUNCTION IN TERMS OF NOT OPENING AND CLOSING PROPERLY. IT WAS INITIALLY REPORTED THAT TWO OF THE ATLAS-WIRE STONE EXTRACTOR DEVICES MALFUNCTIONED IN TERMS OF NOT OPENING AND CLOSING PROPERLY AND WERE MISTAKENLY REPORTED (REPORTED

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8759749 ↗

FDA clearance and classification

The FDA regulates product code FFL (dislodger, stone, basket, ureteral, metal) under 21 CFR 876.4680 as a Class II device. 21 CFR 876.4680 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Atlas Wire G32824

Has the Atlas Wire G32824 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00827002328242), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Atlas Wire G32824?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jul 3, 2019 and Oct 12, 2023. In those reports, the most common specific device problems are “Break” (1), “Difficult to Open or Close” (1) and “Material Frayed” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Atlas Wire G32824?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Atlas Wire G32824 FDA approved?

Devices of this type (product code FFL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cook Incorporated.

Who makes the Atlas Wire G32824, and is it still sold?

The FDA's device database lists Cook Incorporated as the company responsible for it (brand Atlas Wire, model G32824). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other DISLODGER, STONE, BASKET, URETERAL, METAL devices

All DISLODGER, STONE, BASKET, URETERAL, METAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page