NCompass G46027Recalls & reported problems

Cook Incorporated · NCompass, Nitinol Stone Extractor 4 Wire Basket · Class 2

All models of NCompass (5)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00827002460270), received between Mar 28, 2018 and Nov 3, 2025 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Difficult to Open or Close4
  • Device Operates Differently Than Expected2
  • Difficult or Delayed Positioning1
  • Material Separation1

Patient problems

  • No Consequences Or Impact To Patient8
  • No Patient Involvement3
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 3, 2025Break; Break

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23458017 ↗
  • MalfunctionNov 20, 2019Break; Break

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED SINCE THE LAST REPORT WAS SENT ON (B)(6) 2019.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9348279 ↗
  • MalfunctionNov 7, 2019Difficult to Open or Close; Difficult to Open or Close

    THERE IS NO NEW PATIENT OR EVENT INFORMATION TO REPORT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9289484 ↗
  • MalfunctionOct 9, 2019Difficult to Open or Close; Difficult to Open or Close

    NO NEW PATIENT OR EVENT INFORMATION SINCE THE LAST REPORT WAS SUBMITTED ON 09OCT2019.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9172805 ↗
  • MalfunctionOct 1, 2019Break; Break

    ADDITIONAL INFORMATION WAS RECEIVED ON 01OCT2019: A NEW SAME TYPE DEVICE WAS OPENED AND USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9140094 ↗
  • MalfunctionSep 30, 2019Difficult to Open or Close; Difficult to Open or Close

    NO ADDITIONAL PATIENT OR EVENT INFORMATION HAS BEEN RECEIVED SINCE THE LAST REPORT WAS SUBMITTED ON 30SEP2019.

    Patient problems: No Patient Involvement

    MAUDE report 9135656 ↗
  • MalfunctionMay 1, 2019Difficult to Open or Close; Difficult to Open or Close

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8570277 ↗
  • MalfunctionFeb 22, 2019Break; Break

    IT WAS REPORTED DURING A URETEROSCOPY WITH STONE REMOVAL PROCEDURE, TWO NCOMPASS NITINOL TIPLESS STONE EXTRACTOR BASKETS BROKE. DR. SAID THE DISTAL END WITH THE ¿STICK¿ IS MORE FLIMSY THAN ANOTHER MANUFACTURER'S ONE AND THAT HE IS MORE CAREFUL WHILE USING THIS PRODUCT. NO UNINTENDED PART OF THE DEVICE WAS LEFT INSIDE THE PATIENT'S BODY. NO ADDITIONAL PROCEDURES WERE REQUIRED. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THESE OCCURRENCES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8363299 ↗
  • MalfunctionFeb 20, 2019Break; Material Separation; Break; Material Separation

    IT WAS INITIALLY REPORTED THE DEVICES BROKE/BENT AT THE ORANGE JACKET THAT CONNECTS THE BASKET WIRE TO THE HANDLE DURING USE. THIS HAS CAUSED THE DEVICE TO FAIL AND THE BASKET TO NO LONGER OPEN. ADDITIONAL INFORMATION WAS PROVIDED ON 18-FEB-2019. THIS IS ONE COMPLAINT FOR A QUANTITY OF FOUR (4) NCOMPASS NITINOL TIPLESS STONE EXTRACTORS FROM LOT #9342377. DURING THE URETEROSCOPY W/LASER LITHOTRIPSY PROCEDURE WHILE USING A FLEXIBLE URETEROSCOPE AND A WORKING SHEATH APPROXIMATELY 3.6 FRENCH, THE ORANGE TUBING SHEATH THAT IS BEYOND THE HANDLE BROKE/DISCONNECTED IN ALL INSTANCES. THE DEVICE WAS TESTED PRIOR TO USE. THERE WERE NO ISSUES WITH PATIENT ANATOMY THAT KEPT THE BASKET FROM OPENING/CLOSING. THE DOCTOR USED A 2.2 NCIRCLE BASKET TO FINISH GRABBING THE STONE FRAGMENTS. FURTHER CLARIFICATION WAS RECEIVED ON 20-FEB-2019. NO UNINTENDED PART OF THE DEVICE WAS LEFT INSIDE THE PATIENT¿S BODY. NO ADDITIONAL PROCEDURES WERE REQUIRED. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8355433 ↗
  • MalfunctionMay 9, 2018Difficult or Delayed Positioning; Difficult or Delayed Positioning

    IT WAS REPORTED THAT THE USER OPENED THE PACKAGE AND CHECKED THE INTEGRITY OF THE NCOMPASS NITINOL TIPLESS STONE EXTRACTOR AND FOUND THAT THE BASKET CANNOT BE RELEASED. ANOTHER EXTRACTOR WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT AS A RESULT OF THIS REPORTED ISSUE. THE DEVICE DID NOT MAKE CONTACT WITH THE PATIENT.

    Patient problems: No Patient Involvement

    MAUDE report 7500380 ↗
  • MalfunctionMay 1, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected

    IT WAS REPORTED THAT THE USER CHECKED THE INTEGRITY OF THE DEVICE AND FOUND THAT THE HANDLE DIDN'T TRIGGER THE BASKET. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT AS A RESULT OF THIS REPORTED ISSUE. THE DEVICE DID NOT MAKE CONTACT WITH THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7476210 ↗
  • MalfunctionMar 28, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected

    IT WAS REPORTED DURING PREOPERATIVE TESTING, THE BASKET ON A NCOMPASS NITINOL TIPLESS STONE EXTRACTOR WAS UNABLE TO CLOSE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT AS A RESULT OF THIS REPORTED ISSUE. THE DEVICE DID NOT MAKE CONTACT WITH THE PATIENT.

    Patient problems: No Patient Involvement

    MAUDE report 7378544 ↗

FDA clearance and classification

The FDA regulates product code FFL (dislodger, stone, basket, ureteral, metal) under 21 CFR 876.4680 as a Class II device. 21 CFR 876.4680 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NCompass G46027

Has the NCompass G46027 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00827002460270), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NCompass G46027?

The FDA's MAUDE database lists 12 adverse event reports for this device, received between Mar 28, 2018 and Nov 3, 2025. In those reports, the most common specific device problems are “Break” (5), “Difficult to Open or Close” (4) and “Device Operates Differently Than Expected” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NCompass G46027?

None of the reports are classified as deaths, 0 as injuries and 12 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NCompass G46027 FDA approved?

Devices of this type (product code FFL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cook Incorporated.

Who makes the NCompass G46027, and is it still sold?

The FDA's device database lists Cook Incorporated as the company responsible for it (brand NCompass, model G46027). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other DISLODGER, STONE, BASKET, URETERAL, METAL devices

All DISLODGER, STONE, BASKET, URETERAL, METAL devices →