Skip to content

Innomed, Inc. 6890Recalls & reported problems

Innomed, Inc. · Namba Hip Slide for 22-40 mm Heads · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840277103925), received between Mar 31, 2016 and Feb 15, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 15, 2024Break; Break

    WHILE SURGEON WAS USING SKID, THE TIP BROKE AND ALL PARTICLES REMOVED. SKID REMOVED AND REPLACED. BROKEN SKID SENT CPD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18716579 ↗
  • MalfunctionMar 31, 2016Break; Break

    THE FOLLOWING DESCRIPTION WAS RECEIVED BY THE CDRH ON 02/17/2016, (B)(4). IMMEDIATELY FOLLOWING A RIGHT TOTAL HIP ARTHROPLASTY, POST-OPERATIVE RADIOGRAPHY VIEWED BY THE SURGEON WAS NOTED TO BE ABNORMAL. THE PATIENT WAS RETURNED BACK TO SURGERY AND THE SURGEON EXPLORED THE SIGHT FINDING A FOREIGN OBJECT WITHIN THE HIP. THE FOREIGN OBJECT WAS A CORNER OF THE NAMBA HIP SLIDE THAT HAD BROKEN OFF. APPROXIMATE LENGTH- 3.5 CM. APPROXIMATE DEPTH- 1.0 CM. DID REQUIRE REPLACEMENT OF NEW HARDWARE. RESPONSE: THE REPORT WAS RECEIVED ON 2/22/2016. THE INITIAL REPORTER WAS CONTACTED AND A REQUEST WAS MADE TO HAVE THE DEVICE RETURNED FOR AN EVALUATION. THE INITIAL REPORTER RELAYED THAT THE DEVICE COULD NOT BE RETURNED FOR EVALUATION PURPOSES. IT WAS THEN ASKED OF THE INITIAL REPORTER IF PICTURES OF THE BROKEN ITEM COULD BE FORWARDED IN LIEU OF THE PHYSICAL ITEM. THE INITIAL REPORTER RESPONDED THAT THE PICTURES WOULD BE FORWARDED. A FOLLOW UP EMAIL WAS SENT TO THE INITIAL REPORTER ON 2/29/2016. IT WAS AGAIN REQUESTED TO FORWARD PHOTOGRAPHS OF THE BROKEN ITEM, AND IT WAS ALSO EXPLAINED THAT IT WOULD BE DIFFICULT TO PERFORM A SUITABLE EVALUATION WITHOUT EVIDENCE. ON 3/8/2016, THE INITIAL REPORTER WAS

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5539717 ↗

Other Retractor devices

All Retractor devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page