Skip to content

JuggerKnot Soft Anchor SP99529BRecalls & reported problems

Riverpoint Medical, LLC · Soft Anchor · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00840277402431), received between Apr 10, 2024 and Apr 10, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 3 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deformation Due to Compressive Stress3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 10, 2024Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    ACCORDING TO THE REPORTER: "ALL-SUTURE ANCHOR INDICATED FOR RCR REPAIR. ON (B)(6) 2024 DR. (B)(6) , USED THE PRODUCT ABOVE, AND IT WAS UNSATISFACTORY. AS DR. (B)(6) WAS HOLDING THE ANCHOR THE HUMERAL HEAD AND HIS FELLOW WAS APPLYING FORCE WITH A MALLET, THE ANCHOR INSERTER BENT AND BROKE ON TWO OCCASIONS. THE BEND WAS SIGNIFICANT LEFT THE PRODUCT DAMAGED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19084312 ↗
  • MalfunctionApr 10, 2024Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    ACCORDING TO THE REPORTER: "ALL-SUTURE ANCHOR INDICATED FOR RCR REPAIR. ON (B)(6)2024 DR. (B)(6) USED THE PRODUCT ABOVE, AND IT WAS UNSATISFACTORY. AS DR.(B)(6) WAS HOLDING THE ANCHOR THE HUMERAL HEAD AND HIS FELLOW WAS APPLYING FORCE WITH A MALLET, THE ANCHOR INSERTER BENT AND BROKE ON TWO OCCASIONS. THE BEND WAS SIGNIFICANT LEFT THE PRODUCER DAMAGED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19083166 ↗
  • MalfunctionApr 10, 2024Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    ACCORDING TO THE REPORTER, "ATTACHED IS THE EVENT INFORMATION FROM THE 3 BROKEN JUGGERNAUTS AT (B)(6) HOSPITAL. I WILL ALSO ATTACH PICTURES OF THE BROKEN PRODUCTS IF THAT HELPS. WE HAVE TRIED 6X WITH THIS PRODUCT, ALL ANCHORS HAD THE SAME LOT # 23081820, 3 WERE SUCCESSFUL AND 3 FAILED. "

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19081910 ↗

Other Fastener, Fixation, Nondegradable, Soft Tissue devices

All Fastener, Fixation, Nondegradable, Soft Tissue devices →

Sources and updates