Skip to content

Iconix 3911-714-522Recalls & reported problems

Riverpoint Medical, LLC · Iconix Knotless, 1.4mm Anchor with XBraid™ S (Blue/Black/White) · Class 2 · Implant

All models of Iconix (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00840277404756), received between Apr 1, 2025 and Mar 18, 2026 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 3 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 3 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Integrity Problem4
  • Patient Device Interaction Problem4
  • Fracture3
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2026Material Integrity Problem

    ACCORDING TO THE REPORTER, "SURGEON WAS USING MALLET TO INSERT ICONIX KNOTLESS 1.4MM INTO THE ACETABULUM, ON INSERTION OF THE ANCHOR THE TIP OF THE INSERTED SHEATH SNAPPED OFF AND STAYED IN THE ACETABULUM, DUE TO THE NATURE OF WHERE IT WAS THE SURGEON DECIDED AGAINST REMOVING IT, ANOTHER ANCHOR WAS USED IN A DIFFERENT LOCATION WITH NIL ISSUES."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24630015 ↗
  • MalfunctionJan 29, 2026Patient Device Interaction Problem

    ACCORDING TO THE REPORTER:"DR. (B)(6) HAD EXPERIENCED SOFT BONE SETTINGS WHERE IMPLANTS DID NOT SET. SALESFORCE CASE NUMBER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24215447 ↗
  • MalfunctionJan 6, 2026Patient Device Interaction Problem

    ACCORDING TO THE REPORTER:"THE ITEMS WERE WASTED BECAUSE THE IMPLANTS PULLED OUT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23981830 ↗
  • MalfunctionJan 5, 2026Material Twisted/Bent

    ACCORDING TO THE REPORTER: "THE ANCHOR BENT ON INSERTION & PULLED OUT".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23958784 ↗
  • MalfunctionDec 30, 2025Fracture; Fracture

    ACCORDING TO THE REPORTER: "THE SURGERY WAS COMPLETED SUCCESSFULLY. THIS PATIENT CAME BACK IN FOR POST-OP X-RAYS AND THERE WAS A SMALL PIECE OF METAL THAT WAS DISCOVERED IN THEIR SHOULDER. THE PATIENT AND DR. HATZIDAKIS DECIDED TO SCHEDULE THE PATIENT FOR A SURGERY ON (B)(6) 2025. DURING THAT SURGERY ON (B)(6) 2025 IT WAS DISCOVERED THAT THIS PIECE OF METAL WAS A PART OF THE 1.4MM ICONIX KNOTLESS INSERTER. DR. HATZIDAKIS WAS ABLE TO SUCCESSFULLY REMOVE THIS PIECE OF THE 1.4MM ICONIX KNOTLESS INSERTER FROM THE PATIENTS SHOULDER DURING THAT SURGERY.."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23924111 ↗
  • MalfunctionDec 22, 2025Fracture; Fracture

    ACCORDING TO THE REPORTER: "WE WERE INFORMED BY A SURGEON TODAY THAT UPON DOING POST-OP X-RAYS IN HIS OFFICE, HE DISCOVERED A SMALL PIECE OF WHAT HE BELIEVED TO BE OF A STRYKER KNOTLESS ANCHOR THAT WAS IN THE SHOULDER. HE STATED THIS WAS HIS FIRST TIME USING THE STRYKER KNOTLESS. WE HAVE ALREADY CONTACTED OUR LOCAL STRYKER SALE REP AND LET THEM KNOW OF THE ISSUE. HE DID NOT REPORT KNOWING OF ANY OTHER ISSUES WITH THIS ANCHOR."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23866171 ↗
  • MalfunctionDec 16, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    ACCORDING TO THE REPORTER: "ICONIX KNOTLESS ANCHOR IMPLANTED. SLIDING MECHANISM OF THE ANCHOR FAILED AND WAS STUCK. HAD TO TIE DOWN KNOTS LIKE A KNOTTED ANCHOR INSTEAD OF SLIDING INTO A KNOTLESS ANCHOR. NOTICED AFTER IMPLANTATION, WHEN TRYING TO PASS REPAIR STRAND."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23823791 ↗
  • MalfunctionOct 22, 2025Material Integrity Problem; Material Integrity Problem

    ACCORDING TO THE REPORTER, "ANCHOR PULLED OUT, SOFT BONE, TOUGH ANGLE. ITEM WASTED ON BILL ONLY."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23360982 ↗
  • MalfunctionOct 22, 2025Material Integrity Problem; Material Integrity Problem

    ACCORDING TO THE REPORTER, "YES, THIS ANCHOR DID NOT PERFORM AS INTENDED. IT PULLED OUT OF THE BONE WITHOUT DEPLOYING. SO, IT DID NOT GET IMPLANTED INTO THE PATIENT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23359265 ↗
  • MalfunctionOct 22, 2025Material Integrity Problem; Material Integrity Problem

    ACCORDING TO THE REPORTER, "ANCHOR FAILED. CUSTOMER WAS UNHAPPY."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23358385 ↗
  • MalfunctionMay 2, 2025Fracture; Fracture

    ACCORDING TO THE REPORTER: "ICONIX INSERTER BROKE OFF IN BONE WHEN PLACING ANCHOR."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21952116 ↗
  • MalfunctionApr 1, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    ACCORDING TO THE REPORTER: "AFTER CUTTING REPAIR STRANDS. ANCHOR LOOSENED AND THEN SUTURES WERE CUT OUT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21747195 ↗

Other Fastener, Fixation, Nondegradable, Soft Tissue devices

All Fastener, Fixation, Nondegradable, Soft Tissue devices →

Sources and updates