Anastoclip AC Closure System 4000-07Recalls & reported problems
Lemaitre Vascular, Inc. · Anastoclip AC Closure System, Large (L) · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00840663102372), received between Jan 22, 2016 and Mar 27, 2019 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (2 reports).
7 other recalls cover devices cleared under the same FDA submission (K962043); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Mechanical Jam2
- Migration or Expulsion of Device2
Patient problems
- Cerebrospinal Fluid Leakage1
- Failure to Anastomose1
- Fistula1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 27, 2019Mechanical Jam; Mechanical Jam
ONE LARGE ANASTOCLIP VCS APPLIERS FAILED TO CLOSE THE VASCULAR TISSUE PROPERLY DURING ANASTOMOSIS. THIS IS REPORT 1 OF 2.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8458459 ↗MalfunctionMar 27, 2019Mechanical Jam; Mechanical Jam
ONE LARGE ANASTOCLIP VCS APPLIERS FAILED TO CLOSE THE VASCULAR TISSUE PROPERLY DURING ANASTOMOSIS. THIS IS REPORT 2 OF 2.
Patient problems: Failure to Anastomose
MAUDE report 8458458 ↗InjuryJan 22, 2016Migration or Expulsion of Device; Migration or Expulsion of Device
SPINE DURAL RUPTURE DURING LAMINECTOMY. 10MM INCISION CLIPPED WITH ANSTOCLIP VCS SIZE L. CLIPS WERE REPORTED TO HAVE BEEN APPLIED NO MORE THAN 0.8-1.0 MM BETWEEN EACH OTHER. ADDITIONAL HEMOSTATIC AGENT ON TOP. NO FILTRATION AFTER VALSAVA TECHNIQUE APPLIED. THE PROCEDURE NEEDED REVISION SURGERY 8 DAYS POST SURGERY BECAUSE OF DURAL FISTULA. DURING REVISION SURGERY DR. (B)(6) FOUND: CLIPS WERE ATTACHED JUST TO ONE SIDE OF TISSUE; CLIPS WERE OUT OF PLACE; LEAKAGE THROUGH THE SPACE BETWEEN CLIPS.
Patient problems: Fistula
MAUDE report 5385931 ↗InjuryJan 22, 2016Migration or Expulsion of Device; Migration or Expulsion of Device
SPINE DURAL RUPTURE DURING LAMINECTOMY. 15MM INCISION CLIPPED WITH ANASTOCLIP VCS SIZE LARGE. CLIPS WERE REPORTED TO HAVE BEEN APPLIED NO MORE THAN 0.8-1.0 MM BETWEEN EACH OTHER. ADDITIONAL HEMOSTATIC AGENT ON TOP. NO FILTRATION AFTER VALSAVA TECHNIQUE APPLIED. THE PROCEDURE NEEDED REVISION SURGERY 24HRS POST OP. BECAUSE OF CSF LEAKAGE. IMAGING SHOWED INFILTRATION INTO THE PSOAS AND SURGICAL INCISION. DURING REVISION SURGERY DR. (B)(6) FOUND THAT: 1. CLIPS WERE OUT OF PLACE. 2. CLIPS WERE ATTACHED JUST TO ONE SIDE OF TISSUE. 3. CRACKED BORDERS OF THE TISSUE.
Patient problems: Cerebrospinal Fluid Leakage
MAUDE report 5385923 ↗
FDA clearance and classification
Cleared through 510(k) K962043 on Sep 23, 1996 (decision: Substantially Equivalent). Submitted by United States Surgical, A Division of Tyco Healthc as “Auto Suture Modified Vcs Clip Applier”. The FDA review took 118 days. FDA record ↗
The FDA regulates product code FZP (clip, implantable) under 21 CFR 878.4300 as a Class II device. 21 CFR 878.4300 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K962043. One submission often covers many models, so check the product and lot details in each record.
- Z-2283-2016 · Jun 22, 2016 · TerminatedAnastoclip AC 8cm 35 Clips (1 Each} Medium REF # 4000-06 The Anastoclip Clip applier consists of a rotating shaft and an integral cartridge containing titanium clips. The Anastoc
- Z-2285-2016 · Jun 22, 2016 · TerminatedAnasto AC Closure System and Anasto Universal Clip Remover- Anasto AC 8cm 25 Clips (1 Each} X-Large REF # 4000-08 The Anastoclip Clip applier consists of a rotating shaft and
- Z-2288-2016 · Jun 22, 2016 · TerminatedAnastoCiip Universal Clip Remover (6 Pack} REF# e4001-00 The Anastoclip Clip applier consists of a rotating shaft and an integral cartridge containing titanium clips. The Anasto
- Z-2287-2016 · Jun 22, 2016 · TerminatedAnastoCiip Universal Clip Remover (1Each} REF# 4001-04 The Anastoclip Clip applier consists of a rotating shaft and an integral cartridge containing titanium clips. The Anastocl
- Z-2286-2016 · Jun 22, 2016 · TerminatedAnastoCiip Universal Clip Remover (6 Pack} REF# 4001-00 The Anastoclip Clip applier consists of a rotating shaft and an integral cartridge containing titanium clips. The Anasto
- Z-2284-2016 · Jun 22, 2016 · TerminatedAnastoClip AC 8cm 35 Clips (1 Each} Large REF # 4000-07 The Anastoclip Clip applier consists of a rotating shaft and an integral cartridge containing titanium clips. The Anastoc
- Z-2289-2016 · Jun 22, 2016 · TerminatedAnastoCiip Universal Clip Remover (1Each} REF# e4001-04 The Anastoclip Clip applier consists of a rotating shaft and an integral cartridge containing titanium clips. The Anastocl
Common questions about the Anastoclip AC Closure System 4000-07
Has the Anastoclip AC Closure System 4000-07 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00840663102372), based on the FDA recall file dated Nov 27, 2026. However, 7 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the Anastoclip AC Closure System 4000-07?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jan 22, 2016 and Mar 27, 2019. In those reports, the most common specific device problems are “Mechanical Jam” (2) and “Migration or Expulsion of Device” (2); the most common patient problems are “Cerebrospinal Fluid Leakage” (1), “Failure to Anastomose” (1) and “Fistula” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Anastoclip AC Closure System 4000-07?
None of the reports are classified as deaths, 2 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Anastoclip AC Closure System 4000-07 FDA approved?
It was cleared through 510(k) K962043 on Sep 23, 1996. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Anastoclip AC Closure System 4000-07, and is it still sold?
The FDA's device database lists Lemaitre Vascular, Inc. as the company responsible for it (brand Anastoclip AC Closure System, model 4000-07). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Clip, Implantable devices
- LIGAMAX EL5MLEthicon Endo-Surgery, LLC · CLIP, IMPLANTABLE3 deaths3,769 reports
- Endo Clip III 176630Covidien LP · CLIP, IMPLANTABLE1 death1,943 reports
- Epix® Universal Clip Applier CA500Applied Medical Resources Corporation · Clip, Implantable1 recall1 death1,176 reports
- Endo Clip II 176657Covidien LP · CLIP, IMPLANTABLE1,176 reports
- LIGACLIP ER320Ethicon Endo-Surgery, LLC · CLIP, IMPLANTABLE1,098 reports
- Direct Drive® Clip Applier CA090Applied Medical Resources Corporation · Clip, Implantable1 death806 reports
- LAPRA-TY XC200Ethicon Inc. · CLIP, IMPLANTABLE779 reports
- Endo Clip 176625Covidien LP · CLIP, IMPLANTABLE1 death767 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page