Skip to content

Brivo 615 5376496-15Recalls & reported problems

GE Hualun Medical Systems Co., Ltd · TANDEM GANTRY-B615 Medical Device · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K113350); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Dec 18, 2024Class IIOpen, Classified

    GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

    Z-0889-2025 · Root cause: Package design/selection · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113350. One submission often covers many models, so check the product and lot details in each record.

  • Z-1055-2023 · Dec 19, 2022 · Open, Classified
    GE Nuclear Medicine 600/800 series systems, specifically: a) BRIVO B615 3/8 inch DHP BJG NR, b) BRIVO 615 3/8 inch, c) BRIVO B615 3/8 inch DHP BJG NR, d) B615 with new rotor,
  • Z-0367-2014 · Oct 25, 2013 · Terminated
    GE Healthcare Discovery NMCT 670, Discovery NM 630, Optima NMCT 640, & Brivo NM 615. The GE Discovery NMJCT 670 system is a medical tool intended for use by appropriately trained
  • Z-1790-2013 · Jun 13, 2013 · Terminated
    Brivo NM615, is an all-purpose, single detector integrated nuclear imaging system. Product Usage: The Brivo NM 615 is an emission computed tomography system intended to detect

Other System, x-ray, tomography, computed devices

All System, x-ray, tomography, computed devices →

Sources and updates