Discovery RTRecalls & reported problems
GE Hangwei Medical Systems Company Ltd. · Discovery RT designated medical device · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
4 other recalls cover devices cleared under the same FDA submission (K093581, K132410); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Nov 5, 2024Class IIOpen, Classified
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
Z-0600-2025 · Root cause: Component design/selection · Recall details · FDA record ↗
- Jun 2, 2022Class IIOpen, Classified
As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.
Z-1511-2022 · Root cause: Error in labeling · Recall details · FDA record ↗
FDA clearance and classification
Cleared through 510(k) K132410 on Jan 17, 2014 (decision: Substantially Equivalent). Submitted by Ge Medical Systems, LLC as “Discovery Ct 590 Rt, Optima Ct 580, Optima Ct 580 Rt, Optima Ct 580 W”. The FDA review took 168 days. FDA record ↗
Cleared through 510(k) K093581 on Apr 19, 2011 (decision: Substantially Equivalent). Submitted by Ge Medical Systems, LLC as “Discovery Model Ct590 Rt And Optima Model Ct580 Ct Systems”. The FDA review took 517 days. FDA record ↗
The FDA regulates product code JAK (system, x-ray, tomography, computed) under 21 CFR 892.1750 as a Class II device. 21 CFR 892.1750 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K093581, K132410. One submission often covers many models, so check the product and lot details in each record.
- Z-0613-2025 · Nov 5, 2024 · Open, ClassifiedGE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B79632DA: EU: RT Refresh Package, Non-EU: Discovery RT, Optima CT580, or Discovery CT590 RT; X-ray/comput
- Z-2581-2018 · May 24, 2018 · TerminatedDiscovery RT labeled as: a. MID BJG; b. HVY BJG
- Z-2184-2017 · May 19, 2017 · TerminatedGE Healthcare Optima CT580; Discovery RT, CT590 RT X-ray system
- Z-1245-2014 · Feb 14, 2014 · TerminatedGE Healthcare Optima CT580/Discovery CT590RT scanners running software version 13BW16.3. The Discovery CT 590 RT and Optima CT 580 series of Computed Tomography systems are intend
Common questions about the Discovery RT
Has the Discovery RT been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-0600-2025 (Class II), began on Nov 5, 2024: GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical,… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
Is the Discovery RT FDA approved?
It was cleared through 510(k) K093581 on Apr 19, 2011; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Discovery RT, and is it still sold?
The FDA's device database lists GE Hangwei Medical Systems Company Ltd. as the company responsible for it (brand Discovery, model RT). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other System, x-ray, tomography, computed devices
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths49 reports
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls23 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
- Spectral CTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed11 recalls3 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, X-Ray, Tomography, Computed9 recalls20 reports
- Brilliance iCTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed8 recalls5 reports
- Revolution Apex EliteGE Medical Systems, L.L.C. · System, x-ray, tomography, computed8 recalls0 reports
- Brilliance CT Big BorePhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed7 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page