Skip to content

Discovery NMCT 670 CZT H3906ABRecalls & reported problems

G E Medical Systems Israel Ltd. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

6 other recalls cover devices cleared under the same FDA submission (K153402); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Dec 18, 2024Class IIOpen, Classified

    GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

    Z-0895-2025 · Root cause: Package design/selection · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K153402. One submission often covers many models, so check the product and lot details in each record.

  • Z-1057-2023 · Dec 19, 2022 · Open, Classified
    GE Nuclear Medicine 600/800 series systems, specifically: a) 870 CZT O540+ Gantry 7.25mm, b) D670 CZT Gantry 7.25mm, c) DISCOV NM/CT 870 CZT, d) Discovery NMCT 670 CZT O540+,
  • Z-0855-2021 · Dec 10, 2020 · Terminated
    GE NM/CT 870 CZT system Model Number H3906CW - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient bo
  • Z-2993-2020 · Aug 14, 2020 · Terminated
    NM/CT 870 CZT Model # H3906CW
  • Z-1264-2019 · Mar 20, 2019 · Terminated
    NM/CT 870 CZT, Model/Catalogue/Code # H3906CW Product Usage: The GE Discovery NM/CT 670 CZT system is intended for general Nuclear Medicine imaging procedures for detection of
  • Z-1258-2019 · Mar 20, 2019 · Terminated
    Discovery NM/CT 670 CZT, Model/Catalogue/Code H3906CT Product Usage: The GE Discovery NM/CT 670 CZT system is intended for general Nuclear Medicine imaging procedures for detec
  • Z-2195-2017 · May 19, 2017 · Terminated
    GE Healthcare Discovery NM/CT 670 CZT X-ray system

Other System, x-ray, tomography, computed devices

All System, x-ray, tomography, computed devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page