Skip to content

INTERACT Discovery RT 001Recalls & reported problems

Ge Medical Systems · INTERACT DISCOVERY RT 001 UD · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K091673); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Mar 18, 2024Open, Classified

    GE HealthCare has become aware of the potential that adhesive material might not have been applied to some bolts of certain Innova IGS, Discovery IGS, Interact IGS, Allia IGS and Allia Pulse IGS products. This could result in a potential fall of the Detector/Elevator or X-ray tube of the frontal Gantry assembly.

    Z-1779-2024 · Root cause: Process control · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K091673. One submission often covers many models, so check the product and lot details in each record.

  • Z-2187-2017 · May 19, 2017 · Terminated
    GE Healthcare lnnova IGS 540 ACT X-ray system

Other System, x-ray, tomography, computed devices

All System, x-ray, tomography, computed devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.