Revolution AceRecalls & reported problems

GE Hangwei Medical Systems Company Ltd. · Revolution Ace CT designated medical device · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840682146159), received between Nov 8, 2024 and Nov 8, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

4 other recalls cover devices cleared under the same FDA submission (K192686); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 5, 2024Open, Classified

    GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

    Z-0603-2025 · Root cause: Component design/selection · FDA record ↗

  2. Jun 2, 2022Open, Classified

    As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.

    Z-1511-2022 · Root cause: Error in labeling · FDA record ↗

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method1

Patient problems

  • Limb Fracture1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 8, 2024Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT A PATIENT SUSTAINED AN ELBOW FRACTURE WHEN THEIR ARM (WHICH WAS NOT SECURED WITH PATIENT STRAPS) MOVED OFF THE TABLE AND STRUCK THE GANTRY WHILE THE TABLE WAS MOVING INTO THE BORE. THERE IS NO INDICATION OF DEVICE MALFUNCTION.

    Patient problems: Limb Fracture

    MAUDE report 20641460 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K192686. One submission often covers many models, so check the product and lot details in each record.

  • Z-0612-2025 · Nov 5, 2024 · Open, Classified
    GE Healthcare Revolution Maxima Select, Model Number, 6670000-55; X-ray/computed tomography system
  • Z-0606-2025 · Nov 5, 2024 · Open, Classified
    GE Healthcare Revolution Ascend Elite, Model Number 6967000-100; X-ray/computed tomography system
  • Z-0610-2025 · Nov 5, 2024 · Open, Classified
    GE Healthcare Revolution Maxima, Model Numbers: 1) 6665000-2, 2) 6670000-2, 3) 6670000-200, 4) 6670000-6; X-ray/computed tomography system
  • Z-0611-2025 · Nov 5, 2024 · Open, Classified
    GE Healthcare Revolution Maxima M, Model Number, 6670000-2; X-ray/computed tomography system

Other System, x-ray, tomography, computed devices

All System, x-ray, tomography, computed devices →