Skip to content

Revolution CT with Apex editionRecalls & reported problems

GE Medical Systems, L.L.C. · Revolution CT with Apex edition - Designated System Medical Device · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

18 other recalls cover devices cleared under the same FDA submission (K191777); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Feb 18, 2025Class IIOpen, Classified

    GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

    Z-1302-2025 · Root cause: Process change control · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K191777. One submission often covers many models, so check the product and lot details in each record.

  • Z-2154-2026 · Mar 26, 2026 · Open, Classified
    Revolution Apex, X-ray Computed Tomography, Model Number 5590000-20
  • Z-2153-2026 · Mar 26, 2026 · Open, Classified
    Revolution CT ES, X-ray Computed Tomography, Model Number 5590000-15
  • Z-2150-2026 · Mar 26, 2026 · Open, Classified
    Revolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6
  • Z-2152-2026 · Mar 26, 2026 · Open, Classified
    Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25
  • Z-1303-2025 · Feb 18, 2025 · Open, Classified
    GE Healthcare Revolution CT ES, System, X-ray, Tomography, Computed
  • Z-1300-2025 · Feb 18, 2025 · Open, Classified
    GE Healthcare Revolution Apex Plus, System, X-ray, Tomography, Computed
  • Z-1297-2025 · Feb 18, 2025 · Open, Classified
    GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed
  • Z-0816-2022 · Jun 30, 2021 · Open, Classified
    Revolution CT, Revolution CT ES
  • Z-0815-2022 · Jun 30, 2021 · Open, Classified
    Revolution Apex, Revolution CT with Apex Edition
  • Z-2021-2021 · Jun 30, 2021 · Open, Classified
    The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission projection data from the same axial plane taken at different a
  • Z-0242-2022 · Jun 30, 2021 · Open, Classified
    Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
  • Z-2022-2021 · Jun 30, 2021 · Open, Classified
    The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission projection data from the same axial plane taken at different a
  • Z-2004-2021 · Mar 19, 2021 · Completed
    Revolution CT
  • Z-1130-2021 · Dec 17, 2020 · Terminated
    System, X-Ray, Tomography, Computed: Revolution CT, Revolution CT ES.
  • Z-1129-2021 · Dec 17, 2020 · Terminated
    System, X-Ray, Tomography, Computed: Revolution Apex, Revolution CT with Apex Edition

…and 3 more.

Other System, x-ray, tomography, computed devices

All System, x-ray, tomography, computed devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page