BoneScalpel MXB-20Recalls & reported problems
Misonix, Inc. · BoneScalpel 20mm, Standard Blunt Blade · Class U
All models of BoneScalpel (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00841626101142), received between Nov 30, 2017 and Jan 19, 2018 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Human Factors Issue” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Human Factors Issue1
- Human-Device Interface Problem1
- Use of Device Problem1
- Use of Incorrect Control/Treatment Settings1
Patient problems
- Disc Impingement1
- Headache1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 19, 2018Human-Device Interface Problem; Human-Device Interface Problem
CUSTOMER STATED EVENT: DURING A LATERAL TRANSTHORACIC RETRO-PLEURAL CORPECTOMY WITH REMOVAL OF HERNIATED DISC, THE SURGEON WAS USING THE BONE SCALPEL AND THE TIP OF THE BLADE BROKE OFF. AN X-RAY WAS PERFORMED TO ASSURE THAT THE TIP WAS NOT IN THE SURGICAL WOUND. THE X-RAY WAS REPORTED AS NEGATIVE BY THE RADIOLOGIST. IT IS BELIEVED THAT THE TIP WAS SUCTIONED INTO THE SUCTION CANISTER. THERE WAS NO HARM TO THE PATIENT.
Patient problems: Disc Impingement
MAUDE report 7205621 ↗InjuryNov 30, 2017Use of Incorrect Control/Treatment Settings; Use of Device Problem; Human Factors Issue; Use of Incorrect Control/Treatment Settings; Use of Device Problem; Human Factors Issue
THE VERBATIM REPORT FROM THE USER IS AS FOLLOWS: DETAILS : (B)(6) 2017 - PROF (B)(6) WAS PERFORMING A LAMINECTOMY OF L5-S1. AT THE END OF THE CASE WE NOTICED THE BLADE HAD BLACKENED, SUGGESTING OVERHEATING. WE HAD REDUCED THE IRRIGATION TO 30% DUE TO COMPLAINTS ABOUT TOO MUCH WATER IN THE WOUND. THE PULSE HAD BEEN RAISED TO 9. AFTER SHOWING NO ADVERSE EFFECTS POST OP FOR 2 OR 3 DAYS, THE PATIENT SHOWED SYMPTOMS CONSISTENT WITH A DURAL LEAK. THE PATIENT WAS REVISED 5 DAYS LATER ON (B)(6) 2017. PROF (B)(6) REPORTS A NECROSED SCAR ON THE DURA IN LINE WITH THE LAMINA REMOVAL. HIS STRONG BELIEF IS THAT THE BLADE HAD BURNT THE DURA AND THE NECROSED TISSUE HAS GRADUALLY GIVEN WAY, LEADING TO THE LEAK.
Patient problems: Headache
MAUDE report 7071144 ↗
Other Instrument, Ultrasonic Surgical devices
- Harmonic ACE HARH36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical4 deaths7,609 reports
- Harmonic HARHD36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical1 recall2 deaths5,037 reports
- Harmonic ACE HAR36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical2,726 reports
- Harmonic HAR1136Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical2 deaths1,538 reports
- THUNDERBEAT 5 mm,35 cm, Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls1,373 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls2 deaths1,128 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical2 recalls1 death970 reports
- Harmonic ACE HARH23Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical935 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page