Skip to content

Drills DRILL-1.6/095CRecalls & reported problems

Trimed, INC. · Drill 1.6, 95mm Cannulated · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00842188109607), received between Feb 9, 2018 and Jul 31, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Break1
  • Device Issue1

Patient problems

  • Foreign Body In Patient1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 31, 2023Break; Break

    SALES REP REPORTED ON (B)(6) 2023 THAT WHEN THE SURGEON WAS DRILLING INTO THE PATIENT'S BONE, THE TIP OF DRILL-1.6/095C STARTED TO SPLIT. IT WAS REPORTED THAT THE SURGEON MADE IT INTO JUST A SMALL PORTION OF THE BONE BEFORE NOTICING SOMETHING DID NOT SEEM RIGHT. IT WAS ALSO STATED THAT THE PATIENT HAD GOOD BONE QUALITY AND THE PROPER TECHNIQUE WAS BEING FOLLOWED. THE INCIDENT CAUSED A 45-MINUTE DELAY IN SURGERY AS THE SURGEON ATTEMPTED TO REMOVE THE DRILL BITS, HOWEVER THEY WERE IMBEDDED IN THE SCAPHOID AND THEY COULD NOT FIND THEM. THE REP REPORTED THERE ARE 4 PIECES OF THE DRILL LEFT IN THE PATIENT WHICH COULD LEAD TO COMPLICATIONS IN THE FUTURE IF ANY OF THE PIECES ARE OUTSIDE OF THE SCAPHOID WHICH COULD DAMAGE THE SURROUNDING SOFT TISSUE. X-RAYS AND A PICTURE WERE PROVIDED HOWEVER THE REMAINDER OF THE DRILL WAS DISCARDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 17427172 ↗
  • MalfunctionFeb 9, 2018Device Issue; Device Issue

    DURING SURGERY ON (B)(6) 2017, A SURGEON INSERTED DRILL-1.6/095 (LOT#196758) OVER A BENT K-WIRE, AND THE DRILL SPLIT OPEN AT THE BEND IN K-WIRE, OUTSIDE THE BONE. THE DRILL AND K-WIRE WERE REMOVED, AND THE REP CONFIRMED THAT THERE WERE NO BROKEN PIECES OR DEBRIS CREATED FROM THE DRILL SPLITTING. THIS IS THE FIRST OCCURRENCE OF THIS KIND FOR THIS TYPE OF PRODUCT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7260133 ↗

FDA clearance and classification

The FDA regulates product code HTW (bit, drill) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Drills DRILL-1.6/095C

Has the Drills DRILL-1.6/095C been recalled?

No FDA recall record in our data names this exact device identifier (DI 00842188109607), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Drills DRILL-1.6/095C?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Feb 9, 2018 and Jul 31, 2023. In those reports, the most common specific device problems are “Break” (1) and “Device Issue” (1); the most common patient problem is “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Drills DRILL-1.6/095C?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Drills DRILL-1.6/095C FDA approved?

Devices of this type (product code HTW) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Trimed, INC.

Who makes the Drills DRILL-1.6/095C, and is it still sold?

The FDA's device database lists Trimed, INC. as the company responsible for it (brand Drills, model DRILL-1.6/095C). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other BIT, DRILL devices

All BIT, DRILL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page