Prodisc C 09.820.035SRecalls & reported problems
Centinel Spine, LLC · PRODISC-C IMPLANT MEDIUM DEEP 5MM-STERILE · Class 3 · Implant
All models of Prodisc C (12)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00843193112781), received between Jun 30, 2020 and May 3, 2023 — classified as 3 injuries.
The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (1 report).
4 other recalls cover devices cleared under the same FDA submission (P070001); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Inadequacy of Device Shape and/or Size1
- Malposition of Device1
- Migration1
- Migration or Expulsion of Device1
- Patient-Device Incompatibility1
Patient problems
- Neck Pain2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMay 3, 2023Migration or Expulsion of Device; Malposition of Device; Migration or Expulsion of Device; Malposition of Device
IT WAS REPORTED THAT A PATIENT RECEIVED A PRODISC C IMPLANT ON (B)(6) 2022. THE PATIENT EXPERIENCED INCREASED NECK AND ARM PAIN AND AN X-RAY FOUND THAT THERE WAS A CHANGE IN IMPLANT POSITION RESULTING IN KYPHOSIS AND FISH MOUTHING OF THE IMPLANT. A PRODISC C REMOVAL WAS COMPLETED ON (B)(6) 2023. PATIENT WAS REVISED BY A CORELINK 3D PRINTED TITANIUM CAGE.
Patient problems: Neck Pain
MAUDE report 16859741 ↗InjurySep 24, 2021Migration; Migration
PATIENT RECEIVED A PRODISC C US IMPLANT MEDIUM DEEP 5MM ON (B)(6) 2021 IN THE C5/6 DISC SPACE. FOLLOW UP IMAGING REVEALED THE PRODISC C DEVICE MIGRATED ANTERIORLY. LATERAL X-RAYS PROVIDED CONFIRM THIS REPORTED MALFUNCTION. THE PATIENT WAS SCHEDULED FOR REMOVAL TO PRECLUDE SERIOUS INJURY. THERE WAS NO INDICATION OF PATIENT COMPLICATIONS OR SYMPTOMS. THERE WERE NO REPORTED COMORBIDITIES OR PRE-EXISTING CONDITIONS. ON (B)(6)2021 THE PRODISC C DEVICE WAS REMOVED AND CONVERTED TO A FUSION. THERE WERE NO INDICATIONS OF COMPLICATIONS DURING THE REMOVAL PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12526721 ↗InjuryJun 30, 2020Inadequacy of Device Shape and/or Size; Patient-Device Incompatibility; Inadequacy of Device Shape and/or Size; Patient-Device Incompatibility
A PATIENT REQUIRED SURGICAL INTERVENTION TO REMOVE A PRODISC C IMPLANT DUE TO CONTINUED PAIN. THE PATIENT WAS EXPERIENCING PAIN, SPECIFICALLY WHEN LOOKING UP. THIS PATIENT WAS IMPLANTED WITH THE DEVICE ON (B)(6) 2019. THE SURGEON INDICATED THAT THE IMPLANTED PRODISC C DEVICE WAS TOO TALL DESPITE BEING A 5MM (SHORTEST) HEIGHT IMPLANT.
Patient problems: Neck Pain
MAUDE report 10212572 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P070001. One submission often covers many models, so check the product and lot details in each record.
- Z-2051-2026 · Mar 20, 2026 · Open, ClassifiedProdisc C SK U.S. IMPLANT EXTRA LARGE 6MM. Model Number: PDSXL6. Total cervical disc replacement.
- Z-2050-2026 · Mar 20, 2026 · Open, ClassifiedProdisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.
- Z-1426-2026 · Dec 31, 2025 · Open, ClassifiedBrand Name: Prodisc Product Name: Prodisc C SK Model/Catalog Number: PDSL6 Software Version: N/A Product Description: Total cervical disc replacement Component: No
- Z-1406-2021 · Mar 5, 2021 · TerminatedProdisc C - prosthesis, intervertebral disc - Product Usage: indicated in skeletally mature patients for reconstruction of the disc from C3-C7 following single-level discectomy for
Other Prosthesis, Intervertebral Disc devices
- Prodisc C SK PDSM5Centinel Spine, LLC · Prosthesis, Intervertebral Disc2 recalls4 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall12 reports
- Prodisc C 09.820.055SCentinel Spine, LLC · Prosthesis, Intervertebral Disc1 recall9 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall4 reports
- Prodisc C SK PDSXL5Centinel Spine, LLC · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.