Skip to content

Prodisc C Vivo PDVMD5Recalls & reported problems

Centinel Spine, LLC · Prodisc C Vivo, Medium Deep, 5mm · Class 3 · Implant

All models of Prodisc C Vivo (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00843193113702), received between Jul 1, 2024 and May 8, 2026 — classified as 5 injuries.

The most frequently reported specific device problem is “Migration” (3 reports).

4 other recalls cover devices cleared under the same FDA submission (P070001); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration3
  • Adverse Event Without Identified Device or Use Problem1
  • Patient-Device Incompatibility1

Patient problems

  • Hypersensitivity/Allergic reaction1
  • Insufficient Information1
  • Limb Pain1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 8, 2026Adverse Event Without Identified Device or Use Problem

    A PATIENT (42YO, FEMALE) WAS IMPLANTED WITH A PDC VIVO DEVICE AT C5-6 ON (B)(6) 2026. NO INFORMATION ON PATIENT SYMPTOMS WAS PROVIDED. HOWEVER, IT WAS STATED THAT THE POSTERIOR AREA MAY NOT HAVE BEEN ADDRESSED WELL ENOUGH AND ADDITIONALLY THAT THE IMPLANT WAS REMOVED ON (B)(6) 2026 AND THE PATIENT WAS CONVERTED TO A THREE LEVEL FUSION SO THERE MAY HAVE BEEN OTHER AFFECTED LEVELS. HOWEVER, NO REASON FOR REMOVAL WAS CONFIRMED.

    Patient problems: Insufficient Information

    MAUDE report 25124277 ↗
  • InjurySep 10, 2025Patient-Device Incompatibility; Patient-Device Incompatibility

    A PATIENT WAS IMPLANTED WITH TWO PRODISC C VIVO IMPLANTS AT C5-6 AND C6-7 ON (B)(6) 2024. PATIENT BECAME SYMPTOMATIC AND RECEIVED A FACET INJECTION IN (B)(6) 2024 TO NO SIGNIFICANT IMPROVEMENT. IT WAS FOUND THAT THE PATIENT HAD A PREVIOUSLY UNKNOWN NICKEL ALLERGY. A SECOND SURGEON REMOVED THE IMPLANTS ON (B)(6) 2025 AND REPLACED THEM WITH A NICKEL FREE DISC REPLACEMENT DEVICE.

    Patient problems: Hypersensitivity/Allergic reaction

    MAUDE report 23016692 ↗
  • InjuryOct 14, 2024Migration; Migration

    A PATIENT WAS IMPLANTED WITH TWO PDC VIVO DEVICES ON (B)(6) 2024. AT A FOLLOW UP IT WAS FOUND THAT ONE OF THE LEVELS HAD MIGRATED, THE SURGEON COMPLETED A REMOVAL OF LEVEL C5-6 AND FUSED THAT SEGMENT ON (B)(6) 2024.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20438956 ↗
  • InjuryJul 16, 2024Migration; Migration

    A PATIENT WAS IMPLANTED WITH A PRODISC C VIVO IMPLANT ON (B)(6) 2023. PATIENT WAS EXPERIENCING PAIN AND IT WAS FOUND THAT THE IMPLANT HAD MIGRATED ANTERIORLY. THE SURGEON REMOVED THE IMPLANT ON (B)(6) 2024 AND THE PATIENT WAS FUSED.

    Patient problems: Pain

    MAUDE report 19751237 ↗
  • InjuryJul 1, 2024Migration; Migration

    A PATIENT RECEIVED TWO PRODISC C VIVO IMPLANT ON (B)(6) 2023. ONE OF THE IMPLANTS WAS FOUND TO BE DISPLACED AND WAS CAUSING RADICULOPATHY IN THE PATIENT. NO DEVICE FAILURE INDICATED. THE DISPLACED IMPLANT WAS REMOVED ON (B)(6) 2024 AND REPLACED WITH A SEASPINE IMPLANT.

    Patient problems: Limb Pain

    MAUDE report 19650387 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P070001. One submission often covers many models, so check the product and lot details in each record.

  • Z-2050-2026 · Mar 20, 2026 · Open, Classified
    Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.
  • Z-2051-2026 · Mar 20, 2026 · Open, Classified
    Prodisc C SK U.S. IMPLANT EXTRA LARGE 6MM. Model Number: PDSXL6. Total cervical disc replacement.
  • Z-1426-2026 · Dec 31, 2025 · Open, Classified
    Brand Name: Prodisc Product Name: Prodisc C SK Model/Catalog Number: PDSL6 Software Version: N/A Product Description: Total cervical disc replacement Component: No
  • Z-1406-2021 · Mar 5, 2021 · Terminated
    Prodisc C - prosthesis, intervertebral disc - Product Usage: indicated in skeletally mature patients for reconstruction of the disc from C3-C7 following single-level discectomy for

Other Prosthesis, Intervertebral Disc devices

All Prosthesis, Intervertebral Disc devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page