Prodisc C Vivo PDVL5Recalls & reported problems

Centinel Spine, LLC · Prodisc C Vivo, Large, 5mm · Class 3 · Implant

All models of Prodisc C Vivo (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
20
Deaths reported
0
Injuries reported
19
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 00843193113733), received between Jun 21, 2023 and Sep 15, 2026 — classified as 19 injuries, 1 malfunction.

The most frequently reported specific device problem is “Migration” (11 reports).

4 other recalls cover devices cleared under the same FDA submission (P070001); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 4 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration11
  • Adverse Event Without Identified Device or Use Problem4
  • Osseointegration Problem2
  • Patient-Device Incompatibility2
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Insufficient Information8
  • Neck Pain5
  • Limb Pain4
  • Pain3
  • Hypersensitivity/Allergic reaction2
  • Burning Sensation1
  • Extraskeletal ossification1
  • Osteolysis1

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • InjurySep 15, 2026Migration

    A patient was implanted with a 2-level Prodisc C Vivo device on (b)(6) 2025. The patient reporting having neck pain. Imaging showed that the implant had migrated anteriorly and possible osteolysis. The implants were removed on (b)(6) 2026 and the patient was fused.

    Patient problems: Neck Pain

    MAUDE report 50773739 ↗
  • InjuryJul 29, 2026Migration

    A patient was implanted with a Prodisc C Vivo device at C5-6 on (b)(6) 2025. The patient was experiencing neck pain. It was found that the superior endplate had migrated anteriorly. The implant was removed on (b)(6) 2026 and the patient was fused.

    Patient problems: Neck Pain

    MAUDE report 50073418 ↗
  • InjuryJun 5, 2026Migration

    A PATIENT (53 YO, FEMALE) WAS IMPLANTED WITH A PDC VIVO DEVICE AT LEVEL C5-6 ON (B)(6) 2024. THE PATIENT WAS EXPERIENCING NECK PAIN AND MILD RIGHT ARM PAIN. IT WAS FOUND THAT THE IMPLANT HAD MIGRATED ANTERIORLY. THE IMPLANT WAS REMOVED ON (B)(6) 2026 AND THE PATIENT WAS FUSED.

    Patient problems: Limb Pain; Neck Pain

    MAUDE report 25407547 ↗
  • MalfunctionMay 5, 2026Migration

    A PATIENT (54YO, MALE) WAS IMPLANTED WITH A PDC VIVO DEVICE AT LEVEL C6-7 IN (B)(6) 2025. AT 3 MONTHS POSTOP IT WAS FOUND THAT THE IMPLANT HAD MIGRATED.

    Patient problems: Insufficient Information

    MAUDE report 25087094 ↗
  • InjuryNov 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A PATIENT WAS IMPLANTED WITH A PDC VIVO DEVICE ON (B)(6) 2025. IT WAS STATED THAT THE PATIENT HAD NO REAL SYMPTOMS BUT THAT THE SURGEON JUST DECIDED TO FUSE. THE IMPLANT WAS REMOVED ON (B)(6) 2025.

    Patient problems: Insufficient Information

    MAUDE report 23631438 ↗
  • InjuryNov 14, 2025Migration; Migration

    A PATIENT WAS IMPLANTED WITH TWO PRODISC C VIVO DEVICES ON (B)(6) 2024. IT WAS FOUND THAT THE IMPLANTS WERE LOOSENING. THE VIVO DEVICES WERE REMOVED AND REPLACED WITH PRODISC C LEGACY DEVICES ON (B)(6) 2025.

    Patient problems: Insufficient Information

    MAUDE report 23555978 ↗
  • InjuryNov 12, 2025Migration; Migration

    A PATIENT WAS IMPLANTED WITH THREE PRODISC C VIVO IMPLANTS ON (B)(6) 2023. IT WAS FOUND THAT THE VIVO IMPLANTS APPEARED TO BE LOOSENING IN THE PATIENT. A REMOVAL WAS COMPLETED ON (B)(6) 2025 AND REPLACED WITH PRODISC C LEGACY DEVICES.

    Patient problems: Insufficient Information

    MAUDE report 23538330 ↗
  • InjuryNov 12, 2025Migration; Migration

    A PATIENT WAS IMPLANTED WITH THREE PRODISC C VIVO IMPLANTS ON (B)(6) 2023. IT WAS FOUND THAT THE VIVO IMPLANTS APPEARED TO BE LOOSENING IN THE PATIENT. A REMOVAL WAS COMPLETED ON (B)(6) 2025 AND REPLACED WITH PRODISC C LEGACY DEVICES.

    Patient problems: Insufficient Information

    MAUDE report 23538308 ↗
  • InjurySep 16, 2025Osseointegration Problem; Osseointegration Problem

    A PATIENT WAS IMPLANTED WITH A PRODISC C VIVO DEVICE ON (B)(6) 2023. THE PATIENT BEGAN TO EXPERIENCE A REOCCURRENCE OF ARM PAIN AND IT WAS FOUND THAT THE IMPLANT HAD SUBSIDED. THE SURGEON REMOVED THE IMPLANT ON (B)(6) 2025 AND FUSED THE PATIENT.

    Patient problems: Limb Pain

    MAUDE report 23062433 ↗
  • InjuryAug 6, 2025Patient-Device Incompatibility; Patient-Device Incompatibility

    A PATIENT WAS IMPLANTED WITH A PDC SK AT C5-6 AND A PDC VIVO AT C6-7 ON (B)(6) 2025. THE PATIENT WAS FOUND TO HAVE A NICKEL ALLERGY AND REQUIRED THE IMPLANTS TO BE REMOVED. THE IMPLANTS WERE BOTH REMOVED ON (B)(6) 2025 AND THE PATIENT WAS FUSED AT BOTH LEVELS.

    Patient problems: Hypersensitivity/Allergic reaction

    MAUDE report 22725959 ↗
  • InjuryJul 14, 2025Migration; Migration

    A PATIENT WAS IMPLANTED WITH TWO PDC VIVO DEVICES ON (B)(6) 2024. IT WAS FOUND THAT BOTH IMPLANTS HAD MIGRATED ANTERIORLY. NO SYMPTOMS WERE REPORTED. THE DECISION WAS MADE TO REMOVE THE IMPLANTS, THE REMOVAL SURGERY WAS COMPLETED ON (B)(6) 2025 AND THE PATIENT WAS FUSED.

    Patient problems: Insufficient Information

    MAUDE report 22493817 ↗
  • InjuryJul 14, 2025Migration; Migration

    A PATIENT WAS IMPLANTED WITH TWO PDC VIVO DEVICES ON (B)(6) 2024. IT WAS FOUND THAT BOTH IMPLANTS HAD MIGRATED ANTERIORLY. NO SYMPTOMS WERE REPORTED. THE DECISION WAS MADE TO REMOVE THE IMPLANTS, THE REMOVAL SURGERY WAS COMPLETED ON (B)(6) 2025 AND THE PATIENT WAS FUSED.

    Patient problems: Insufficient Information

    MAUDE report 22493816 ↗
  • InjuryMar 18, 2025Migration; Migration

    A PATIENT WAS IMPLANTED WITH A PDC VIVO IMPLANT AT LEVEL C4-5 ON (B)(6) 2024. THE PATIENT HAD A FOLLOW UP APPOINTMENT WHERE X-RAYS REVEALED DETERIORATING BONE BELOW THE IMPLANT. AT THAT TIME, THE PATIENT HAD NO PAIN. THE PATIENT RECENTLY DEVELOPED PAIN AND ADDITIONAL X-RAYS REVEALED MORE DETERIORATION OF BONE BELOW THE IMPLANT. IT IS UNCERTAIN WHAT CAUSED THE BONE DETERIORATION. THE SURGEON REMOVED THE IMPLANT AND FUSED THE LEVEL ON (B)(6) 2025. THE PATIENT HAD NO ISSUES DURING THE REMOVAL SURGERY AND IS DOING WELL NOW.

    Patient problems: Neck Pain; Osteolysis

    MAUDE report 21633302 ↗
  • InjuryMar 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT WAS IMPLANTED WITH A PRODISC C VIVO DEVICE AT C5-6 ON (B)(6) 2025. PATIENT WAS EXPERIENCING ONGOING PAIN AND AN MRI WAS ORDERED. MRI SHOWED A POSSIBLE HEMATOMA OF UNKNOWN ORIGIN; THE SURGEON WAS UNCERTAIN ABOUT THE HEMATOMA DUE TO DEVICE ARTIFACT ON THE MRI. SURGEON REMOVED THE IMPLANT ON (B)(6) 2025 AND REPLACED IT WITH A KEELED IMPLANT.

    Patient problems: Limb Pain

    MAUDE report 21510209 ↗
  • InjuryAug 28, 2024Migration; Migration

    A PATIENT WAS IMPLANTED WITH A PRODISC C VIVO IMPLANT AT C5-6 ON (B)(6) 2023. PATIENT WAS EXPERIENCING NECK PAIN AND INSTABILITY AND IT WAS FOUND THAT THE IMPLANT HAD MIGRATED POSTERIORLY. THE IMPLANT WAS REMOVED ON (B)(6) 2024 AND THE PATIENT WAS FUSED.

    Patient problems: Neck Pain

    MAUDE report 20085662 ↗
  • InjuryAug 22, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A PATIENT WAS IMPLANTED WITH TWO PDC VIVO IMPLANTS AT C5-6 AND C6-7 ON (B)(6) 2024. THE PATIENT HAD ONGOING RADICULOPATHY. A ONE LEVEL PDC VIVO REMOVAL WAS COMPLETED ON (B)(6) 2024. PATIENT WAS CONVERTED TO A CAGE AND PLATE SYSTEM. NO DEVICE ISSUES WERE SEEN.

    Patient problems: Pain

    MAUDE report 20042435 ↗
  • InjuryJun 7, 2024Osseointegration Problem; Osseointegration Problem

    A PATIENT RECEIVED A PRODISC C VIVO IMPLANT AT LEVEL C5-6 ON (B)(6) 2023. IT WAS FOUND THAT THE IMPLANT HAD SUBSIDED. A REMOVAL WAS COMPLETED ON (B)(6) 2024 AND THE PATIENT WAS CONVERTED TO FUSION.

    Patient problems: Insufficient Information

    MAUDE report 19483280 ↗
  • InjuryMay 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT ORIGINALLY HAD A 2-LEVEL PDC VIVO IMPLANTED ON (B)(6) 2022 AT C5/6 AND C6/7. THE PATIENT DID FINE FOR A LONG PERIOD OF TIME. AT SOME POINT, HE WAS HIT BY LIGHTNING, AND THAT IS WHEN HE STARTED TO HAVE PAIN. IN ADDITION TO THE LIGHTNING STRIKE, HE ALSO HAS A FAILED CARPAL TUNNEL SURGERY SO HE HAD PAIN IN HIS HAND. ALTHOUGH THE PATIENT'S IMPLANTS APPEARED TO BE IN GOOD POSITION AND DID NOT MIGRATE, HE DID HAVE EXCESSIVE BONE GROWTH ANTERIOR TO THE IMPLANT, WHICH APPEARED TO LIMIT MOTION. SINCE THE PATIENT HAD PAIN IN HIS HAND WHICH MAY OR MAY NOT HAVE BEEN GENERATED BY THE IMPLANT AT C5/6, THE SURGEON DECIDED TO REMOVE THE VIVO IMPLANT AT C5/6 AND TO FUSE THAT SEGMENT, WHILE LEAVING THE VIVO IMPLANT AT C6/7 INTACT AND IN PLACE. ADDITIONALLY, THE PATIENT HAD PAIN COMING FOR C4/5 SO THAT LEVEL WAS FUSED AS WELL. THE REVISION SURGERY TOOK PLACE ON (B)(6) 2024 AND RESULTED IN A 2-LEVEL FUSION AT C4/5/6 ABOVE A VIVO IMPLANT AT C6/7 WHICH WAS LEFT INTACT.

    Patient problems: Extraskeletal ossification; Limb Pain

    MAUDE report 19231589 ↗
  • InjuryMar 7, 2024Patient-Device Incompatibility; Patient-Device Incompatibility

    A PATIENT WAS IMPLANTED WITH TWO PRODISC C VIVO IMPLANTS ON (B)(6) 2023. PATIENT EXPERIENCED A BURNING SENSATION AND PAIN. IT WAS DETERMINED THAT THE PATIENT HAD A NICKEL ALLERGY. THE IMPLANTS WERE REMOVED ON (B)(6) 2024 AND REPLACED WITH SIMPLIFY CTDR. IT WAS FOUND AFTER REMOVAL THAT A DEVICE CULTURE DETERMINED THE PRESENCE OF AN INFECTION.

    Patient problems: Burning Sensation; Hypersensitivity/Allergic reaction; Pain; Unspecified Infection

    MAUDE report 18860775 ↗
  • InjuryJun 21, 2023Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    A PATIENT WAS IMPLANTED WITH A PRODISC C VIVO DEVICE AT LEVEL C6-7 ON (B)(6) 2023. THE IMPLANT WAS REMOVED ON (B)(6) 2023. THE PATIENT WAS COMPLAINING ABOUT PAIN IN BOTH ARMS; WITH THE RIGHT ARM BEING WORSE THAN THE LEFT AS THE REASON FOR THE REMOVAL. X-RAY IMAGES WERE PROVIDED. THE SURGEON STATED THAT THE IMPLANT PLACEMENT LOOKS PERFECT, BUT INADEQUATE DECOMPRESSION MAY HAVE OCCURRED ON THE ASYMPTOMATIC SIDE.

    Patient problems: Pain

    MAUDE report 17173396 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P070001. One submission often covers many models, so check the product and lot details in each record.

  • Z-2051-2026 · Mar 20, 2026 · Open, Classified
    Prodisc C SK U.S. IMPLANT EXTRA LARGE 6MM. Model Number: PDSXL6. Total cervical disc replacement.
  • Z-2050-2026 · Mar 20, 2026 · Open, Classified
    Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.
  • Z-1426-2026 · Dec 31, 2025 · Open, Classified
    Brand Name: Prodisc Product Name: Prodisc C SK Model/Catalog Number: PDSL6 Software Version: N/A Product Description: Total cervical disc replacement Component: No
  • Z-1406-2021 · Mar 5, 2021 · Terminated
    Prodisc C - prosthesis, intervertebral disc - Product Usage: indicated in skeletally mature patients for reconstruction of the disc from C3-C7 following single-level discectomy for

Other Prosthesis, Intervertebral Disc devices

All Prosthesis, Intervertebral Disc devices →