Skip to content

Simplify 2-5-CERecalls & reported problems

Nuvasive, Inc. · Simplify Disc – Size 2, Height 5 · Class 3 · Implant

All models of Simplify (13)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00843285101976), received between Mar 1, 2022 and Jun 26, 2023 — classified as 2 injuries, 1 malfunction.

1 other recall cover devices cleared under the same FDA submission (P200022); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3

Patient problems

  • Extraskeletal ossification2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Insufficient Information1
  • Intervertebral Disc Compression or Protrusion1
  • Limb Pain1
  • Neck Pain1
  • Pain1
  • Stenosis1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL ONE-LEVEL CERVICAL TOTAL DISC REPLACEMENT PROCEDURE USING THE SIMPLIFY DISC AT C5/6 AS A PART OF A CLINICAL STUDY AND TO ADDRESS LEFT C5/6 DISC HERNIATION. APPROXIMATELY SEVEN (7) YEARS LATER, A REVISION SURGERY WAS PERFORMED WHERE THE SIMPLIFY DISC WAS REMOVED AND A THREE-LEVEL FUSION PROCEDURE WAS PERFORMED FROM C4 TO C7. THE SURGEON CITED THE BREAKDOWN OF THE CERVICAL LEVELS SUPERIOR AND INFERIOR TO THE SIMPLIFY DISC AS THE REASON FOR THE REMOVAL OF THE DISC AND THE MOVE TO A FUSION PROCEDURE; SPECIFICALLY, TO ADDRESS SEVERE NECK AND ARM PAIN, DECREASED UPPER EXTREMITY REFLEXES AND SENSATION, SEVERE STENOSIS, CERVICAL MYELOPATHY, AND LOSS OF DISC SPACE HEIGHT AT C4/5, AND A COLLAPSING DISC AT C6/7 WITH FORAMINAL STENOSIS AT SEVERE LEFT C6/7. THE REVISION PROCEDURE WAS REPORTEDLY SUCCESSFUL WITH NO PROBLEMS ENCOUNTERED AND THERE WERE NO NOTED ISSUES WITH THE SIMPLIFY DISC ITSELF. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information; Intervertebral Disc Compression or Protrusion; Limb Pain; Neck Pain; Pain; Stenosis

    MAUDE report 17205897 ↗
  • InjuryJul 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED VIA CLINICAL STUDY THAT A PATIENT UNDERWENT A CERVICAL PROCEDURE ON (B)(6) 2018 WITH A SIMPLIFY DISC REPLACEMENT. ON (B)(6) 2022, IT WAS REPORTED THAT THE PATIENT EXPERIENCED HETEROTOPIC OSSIFICATION AT THE INDEX LEVEL OF C5/6. NO TREATMENT HAS BEEN GIVEN AND THERE IS NO PLAN TO REMOVE OR REVISE THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Extraskeletal ossification

    MAUDE report 15033015 ↗
  • MalfunctionMar 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    VIA CLINICAL STUDY, IT WAS REPORTED THAT THE PATIENT EXPERIENCED HETEROTOPIC OSSIFICATION AT THE C6/7 INDEX LEVEL OF A SIMPLIFY IMPLANT. NO CURRENT TREATMENT IS BEING PERFORMED FOR THE HETEROTOPIC OSSIFICATION AND THERE IS NO REPORTED PLAN TO REMOVE OR REVISE THE SIMPLIFY IMPLANT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Extraskeletal ossification

    MAUDE report 13643000 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P200022. One submission often covers many models, so check the product and lot details in each record.

  • Z-1453-2022 · Jun 20, 2022 · Open, Classified
    Description/Part: Simplify Disc Size 1, Ht 5/SM-5; Simplify Disc Size 1, Ht 6/SM-6; Simplify Disc Size 2, Ht 4/MD-4; Simplify Disc Size 2, Ht 5/MD-5; Simplify Disc Size 2, Ht 5, 5

Other Prosthesis, intervertebral disc devices

All Prosthesis, intervertebral disc devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page