Masimo Corporation · Rad-97 with non-invasive blood pressure · Class 2

All models of Rad-97 (3)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Masimo Rad-97 patient monitor tracks vital signs like oxygen levels and blood pressure. Its 84 adverse event reports describe blood pressure readings that stay inaccurate even after calibration, sensors that lose readings or give slow results, unexpected shutdowns, resets, and alarms that don't sound. No deaths or injuries were reported, and most reports noted no patient impact. No recalls name this device. Reports in the latest 12 months (6) were fewer than the prior 12 months (12).

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
84
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 84 adverse event reports that list this exact device (DI 00843997012775), received between Mar 23, 2018 and Sep 28, 2026 — classified as 84 malfunctions.

The most frequently reported specific device problem is “Incorrect Measurement” (18 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 5 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 3 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect Measurement18
  • Sensing Intermittently18
  • Unexpected Shutdown17
  • Inappropriate or Unexpected Reset12
  • High Readings6
  • No Audible Alarm3
  • Low Readings2
  • Overheating of Device2

Patient problems

  • No Clinical Signs, Symptoms or Conditions63
  • No Consequences Or Impact To Patient20
  • Bruise/Contusion1
  • Erythema1
  • Pain1

Recent adverse event reports

The 25 most recent of 84 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Unexpected Shutdown

    The customer reported that the device will not power on most of the time. if it does turn on, it shuts off by itself. There was no patient impact or consequence reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50910730 ↗
  • MalfunctionSep 24, 2026Sensing Intermittently

    The customer reported that the device connects for 2 mins and stops reading. TB20 connection is loose. There was no patient impact or consequence reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50877491 ↗
  • MalfunctionSep 19, 2026Unexpected Shutdown

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50832405 ↗
  • MalfunctionAug 23, 2026Incorrect Measurement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50393591 ↗
  • MalfunctionDec 8, 2025Sensing Intermittently; Sensing Intermittently

    THE CUSTOMER REPORTED A "SATURATION PROBLEM (WHICH SEEMS RELATED TO THE CIRCUIT BOARD)" WITH THE RAD-97. PER ADDITIONAL INFORMATION FROM THE CUSTOMER, THE DEVICE "LOST READING WITHOUT TOUCHING ANYTHING." THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23741527 ↗
  • MalfunctionNov 25, 2025Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THAT THE DEVICE IS "TURNING OFF DURING WORK". THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23653334 ↗
  • MalfunctionOct 31, 2025Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE CUSTOMER REPORTED THAT THE SENSOR PROVIDED SLOW AND INACCURATE READING[S] OF SPO2 AND PR. THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23444051 ↗
  • MalfunctionOct 16, 2025Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THE RAD-97 WAS "NOT STAYING ON." NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23319541 ↗
  • MalfunctionSep 4, 2025Sensing Intermittently; Sensing Intermittently

    THE CUSTOMER REPORTED THAT THE DEVICE SATURATION IS INTERMITTENT. THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22974301 ↗
  • MalfunctionAug 25, 2025Use of Device Problem; Use of Device Problem

    THE CUSTOMER REPORTED A "PAINFUL EXPERIENCE" WITH THE RAD-97 AS SHE "RECEIVED A BRUISE ON THE LEFT ARM AND RED MARKS ON BOTH ARMS WHICH DISSIPATED HOURS AFTER." NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: Bruise/Contusion; Erythema; Pain

    MAUDE report 22891469 ↗
  • MalfunctionAug 14, 2025Sensing Intermittently; Sensing Intermittently

    THE CUSTOMER REPORTED THAT THE DEVICE HAS AN "INTERMITTENT LOSS OF SPO2 AND HEART RATE SIGNAL". THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22811593 ↗
  • MalfunctionAug 12, 2025Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THE RAD-97 "SUDDENLY SHUTDOWN AFTER 3 MINUTES." NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22786613 ↗
  • MalfunctionAug 5, 2025Sensing Intermittently; Sensing Intermittently

    THE CUSTOMER REPORTED THAT AFTER A FEW MINUTES OF MONITORING, THE DEVICE STOPS MONITORING. THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22723248 ↗
  • MalfunctionAug 1, 2025Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THAT THE DEVICE "TURNS OFF BY ITSELF, FOR NO REASON". THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22689290 ↗
  • MalfunctionJul 14, 2025Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THE RAD-97 "SHUTS OFF." NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22501147 ↗
  • MalfunctionApr 24, 2025Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THE RAD-97 "POWER[ED] OFF BY ITSELF." NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21905543 ↗
  • MalfunctionFeb 20, 2025Sensing Intermittently; Sensing Intermittently

    THE CUSTOMER REPORTED THAT ONCE THE DEVICE IS "TURNED ON FOR A FEW MINUTES, IT STOPS TAKING MEASUREMENTS". THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21428416 ↗
  • MalfunctionJan 16, 2025Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED "FOUND BATTERY IS FULLCHARGED 100% AFTER 5 MINUTES SUDDENLY SHUTDOWN". THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21174916 ↗
  • MalfunctionSep 24, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THAT THE DEVICE "TURNS OFF AUTOMATICALLY DURING OPERATION AND THERE IS NO VISIBLE SIGNS OF DEMAGE." THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20293230 ↗
  • MalfunctionDec 15, 2023Sensing Intermittently; Sensing Intermittently

    THE CUSTOMER REPORTED AN "INTERMITTENT CONNECTION OF SATS LEAD". THERE WAS NO CONSEQUENCE OR IMPACT TO THE PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18336385 ↗
  • MalfunctionDec 5, 2023Sensing Intermittently; Sensing Intermittently

    THE CUSTOMER REPORTED THE SPO2 KEEPS LOSING SIGNAL FOR THE RAD-97 DEVICE. THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18271250 ↗
  • MalfunctionSep 7, 2023Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THE RADICAL-97 TURNS OFF ON ITS OWN. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17705662 ↗
  • MalfunctionSep 7, 2023Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THE RAD-97 TURNS OFF AFTER A FEW MINUTES. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17705621 ↗
  • MalfunctionJun 21, 2023Sensing Intermittently; Sensing Intermittently

    THE CUSTOMER REPORTED THE SPO2 MEASUREMENT IS INTERRUPTED SPORADICALLY. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17176204 ↗
  • MalfunctionJun 16, 2023Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THE RAD-97 WAS "GETTING ON AND OFF." NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17148781 ↗

FDA clearance and classification

The FDA regulates product code MWI (monitor, physiological, patient (without arrhythmia detection or alarms)) under 21 CFR 870.2300 as a Class II device. 21 CFR 870.2300 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Rad-97 9797

Has the Rad-97 9797 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00843997012775), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Rad-97 9797?

The FDA's MAUDE database lists 84 adverse event reports for this device, received between Mar 23, 2018 and Sep 28, 2026. In those reports, the most common specific device problems are “Incorrect Measurement” (18), “Sensing Intermittently” (18) and “Unexpected Shutdown” (17); the most common patient problems are “Bruise/Contusion” (1), “Erythema” (1) and “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Rad-97 9797?

None of the reports are classified as deaths, 0 as injuries and 84 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Rad-97 9797, and is it still sold?

The FDA's device database lists Masimo Corporation as the company responsible for it (brand Rad-97, model 9797). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) devices

All Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) devices →