OsteoMed 450-0021-00Recalls & reported problems
Osteomed LLC · Power Control Console 2 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00845694012157), received between Aug 12, 2021 and Nov 16, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Electromagnetic Compatibility Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Electromagnetic Compatibility Problem1
- Mechanical Problem1
- Unintended Electrical Shock1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Electric Shock1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 16, 2021Mechanical Problem; Unintended Movement; Mechanical Problem; Unintended Movement
ON OCTOBER 26, 2021, THE CUSTOMER CONTACTED OSTEOMED PRODUCT MANAGER BY EMAIL. PER THE EMAIL: "THE CONSOLE UNIT WILL SPONTANEOUSLY BEGIN RUNNING ON REVERSE, WITHOUT ANYONE PRESSING THE FOOT PEDAL! " THE INVOLVED PART NUMBERS ARE AS FOLLOWS: COMPLAINT #, PART #, DESCRIPTION, SERIAL #: (B)(4), 450-0084, SERIES II LOW PROFILE MODULAR MOTOR UNIT, (B)(4); (B)(4), 450-0021, POWER CONTROL CONSOLE 2, (B)(4); (B)(4), 450-0390, BMF FOOTSWITCH, (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12820679 ↗MalfunctionAug 30, 2021Electromagnetic Compatibility Problem; Electromagnetic Compatibility Problem
ON (B)(6) 2021, THE CUSTOMER CONTACTED OSTEOMED PRODUCT MANAGER, REPORTING THAT THE OSTEOMED 450-0021, POWER CONTROL CONSOLE 2 WITH SERIAL NUMBER (B)(4) WAS INTERFERING WITH GE INTERCARE AND CRITICARE MONITORS THE INVOLVED PART NUMBERS ARE AS FOLLOWS: COMPLAINT #, PART #, DESCRIPTION, SERIAL #: (B)(4), 450-0021, POWER CONTROL CONSOLE 2 (B)(4); (B)(4), 450-0390, BMF FOOTSWITCH (B)(4); (B)(4), 450-0084, SERIES II LOW PROFILE MODULAR MOTOR UNIT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12389466 ↗MalfunctionAug 12, 2021Unintended Electrical Shock; Unintended Electrical Shock
ON (B)(6) 2021, THE CUSTOMER CONTACTED THEIR OSTEOMED REPRESENTATIVE, REPORTING THAT THE CUSTOMER HAD EXPERIENCED A SIMILAR SENSATION WITH THE LOANER CONSOLE AND MOTOR THAT WAS SENT TO THEM IN EXCHANGE FOR THE UNITS LISTED ON THE FER (B)(6) 2021. 6/21/2021 - EMAIL RECEIVED FROM PRODUCT MANAGER, WHO CONTACTED THE OFFICE AGAIN AND REPORTS THAT THE SURGEON ASSISTANT FIGURED OUT THAT THE ISSUE WAS WITH THE UNIT PLUGGED INTO A SURGE PROTECTOR. AFTER PLUGGING DIRECTLY INTO THE WALL SOCKET THE ISSUE WAS SOLVED. THE SURGEON DOES NOT BELIEVE THE PRODUCT WAS THE ISSUE. 8/6/2021 - COMPLAINT DETERMINED TO BE REPORTABLE.
Patient problems: Electric Shock
MAUDE report 12316298 ↗
Other HANDPIECE, ROTARY BONE CUTTING devices
- OsteoMed 450-0241Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING9 reports
- Surgical Straight Handpiece SGS-E2SNakanishi Inc. · Handpiece, Rotary Bone Cutting8 reports
- OsteoMed 450-0034Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING6 reports
- S-12 surgical handpiece 1:2 30061000W & H Dentalwerk Bürmoos GmbH · Handpiece, Rotary Bone Cutting3 reports
- OsteoMed 450-0390Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING3 reports
- OsteoMed 450-0084Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING3 reports
- S-11 surgical handpiece 1:1 30056000W & H Dentalwerk Bürmoos GmbH · Handpiece, Rotary Bone Cutting2 reports
- SGA-E2SNakanishi Inc. · HANDPIECE, ROTARY BONE CUTTING2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page