Skip to content

S-12 surgical handpiece 1:2 30061000Recalls & reported problems

W & H Dentalwerk Bürmoos GmbH · Surgical Handpiece · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI J021300610001), received between Feb 1, 2019 and Apr 20, 2026 — classified as 3 injuries.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method1
  • Overheating of Device1
  • Temperature Problem1
  • Use of Device Problem1

Patient problems

  • Burn(s)1
  • Partial thickness (Second Degree) Burn1
  • Superficial (First Degree) Burn1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 20, 2026Temperature Problem

    THE HANDPIECE WAS USED WITH COPIOUS SALINE IRRIGATION (40%) ATTACHED TO SURGICAL HANDPIECE IRRIGATION INPUT. SIDE OF HANDPIECE (NOT BUR) OVERHEATED AND BURNED PATIENTS'S FACE/LIP DURING DECORONATING CROWN OF TOOTH #18. HANDPIECE OVERHEATED TO HIGH LEVEL CAUSED 2ND DEGREE BURN TO PATIENT ALONG VERMILLION BORDER AND ONTO SKIN OF CHIN. COPIOUS SALINE IRRIGATION WAS UTILIZED/ATTACHED TO HANDPIECE; HOWEVER, OVERHEATING WAS A RESULT OF OVERHEATING/MALFUNCTION OF THE BODY OF THE HAN.

    Patient problems: Burn(s)

    MAUDE report 24924982 ↗
  • InjuryApr 16, 2019Use of Device Problem; Improper or Incorrect Procedure or Method; Use of Device Problem; Improper or Incorrect Procedure or Method

    DURING A DENTAL SURGICA TREATMENT THE INSTRUMENT GOT HOT AND IT ENDED UP IN BURNING THE PATIENT'S LIP (2ND DEGREE). THE INJURY IS REVERSIBLE AND THE PATIENT IS UNDER MONITORING. AFTER EVALUATION OF THE SUSPECT DEVICE IT CAME OUT THAT THE ADVERSE EVENT OCCURED DUE TO WRONG LUBRICATION OF THE INSTRUMENT.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 8517118 ↗
  • InjuryFeb 1, 2019Overheating of Device

    THE END-USER STATED THAT THE TIP OF THE HANDPIECE S-12 GOT HOT AND BURNED THE PATIENT. ALSO THE ASSISTANT GOT BURN MARKS AFTER TOUCHING THE HANDPIECE.

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 8295675 ↗

Other Handpiece, Rotary Bone Cutting devices

All Handpiece, Rotary Bone Cutting devices →