S-12 surgical handpiece 1:2 30061000Recalls & reported problems
W & H Dentalwerk Bürmoos GmbH · Surgical Handpiece · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI J021300610001), received between Feb 1, 2019 and Apr 20, 2026 — classified as 3 injuries.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method1
- Overheating of Device1
- Temperature Problem1
- Use of Device Problem1
Patient problems
- Burn(s)1
- Partial thickness (Second Degree) Burn1
- Superficial (First Degree) Burn1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 20, 2026Temperature Problem
THE HANDPIECE WAS USED WITH COPIOUS SALINE IRRIGATION (40%) ATTACHED TO SURGICAL HANDPIECE IRRIGATION INPUT. SIDE OF HANDPIECE (NOT BUR) OVERHEATED AND BURNED PATIENTS'S FACE/LIP DURING DECORONATING CROWN OF TOOTH #18. HANDPIECE OVERHEATED TO HIGH LEVEL CAUSED 2ND DEGREE BURN TO PATIENT ALONG VERMILLION BORDER AND ONTO SKIN OF CHIN. COPIOUS SALINE IRRIGATION WAS UTILIZED/ATTACHED TO HANDPIECE; HOWEVER, OVERHEATING WAS A RESULT OF OVERHEATING/MALFUNCTION OF THE BODY OF THE HAN.
Patient problems: Burn(s)
MAUDE report 24924982 ↗InjuryApr 16, 2019Use of Device Problem; Improper or Incorrect Procedure or Method; Use of Device Problem; Improper or Incorrect Procedure or Method
DURING A DENTAL SURGICA TREATMENT THE INSTRUMENT GOT HOT AND IT ENDED UP IN BURNING THE PATIENT'S LIP (2ND DEGREE). THE INJURY IS REVERSIBLE AND THE PATIENT IS UNDER MONITORING. AFTER EVALUATION OF THE SUSPECT DEVICE IT CAME OUT THAT THE ADVERSE EVENT OCCURED DUE TO WRONG LUBRICATION OF THE INSTRUMENT.
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 8517118 ↗InjuryFeb 1, 2019Overheating of Device
THE END-USER STATED THAT THE TIP OF THE HANDPIECE S-12 GOT HOT AND BURNED THE PATIENT. ALSO THE ASSISTANT GOT BURN MARKS AFTER TOUCHING THE HANDPIECE.
Patient problems: Superficial (First Degree) Burn
MAUDE report 8295675 ↗
Other Handpiece, Rotary Bone Cutting devices
- OsteoMed 450-0241Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING9 reports
- Surgical Straight Handpiece SGS-E2SNakanishi Inc. · Handpiece, Rotary Bone Cutting8 reports
- OsteoMed 450-0034Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING6 reports
- OsteoMed 450-0390Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING3 reports
- OsteoMed 450-0021-00Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING3 reports
- OsteoMed 450-0084Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING3 reports
- S-11 surgical handpiece 1:1 30056000W & H Dentalwerk Bürmoos GmbH · Handpiece, Rotary Bone Cutting2 reports
- SGA-E2SNakanishi Inc. · HANDPIECE, ROTARY BONE CUTTING2 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.