- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI J021300610001), received between Feb 1, 2019 and Apr 20, 2026 — classified as 3 injuries.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method1
- Overheating of Device1
- Temperature Problem1
- Use of Device Problem1
Patient problems
- Burn(s)1
- Partial thickness (Second Degree) Burn1
- Superficial (First Degree) Burn1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 20, 2026Temperature Problem
THE HANDPIECE WAS USED WITH COPIOUS SALINE IRRIGATION (40%) ATTACHED TO SURGICAL HANDPIECE IRRIGATION INPUT. SIDE OF HANDPIECE (NOT BUR) OVERHEATED AND BURNED PATIENTS'S FACE/LIP DURING DECORONATING CROWN OF TOOTH #18. HANDPIECE OVERHEATED TO HIGH LEVEL CAUSED 2ND DEGREE BURN TO PATIENT ALONG VERMILLION BORDER AND ONTO SKIN OF CHIN. COPIOUS SALINE IRRIGATION WAS UTILIZED/ATTACHED TO HANDPIECE; HOWEVER, OVERHEATING WAS A RESULT OF OVERHEATING/MALFUNCTION OF THE BODY OF THE HAN.
Patient problems: Burn(s)
MAUDE report 24924982 ↗InjuryApr 16, 2019Use of Device Problem; Improper or Incorrect Procedure or Method; Use of Device Problem; Improper or Incorrect Procedure or Method
DURING A DENTAL SURGICA TREATMENT THE INSTRUMENT GOT HOT AND IT ENDED UP IN BURNING THE PATIENT'S LIP (2ND DEGREE). THE INJURY IS REVERSIBLE AND THE PATIENT IS UNDER MONITORING. AFTER EVALUATION OF THE SUSPECT DEVICE IT CAME OUT THAT THE ADVERSE EVENT OCCURED DUE TO WRONG LUBRICATION OF THE INSTRUMENT.
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 8517118 ↗InjuryFeb 1, 2019Overheating of Device
THE END-USER STATED THAT THE TIP OF THE HANDPIECE S-12 GOT HOT AND BURNED THE PATIENT. ALSO THE ASSISTANT GOT BURN MARKS AFTER TOUCHING THE HANDPIECE.
Patient problems: Superficial (First Degree) Burn
MAUDE report 8295675 ↗
FDA clearance and classification
The FDA regulates product code KMW (handpiece, rotary bone cutting) under 21 CFR 872.4120 as a Class II device. 21 CFR 872.4120 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the S-12 surgical handpiece 1:2 30061000
Has the S-12 surgical handpiece 1:2 30061000 been recalled?
No FDA recall record in our data names this exact device identifier (DI J021300610001), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the S-12 surgical handpiece 1:2 30061000?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Feb 1, 2019 and Apr 20, 2026. In those reports, the most common specific device problems are “Improper or Incorrect Procedure or Method” (1), “Overheating of Device” (1) and “Temperature Problem” (1); the most common patient problems are “Burn(s)” (1), “Partial thickness (Second Degree) Burn” (1) and “Superficial (First Degree) Burn” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the S-12 surgical handpiece 1:2 30061000?
None of the reports are classified as deaths, 3 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the S-12 surgical handpiece 1:2 30061000, and is it still sold?
The FDA's device database lists W & H Dentalwerk Bürmoos GmbH as the company responsible for it (brand S-12 surgical handpiece 1:2, model 30061000). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Handpiece, Rotary Bone Cutting devices
- OsteoMed 450-0241Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING9 reports
- Surgical Straight Handpiece SGS-E2SNakanishi Inc. · Handpiece, Rotary Bone Cutting8 reports
- OsteoMed 450-0034Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING6 reports
- OsteoMed 450-0390Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING3 reports
- OsteoMed 450-0021-00Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING3 reports
- OsteoMed 450-0084Osteomed LLC · HANDPIECE, ROTARY BONE CUTTING3 reports
- SGA-E2SNakanishi Inc. · HANDPIECE, ROTARY BONE CUTTING2 reports
- S-11 surgical handpiece 1:1 30056000W & H Dentalwerk Bürmoos GmbH · Handpiece, Rotary Bone Cutting2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.